A novel EEG artifact in the intensive care unit

被引:7
|
作者
Young, B
Osvath, L
Jones, D
Socha, E
机构
[1] London Hlth Sci Ctr, EEG Lab, London, ON, Canada
[2] XLTEK Corp, Oakville, ON, Canada
关键词
EEG; intensive care unit; artifact; hemoperfusion; psychomotor variant;
D O I
10.1097/00004691-200210000-00011
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The authors describe a novel EEG artifact in two patients whose records were otherwise suppressed. The saw-toothed waveform, containing 5.5- and 11-Hz rhythms, was induced as a movement artifact by a hemoperfusion machine connected to the patient. The artifact is highly characteristic, but could be easily misinterpreted as a cerebral rhythm if one is not aware of the entity. The artifact should be suspected when invariant, saw-toothed rhythms appear beyond the temporal and occipital lobes, and also contaminate the electrocardiographic channel. It can be proved by its disappearance on stopping and reemergence with starting the rotary pump action of the device.
引用
收藏
页码:484 / 486
页数:3
相关论文
共 50 条
  • [1] Rhythmic artifact of physiotherapy in intensive care unit EEG recordings
    Young, Bryan
    Raihan, Syed
    Ladak, H.
    Kelly, Martin
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2007, 24 (03) : 252 - 256
  • [2] Rhythmic artifact of physiotherapy in intensive care unit EEG recordings
    Sethi, Nitin K.
    Torgovnick, Josh
    Sethi, Prahlad K.
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2008, 25 (01) : 62 - 62
  • [3] EEG in the intensive care unit
    Thoemke, F.
    KLINISCHE NEUROPHYSIOLOGIE, 2006, 37 (04) : 233 - 241
  • [4] EEG monitoring in the intensive care unit
    Dunne, John W.
    NEUROLOGY ASIA, 2013, 18 : 19 - 21
  • [5] Continuous EEG monitoring in the intensive care unit
    Kay, J
    CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 1998, 25 (01) : S12 - S15
  • [6] THE EEG IN THE NEUROSURGICAL INTENSIVE-CARE UNIT
    KRAMER, M
    KUHLENDAHL, H
    STAHL, HJ
    KUNKEL, H
    LUKEN, H
    ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1980, 50 (3-4): : P127 - P127
  • [7] EEG and evoked potentials in the intensive care unit
    Guérit, JM
    NEUROPHYSIOLOGIE CLINIQUE-CLINICAL NEUROPHYSIOLOGY, 1999, 29 (04): : 301 - 317
  • [8] Continuous EEG Monitoring in the Intensive Care Unit
    Jeffrey D. Kennedy
    Elizabeth E. Gerard
    Current Neurology and Neuroscience Reports, 2012, 12 : 419 - 428
  • [9] EEG artifacts in the intensive care unit setting
    White D.M.
    Van Cott A.C.
    Neurodiagnostic Journal, 2010, 50 (01) : 8 - 25
  • [10] Continuous EEG monitoring in the intensive care unit
    Nuwer, MR
    CLINICAL NEUROPHYSIOLOGY: FROM RECEPTORS TO PERCEPTION, 1999, 50 : 150 - 155