Data analysis of dose map verification for IMRT

被引:0
作者
Xu Yingjie [1 ]
Dai Jianrong [1 ]
Fu Guishan [1 ]
Zhang Yongqian [1 ]
Zhang Ke [1 ]
Cui Weijie [1 ]
Zhang Dandan [1 ]
Jiang Bo [1 ]
机构
[1] Chinese Acad Med Sci, Dept Radiat Oncol, Inst Canc, Beijing 100021, Peoples R China
关键词
Dose map verification; Intensity-modulated radiation therapy; Beam modeling; MODULATED RADIATION-THERAPY; QUALITY-ASSURANCE; DIODE-ARRAY;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this paper we analyze the results of dose map verifications for patient's IMRT (Intensity-modulated radiation therapy) plans and study the factors that may influence the accuracy of verification. MapCHECK, a two-dimensional diode array, was used to measure the dose maps for 1242 plans (14540 fields) from May 2004 to August 2008. The plans were designed with Pinnacle 3 planning system. The passing rate of beams was determined when the acceptance criterion was 2%/2 mm, 3%/3 mm and 4%/4 mm. And the data with 3%/3 mm criteria was analyzed in more detail. The considered factors were beam modeling, optimization mode and treatment site. The median passing rate of total fields was 93.5%, 98.8%, and 100% when the acceptance criterion was 2%/2 mm, 3%/3 mm. and 4%/4 mm, and the interquartile range were 11.1%, 3.8%, and 1.3%, respectively. The results of direct machine parameter optimization (DMPO) planning mode was better than those of multiple-step mode and beam modeling of new accelerators was better than that of old accelerators. These indicate that beam modeling is the key point of improving passing rate of IMRT verification and the influence of treatment site was little. The factors, the total number of segments, minimum area of segments and minimum monitor unit (MU) of segments, also influence the dosimetric accuracy of IMRT plan verification.
引用
收藏
页码:297 / 301
页数:5
相关论文
共 9 条
  • [1] Both S, 2007, J APPL CLIN MED PHYS, V8, P1
  • [2] Verification of IMRT fields by film dosimetry
    Bucciolini, M
    Buonamici, FB
    Casati, M
    [J]. MEDICAL PHYSICS, 2004, 31 (01) : 161 - 168
  • [3] Comparison of measured and computed portal dose for IMRT treatment
    Cilla, Savino
    Viola, Pietro
    Azario, Luigi
    Grimaldi, Luca
    Craus, Maurizio
    D'Onofrio, Guido
    Fidanzio, Andrea
    Deodato, Francesco
    Macchia, Gabriella
    Digesu, Cinzia
    Morganti, Alessio G.
    Piermattei, Angelo
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (03): : 65 - 79
  • [4] Clinical evaluation of direct aperture optimization when applied to head-and-neck IMRT
    Jones, Stephen
    Williams, Matthew
    [J]. MEDICAL DOSIMETRY, 2008, 33 (01) : 86 - 92
  • [5] A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery
    Jursinic, PA
    Nelms, BE
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 870 - 879
  • [6] Evaluation of a 2D diode array for IMRT quality assurance
    Létourneau, D
    Gulam, M
    Yan, D
    Oldham, M
    Wong, JW
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 70 (02) : 199 - 206
  • [7] Nelms B.E., 2007, J Appl Clin Med Phys, V8, P2448
  • [8] *SUN NUCL CORP, 2004, MAPCHECK MOD 1175 MA
  • [9] Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPID
    Talamonti, C.
    Casati, M.
    Bucciolini, M.
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4367 - 4378