Viscosity Measurement Sensor: A Prototype for a Novel Medical Diagnostic Method Based on Quartz Crystal Resonator

被引:11
作者
Miranda-Martinez, Andres [1 ]
Rivera-Gonzalez, Marco Xavier [1 ]
Zeinoun, Michael [1 ]
Carvajal-Ahumada, Luis Armando [2 ]
Serrano-Olmedo, Jose Javier [1 ,3 ]
机构
[1] Univ Politecn Madrid UPM, Ctr Biomed Technol CTB, Madrid 28040, Spain
[2] Univ Manuela Beltran, Bogota Campus, Bogota 110231, Colombia
[3] Univ Politecn Madrid UPM, Networking Res Ctr Bioengn Biomat & Nanomed CIBER, Madrid 28040, Spain
关键词
quartz crystal resonator; viscosity; synovial fluid; cerebrospinal fluid; diagnostic sensor; CEREBROSPINAL-FLUID; HYALURONIC-ACID; MICROBALANCE; EPIDEMIOLOGY;
D O I
10.3390/s21082743
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Viscosity variation in human fluids, such as Synovial Fluid (SF) or Cerebrospinal Fluid (CSF), can be used as a diagnostic factor; however, the sample volume obtained for analysis is usually small, making it difficult to measure its viscosity. On the other hand, Quartz Crystal Resonators (QCR) have been used widely in sensing applications due to their accuracy, cost, and size. This work provides the design and validation of a new viscosity measurement system based on quartz crystal resonators for low volume fluids, leading to the development of a sensor called "ViSQCT" as a prototype for a new medical diagnostic tool. The proposed method is based on measuring the resonance frequency at the crystal's maximum conductance point through a frequency sweep, where crystals with 10 MHz fundamental resonance frequency were used. For validation purposes, artificial fluids were developed to simulate SFs and CFs in healthy and pathological conditions as experiment phantoms. A commercial QCR based system was also used for validation since its methodology differs from ours. A conventional rotational viscometer was used as a reference for calibration purposes. ViSQCT demonstrates the capability to measure the sample's viscosity differentiation between healthy and pathological fluid phantoms and shows that it can be used as a basis for a diagnostic method of several pathologies related to the studied biological fluids. However, some performance differences between both QCR-based systems compared to the reference system deserves further investigation.
引用
收藏
页数:15
相关论文
共 42 条
[1]   Viscosity measurement of industrial oils using the droplet quartz crystal microbalance [J].
Ash, DC ;
Joyce, MJ ;
Barnes, C ;
Booth, CJ ;
Jefferies, AC .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2003, 14 (11) :1955-1962
[2]   Synovial fluid analysis [J].
Brannan, SR ;
Jerrard, DA .
JOURNAL OF EMERGENCY MEDICINE, 2006, 30 (03) :331-339
[3]   Epidemiology, Diagnosis, and Antimicrobial Treatment of Acute Bacterial Meningitis [J].
Brouwer, Matthijs C. ;
Tunkel, Allan R. ;
van de Beek, Diederik .
CLINICAL MICROBIOLOGY REVIEWS, 2010, 23 (03) :467-+
[4]   Evaluation of Hyaluronic Acid Dilutions at Different Concentrations Using a Quartz Crystal Resonator (QCR) for the Potential Diagnosis of Arthritic Diseases [J].
Carvajal Ahumada, Luis Armando ;
Rivera Gonzalez, Marco Xavier ;
Herrera Sandoval, Oscar Leonardo ;
Serrano Olmedo, Jose Javier .
SENSORS, 2016, 16 (11)
[5]   A new way to find dielectric properties of liquid sample using the quartz crystal resonator (QCR) [J].
Carvajal Ahumada, Luis Armando ;
Pena Perez, Nuria ;
Herrera Sandoval, Oscar Leonardo ;
del Pozo Guerrero, Francisco ;
Serrano Olmedo, Jose Javier .
SENSORS AND ACTUATORS A-PHYSICAL, 2016, 239 :153-160
[6]   Characterization of the behaviour of a quartz crystal resonator fully immersed in a Newtonian liquid by impedance analysis [J].
Cassiede, M. ;
Daridon, J. -L. ;
Paillol, J. H. ;
Pauly, J. .
SENSORS AND ACTUATORS A-PHYSICAL, 2011, 167 (02) :317-326
[7]   Viral meningitis [J].
Chadwick, David R. .
BRITISH MEDICAL BULLETIN, 2005, 75-76 :1-14
[8]   Immobilization of serum albumin and peptide aptamer for EPC on polydopamine coated titanium surface for enhanced in-situ self-endothelialization [J].
Chen, Zhuoyue ;
Li, Quanli ;
Chen, Jialong ;
Luo, Rifang ;
Maitz, Manfred F. ;
Huang, Nan .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2016, 60 :219-229
[9]  
Damiano J., 2004, EMC RHUMATOLOGIE ORT, V1, P2, DOI [10.1016/j.emcrho.2003.11.002, DOI 10.1016/J.EMCRHO.2003.11.002]
[10]   A review of quartz crystal microbalances for space applications [J].
Dirri, Fabrizio ;
Palomba, Ernesto ;
Longobardo, Andrea ;
Zampetti, Emiliano ;
Saggin, Bortolino ;
Scaccabarozzi, Diego .
SENSORS AND ACTUATORS A-PHYSICAL, 2019, 287 :48-75