Validation of speed-resolved laser Doppler perfusion in a multimodal optical system using a blood-flow phantom

被引:17
|
作者
Jonasson, Hanna [1 ]
Fredriksson, Ingemar [1 ,2 ]
Larsson, Marcus [1 ]
Stromberg, Tomas [1 ]
机构
[1] Linkoping Univ, Dept Biomed Engn, Linkoping, Sweden
[2] Perimed AB, Stockholm, Sweden
关键词
laser Doppler; perfusion; blood flow; phantom; MONTE-CARLO SIMULATIONS; TURBID MEDIUM; DECOMPOSITION; SPECTRUM; HEAD;
D O I
10.1117/1.JBO.24.9.095002
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The PeriFlux 6000 EPOS system combines diffuse reflectance spectroscopy (DRS) and laser Doppler flowmetry (LDF) for the assessment of oxygen saturation (expressed in percentage), red blood cell (RBC) tissue fraction (expressed as volume fraction, %RBC), and perfusion (%RBC x mm/s) in the microcirculation. It also allows the possibility of separating the perfusion into three speed regions (0 to 1, 1 to 10, and >10 mm/s). We evaluate the speed-resolved perfusion components, i.e., the relative amount of perfusion within each speed region, using a blood-flow phantom. Human blood was pumped through microtubes with an inner diameter of 0.15 mm. Measured DRS and LDF spectra were compared to Monte Carlo-simulated spectra in an optimization routine, giving the best-fit parameters describing the measured spectra. The root-mean-square error for each of the three speed components (0 to 1, 1 to 10, and >10 mm/s, respectively) when describing the blood-flow speed in the microtubes was 2.9%, 8.1%, and 7.7%. The presented results show that the system can accurately discriminate blood perfusion originating from different blood-flow speeds, which may enable improved measurement of healthy and dysfunctional microcirculatory flow. (C) The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License.
引用
收藏
页数:8
相关论文
共 13 条
  • [1] Full-field high-speed laser Doppler imaging system for blood-flow measurements
    Serov, Alexandre
    Lasser, Theo
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS IV, 2006, 6080
  • [2] Speed-resolved perfusion imaging using multi-exposure laser speckle contrast imaging and machine learning
    Hultman, Martin
    Larsson, Marcus
    Stromberg, Tomas
    Fredriksson, Ingemar
    JOURNAL OF BIOMEDICAL OPTICS, 2023, 28 (03) : 36007
  • [3] Mucosal blood flow measurements using laser Doppler perfusion monitoring
    Dag Arne Lihaug Hoff
    Hans Gregersen
    Jan Gunnar Hatlebakk
    World Journal of Gastroenterology, 2009, 15 (02) : 198 - 203
  • [4] Mucosal blood flow measurements using laser Doppler perfusion monitoring
    Hoff, Dag Arne Lihaug
    Gregersen, Hans
    Hatlebakk, Jan Gunnar
    WORLD JOURNAL OF GASTROENTEROLOGY, 2009, 15 (02) : 198 - 203
  • [5] Cortical blood flow autoregulation revisited using laser Doppler perfusion imaging
    Kimme, P
    Ledin, T
    Sjöberg, F
    ACTA PHYSIOLOGICA SCANDINAVICA, 2002, 176 (04): : 255 - 262
  • [6] 2-DIMENSIONAL BLOOD-FLOW DETERMINATIONS IN ALLERGIC REACTIONS USING LASER-DOPPLER SCANNING
    STUCKER, M
    AUER, T
    HOFFMANN, K
    ALTMEYER, P
    CONTACT DERMATITIS, 1995, 33 (05) : 299 - 303
  • [7] Full-field laser Doppler blood-flow imaging and monitoring using an intelligent CMOS camera and area illumination
    Serov, A
    Steinacher, B
    Lasser, T
    SARATOV FALL MEETING 2004: OPTICAL TECHNOLOGIES IN BIOPHYSICS AND MEDICINE VI, 2005, 5771 : 95 - 102
  • [8] MEASUREMENT OF DIGITAL BLOOD-FLOW USING THE LASER DOPPLER, IMPEDANCE AND STRAIN-GAUGE METHODS
    YAMAMOTO, Y
    OBERG, PA
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (02) : 113 - 118
  • [9] SIMPLE SYSTEM FOR AUTOMATIC INTERMITTENT RECORDING OF BLOOD-FLOW IN FEMORODISTAL BYPASS GRAFTS USING DOPPLER ULTRASOUND
    THRUSH, AJ
    EVANS, DH
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (02) : 193 - 195
  • [10] Skeletal muscle blood flow after prolonged tourniquet ischaemia and reperfusion with and without intervening reoxygenation:: An experimental study in rats using laser Doppler perfusion imaging
    Sirsjö, A
    Gidlöf, A
    Nilsson, G
    Povlsen, B
    SCANDINAVIAN JOURNAL OF PLASTIC AND RECONSTRUCTIVE SURGERY AND HAND SURGERY, 1999, 33 (03): : 281 - 285