共 7 条
Transmission of Salmonella enteritidis after endoscopic retrograde cholangiopancreatography because of inadequate endoscope decontamination
被引:22
|作者:
Robertson, Paul
[1
]
Smith, Andrew
[2
]
Anderson, Margaret
[3
]
Stewart, Jackie
[2
]
Hamilton, Kate
[2
]
McNamee, Sandra
[2
]
Curran, Evonne T.
[2
]
机构:
[1] Queen Elizabeth Univ Hosp, Microbiol Dept, 1345 Govan Rd, Glasgow G51 4TF, Lanark, Scotland
[2] North Glasgow Univ Hosp Natl Hlth Serv Trust, Infect Control Team, Glasgow, Lanark, Scotland
[3] Stobhill Hosp, Day Surg Unit, Glasgow, Lanark, Scotland
关键词:
Endoscope;
Decontamination;
Salmonella;
GASTROINTESTINAL ENDOSCOPY;
INFECTION-CONTROL;
D O I:
10.1016/j.ajic.2016.11.024
中图分类号:
R1 [预防医学、卫生学];
学科分类号:
1004 ;
120402 ;
摘要:
We report a historic nosocomial outbreak of Salmonella enteritidis affecting 4 inpatients who underwent endoscopic retrograde cholangiopancreatography. The cause was attributed to inadequate decontamination of an on-loan endoscope used over a weekend. This report highlights the risks of using on-loan endoscopes, particularly regarding their commissioning and adherence to disinfection protocols. In an era of increasing antibiotic resistance, transmission of Enterobacteriaceae by endoscopes remains a significant concern. (C) 2017 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.