A Clampless and Sutureless Aorto-Prosthetic End-to-Side Anastomotic Device: An Experimental Study

被引:6
|
作者
Alimi, Y. S. [1 ,2 ]
Saint Lebes, B. [1 ,2 ]
Garitey, V. [3 ]
Afrapoli, A. [1 ,2 ]
Boufi, M. [1 ,2 ]
Hartung, O. [1 ,2 ]
Garcia, S. [4 ]
Mouret, F. [3 ]
Berdah, S. [1 ,2 ]
机构
[1] Ctr Hosp Univ Nord, Serv Chirurg Vasc, Marseille, France
[2] Univ Mediterranee, Ctr Enseignement & Rech Chirurg, Marseille, France
[3] Protomed, Marseille, France
[4] Ctr Hosp Univ Nord, Serv Anatomopathol, Marseille, France
关键词
Aortic surgery; Anastomosis; Pigs; Study; OCCLUSIVE DISEASE; SURGERY; RECONSTRUCTION; BYPASS;
D O I
10.1016/j.ejvs.2009.07.003
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: A feasibility study. Methods: Eight pigs (all females; mean weight: 29 kg) underwent a conventional transperitoneal aortic approach with implantation of an aorto-prosthetic end-to-side anastomosis using a Clampless (R) device and deployment of a 5-mm polytetrafluoroethylene (PTFE) graft. After proximal ligature, a conventional end-to-end anastomosis was then performed between the graft and the left iliac artery. Results: The first pig died during the procedure due to graft misplacement. The seven other procedures were successful with a mean operative and anastomosis time of 101 min (range: 81-115 min) and 3.35 min (range: 2.25-4.25 min), respectively; mean blood loss was 152 ml (range: 30-235 ml). Another pig with a patent graft died at day 4 as a result of a severe unrelated pneumonopathy. The angiogram performed during the procedure and before sacrifice, at 2 (n = 2), 4 (n = 2) and 6 weeks (n = 2), showed no graft stenosis or thrombosis. Microscopic examination revealed a tissue covering the intraluminal stent, which evolved over time, with no visible endothelial proliferation or inflammation. Conclusion: An aorto-prosthetic anastomosis can be performed safety and efficiently with our new clampless and sutureless device. The next step will be a laparoscopic Clampless (R) implantation. (C) 2009 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:597 / 602
页数:6
相关论文
共 40 条
  • [1] A clampless and sutureless aorto-prosthetic anastomosis device
    Boufi, M.
    Alimi, Y. a
    E-MEMOIRES DE L ACADEMIE NATIONALE DE CHIRURGIE, 2011, 10 (01): : 59 - 64
  • [2] New microvascular anastomotic device for end-to-side anastomosis using negative pressure; a preliminary study
    Tachi, Kazufumi
    Furukawa, Katsuko S.
    Koshima, Isao
    Ushida, Takashi
    JOURNAL OF PLASTIC SURGERY AND HAND SURGERY, 2020, 54 (03) : 167 - 171
  • [3] Hemodynamics of an End-to-Side Anastomotic Graft for a Pulsatile Pediatric Ventricular Assist Device
    Yang, Ning
    Deutsch, Steven
    Paterson, Eric G.
    Manning, Keefe B.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2010, 132 (03):
  • [4] Flow waveform effects on end-to-side anastomotic flow patterns
    Ethier, CR
    Steinman, DA
    Zhang, X
    Karpik, SR
    Ojha, M
    JOURNAL OF BIOMECHANICS, 1998, 31 (07) : 609 - 617
  • [5] CorLink™ sutureless aortic anastomotic device:: Results of an animal study
    Bar-El, Y
    Tio, FO
    Shofti, R
    JOURNAL OF SURGICAL RESEARCH, 2003, 115 (01) : 127 - 132
  • [6] A comparative study on the complications of conventional and end-to-side inserting pancreatojejunostomy after pancreaticoduodenectomy
    Hashimoto, Daisuke
    Chikamoto, Akira
    Harimoto, Norifumi
    Ikegami, Toru
    Uchiyama, Hideaki
    Yoshizumi, Tomoharu
    Baba, Hideo
    Maehara, Yoshihiko
    SURGERY TODAY, 2017, 47 (02) : 238 - 244
  • [7] Preliminary Results from a Prospective Study of Laparoscopic Aortobifemoral Bypass Using a Clampless and Sutureless Aortic Anastomotic Technique
    Segers, B.
    Horn, D.
    Lemaitre, J.
    Roman, A.
    Stevens, E.
    Van den Broeck, V.
    Hizette, P.
    Bosschaerts, T.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2014, 48 (04) : 400 - 406
  • [8] Microvascular anastomotic coupler for venous end-to-side anastomoses in head and neck reconstructive surgery
    Ulrich Kisser
    Sophie Koepernik
    European Archives of Oto-Rhino-Laryngology, 2023, 280 : 4709 - 4712
  • [9] Comparative Study of Continuous and Pulsatile Left Ventricular Assist Devices on Hemodynamics of a Pediatric End-to-Side Anastomotic Graft
    Yang N.
    Deutsch S.
    Paterson E.G.
    Manning K.B.
    Cardiovascular Engineering and Technology, 2010, 1 (01) : 88 - 103
  • [10] Microvascular anastomotic coupler for venous end-to-side anastomoses in head and neck reconstructive surgery