Development of bone chip-vacuum system in orthopedic drilling process

被引:11
作者
Gok, Kadir [1 ]
Erdem, Mehmet [2 ]
Kisioglu, Yasin [3 ]
Gok, Arif [4 ]
Tumsek, Mert [5 ]
机构
[1] Izmir Bakircay Univ, Dept Biomed Engn, Fac Engn & Architecture, TR-35660 Izmir, Turkey
[2] Kutahya Dumlupinar Univ, Kutahya Vocat Sch Tech Sci Mech & Met Technol, Germiyan Campus, Kutahya, Turkey
[3] Kocaeli Univ, Technol Fac, Dept Biomed Engn, TR-41380 Kocaeli, Turkey
[4] Kutahya Dumlupinar Univ, Architecture Fac, Dept Ind Design, Kutahya, Turkey
[5] Manisa Celal Bayar Univ, Nat Sci Inst, Dept Mech & Mfg Engn, Manisa, Turkey
关键词
Orthopedic surgical drilling; Necrosis; Collecting bone chips; Statistical analysis;
D O I
10.1007/s40430-021-02959-w
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Overheating occurs during the bone drilling process using drill bits in orthopedic surgery. The temperatures frequently exceed 47 degrees C, usually accepted as critical value, since the bones and surrounding tissues burn and then cause the necrosis during the drilling process. To prevent this, orthopedic surgeons either spray coolants to the drilling zone or pause the drilling process until the bones and tissues cool down. Such heat damage in bones and soft tissues is undesirable event for surgeons. The bone chips, sometimes, are also complicating the bone fracture fixations. In this study, an investigation is described the effects of the processing parameters on bone chip formations, temperature levels and chip-vacuuming performances during drilling. The temperature levels are also measured using thermocouple devices. A bone chip collecting system is developed as a specific device to collect the chips or fragments using vacuuming technique. The collected bone chips are stored in a pot (reservoir) and delivered to a reparative site in case of necessary. Some design parameters for the developed vacuum system are identified considering the temperature rise during the drilling process. Analyzing the effects of these parameters on chip formation and developed vacuum system statistically and some results are presented. At the end of the study, the bone chip-vacuum device was developed and performed successfully to collect the whole fragments in the bone drilling experimental tests. The chip-collecting device was also useful to remove overheat from the drilling zone. The optimal level of the processing parameters was calculated as A1B3C1 for both vacuumed and non-vacuumed conditions. The fractured powder-type chips were formed as anticipated using the cortical structure of diaphyseal part of calf bone which is helpful for chip vacuuming device.
引用
收藏
页数:11
相关论文
共 36 条
  • [1] MEASUREMENTS OF SHAFT SPEED WHILE DRILLING THROUGH BONE
    ABOUZGIA, MB
    JAMES, DF
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1995, 53 (11) : 1308 - 1315
  • [2] Effect of surface topography on pull-out strength of cortical screw after ultrasonic bone drilling: an in vitro study
    Agarwal, Raj
    Jain, Vivek
    Gupta, Vishal
    Saxena, Sanjai
    Dwibedi, Vagish
    [J]. JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2020, 42 (07)
  • [3] Experimental investigations of forces and torque in conventional and ultrasonically-assisted drilling of cortical bone
    Alam, K.
    Mitrofanov, A. V.
    Silberschmidt, V. V.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2011, 33 (02) : 234 - 239
  • [4] [Anonymous], 2000, DESIGN ANAL EXPT
  • [5] Thermal osteonecrosis and bone drilling parameters revisited
    Augustin, Goran
    Davila, Slavko
    Mihoci, Kristijan
    Udiljak, Toma
    Vedrina, Denis Stjepan
    Antabak, Anko
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (01) : 71 - 77
  • [6] Basiaga M, 2011, ACTA BIOENG BIOMECH, V13, P29
  • [7] Cordioli G, 1997, INT J ORAL MAX IMPL, V12, P186
  • [8] COSTICH EMMETT R., 1964, ORAL SURG ORAL MED ORAL PATHOL, V17, P563, DOI 10.1016/0030-4220(64)90359-7
  • [9] An experimental comparative study of drilling efficiency and temperature elevation with unmodified and modified medical drills in pig tibia bone
    Enokida, Makoto
    Kanaya, Haruhisa
    Uehara, Kazutake
    Ueki, Masaru
    Nagashima, Hideki
    [J]. HELIYON, 2019, 5 (08)
  • [10] HEAT CAUSED BY DRILLING CORTICAL BONE - TEMPERATURE MEASURED INVIVO IN PATIENTS AND ANIMALS
    ERIKSSON, AR
    ALBREKTSSON, T
    ALBREKTSSON, B
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1984, 55 (06): : 629 - 631