A novel EEG artifact in the intensive care unit

被引:7
|
作者
Young, B
Osvath, L
Jones, D
Socha, E
机构
[1] London Hlth Sci Ctr, EEG Lab, London, ON, Canada
[2] XLTEK Corp, Oakville, ON, Canada
关键词
EEG; intensive care unit; artifact; hemoperfusion; psychomotor variant;
D O I
10.1097/00004691-200210000-00011
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The authors describe a novel EEG artifact in two patients whose records were otherwise suppressed. The saw-toothed waveform, containing 5.5- and 11-Hz rhythms, was induced as a movement artifact by a hemoperfusion machine connected to the patient. The artifact is highly characteristic, but could be easily misinterpreted as a cerebral rhythm if one is not aware of the entity. The artifact should be suspected when invariant, saw-toothed rhythms appear beyond the temporal and occipital lobes, and also contaminate the electrocardiographic channel. It can be proved by its disappearance on stopping and reemergence with starting the rotary pump action of the device.
引用
收藏
页码:484 / 486
页数:3
相关论文
共 50 条
  • [21] EEG-Video-Monitoring in the Intensive Care Unit
    Rating, D.
    ZEITSCHRIFT FUR EPILEPTOLOGIE, 2014, 27 (03): : 211 - 212
  • [22] Continuous EEG Monitoring in the Neonatal Intensive Care Unit
    McCoy, Blathnaid
    Hahn, Cecil D.
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2013, 30 (02) : 106 - 114
  • [23] Continuous EEG monitoring in the intensive care unit: An overview
    Hirsch, LJ
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2004, 21 (05) : 332 - 340
  • [24] EEG in the Pediatric Intensive Care Unit: An Irish Experience
    MacDarby, Laura
    Healy, Martina
    Curley, Gerard
    McHugh, John C.
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2021, 38 (02) : 130 - 134
  • [25] Continuous EEG monitoring (eegc) in the intensive care unit setting
    Mesa, T.
    Solari, F.
    Santin, J.
    Godoy, J.
    Kuester, G.
    Andrade, L.
    EPILEPSIA, 2007, 48 : 99 - 99
  • [26] EEG monitoring in the neonatal intensive care unit: A critical juncture
    Boylan, Geraldine B.
    CLINICAL NEUROPHYSIOLOGY, 2011, 122 (10) : 1905 - 1907
  • [27] Continuous EEG recording in the intensive care unit: epileptiform activity
    GB Young
    Critical Care, 7 (Suppl 2):
  • [28] An international survey of EEG use in the neonatal intensive care unit
    Boylan, G. B.
    Burgoyne, L.
    Moore, C.
    O'Flaherty, B.
    Rennie, J. M.
    ACTA PAEDIATRICA, 2010, 99 (08) : 1150 - 1155
  • [29] Sensitivity of quantitative EEG for seizure identification in the intensive care unit
    Haider, Hiba A.
    Esteller, Rosana
    Hahn, Cecil D.
    Westover, M. Brandon
    Halford, Jonathan J.
    Lee, Jong W.
    Shafi, Mouhsin M.
    Gaspard, Nicolas
    Herman, Susan T.
    Gerard, Elizabeth E.
    Hirsch, Lawrence J.
    Ehrenberg, Joshua A.
    LaRoche, Suzette M.
    NEUROLOGY, 2016, 87 (09) : 935 - 944
  • [30] Continuous EEG monitoring in children in the intensive care unit (ICU)
    Nguyen The Tich, S.
    Cheliout-Heraut, F.
    NEUROPHYSIOLOGIE CLINIQUE-CLINICAL NEUROPHYSIOLOGY, 2015, 45 (01): : 75 - 80