Measurement of polyethylene wear in acetabular components inserted with and without cement - A randomized trial

被引:99
作者
Devane, PA
Robinson, EJ
Bourne, RB
Rorabeck, CH
Nayak, NN
Horne, JG
机构
[1] UNIV WESTERN ONTARIO HOSP, DEPT ORTHOPAED SURG, LONDON, ON N6A 5A5, CANADA
[2] UNIV HOSP, DEPT ORTHOPAED SURG, LONDON, ON, CANADA
关键词
D O I
10.2106/00004623-199705000-00007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We measured the three-dimensional wear of polyethylene after total hip arthroplasty with a titanium metal-backed Mallory-Head prosthesis that was inserted with cement in sixty-nine patients (sixty-nine hips) and with a press-fit titanium metal-backed Mallory-Head prosthesis that was inserted without cement in seventy patients (seventy hips), A modular titanium femoral head was used in all of the hips, The patients in the present study were part of a larger double-blind randomized trial comparing the result of total hip arthroplasty performed with cement with that of the same procedure performed without cement in 250 patients, The criterion for inclusion in the study of polyethylene wear was a minimum duration of followup of four years, which was met by 148 patients, As adequate radiographs for digitization were not available for nine patients, 139 patients were included in the present study, The age of the patient, the postoperative Harris hip score, the diameter of the femoral head, the thickness of the liner in the polar region of the acetabular component, and the duration of follow-up were similar for the two groups, The mean rate of volumetric wear of the polyethylene was significantly greater in the prostheses that had been inserted without cement than in those that had been inserted with cement (155.1 cubic millimeters per year compared with 98.5 cubic millimeters per gear; p=0.000008). Thirty-four (49 per cent) of the seventy hips in which the prosthesis had been inserted without cement had evidence of osteolysis on radiographs, compared with twelve (17 per cent) of the sixty-nine hips in the other group (p=0.0002), Osteolysis was associated with an increased rate of polyethylene wear only in the hips in which the prosthesis had been inserted without cement.
引用
收藏
页码:682 / 689
页数:8
相关论文
共 38 条
  • [1] [Anonymous], 1979, CLIN ORTHOP RELAT RE
  • [2] [Anonymous], CLIN ORTHOPAEDICS
  • [3] THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT
    BARTEL, DL
    BICKNELL, VL
    WRIGHT, TM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) : 1041 - 1051
  • [4] DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES
    BRIEN, WW
    SALVATI, EA
    WRIGHT, TM
    NELSON, CL
    HUNGERFORD, DS
    GILLIAM, DL
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) : 1548 - 1550
  • [5] Buly R L, 1992, J Arthroplasty, V7, P315, DOI 10.1016/0883-5403(92)90056-V
  • [6] POLYETHYLENE WEAR IN CEMENTED METAL-BACKED ACETABULAR CUPS
    CATES, HE
    FARIS, PM
    KEATING, EM
    RITTER, MA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02): : 249 - 253
  • [7] Charnley J, 1973, CLIN ORTHOP RELAT R, V95, P9
  • [8] CHILLAG KJ, 1991, CLIN ORTHOP RELAT R, V273, P261
  • [9] COLLIER JP, 1990, CLIN ORTHOP RELAT R, V261, P107
  • [10] Cooper R A, 1992, J Arthroplasty, V7, P285, DOI 10.1016/0883-5403(92)90050-Z