Electrochemical aptasensor for tetracycline detection

被引:165
作者
Kim, Yoon-Jin [1 ]
Kim, Yeon Seok [1 ]
Niazi, Javed H. [1 ]
Gu, Man Bock [1 ]
机构
[1] Korea Univ, Sch Life Sci & Biotechnol, Seoul 136701, South Korea
关键词
Electrochemical detection; Aptasensor; DNA aptamers; Tetracycline; APTAMER-BASED SENSORS; SENSING PLATFORM; MAGNETIC BEADS; GOLD ELECTRODE; ANTIBIOTICS; BIOSENSOR; PROTEIN; ASSAY; DISPLACEMENT; RESIDUES;
D O I
10.1007/s00449-009-0371-4
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
An electrochemical aptasensor was developed for the detection of tetracycline using ssDNA aptamer that selectively binds to tetracycline as recognition element. The aptamer was highly selective for tetracycline which distinguishes minor structural changes on other tetracycline derivatives. The biotinylated ssDNA aptamer was immobilized on a streptavidin-modified screen-printed gold electrode, and the binding of tetracycline to aptamer was analyzed by cyclic voltammetry and square wave voltammetry. Our results showed that the minimum detection limit of this sensor was 10 nM to micromolar range. The aptasensor showed high selectivity for tetracycline over the other structurally related tetracycline derivatives (oxytetracycline and doxycycline) in a mixture. The aptasensor developed in this study can potentially be used for detection of tetracycline in pharmaceutical preparations, contaminated food products, and drinking water.
引用
收藏
页码:31 / 37
页数:7
相关论文
共 47 条
[1]  
Baker BR, 2006, J AM CHEM SOC, V128, P3138, DOI 10.1021/ja056957p
[2]   Displacement enzyme linked aptamer assay [J].
Baldrich, E ;
Acero, JL ;
Reekmans, G ;
Laureyn, W ;
O'Sullivan, CK .
ANALYTICAL CHEMISTRY, 2005, 77 (15) :4774-4784
[3]   Different approaches for the detection of thrombin by an electrochemical aptamer-based assay coupled to magnetic beads [J].
Centi, S. ;
Messina, G. ;
Tombelli, S. ;
Palchetti, I. ;
Mascini, M. .
BIOSENSORS & BIOELECTRONICS, 2008, 23 (11) :1602-1609
[4]   Aptamer-based detection of plasma proteins by an electrochemical assay coupled to magnetic beads [J].
Centi, Sonia ;
Tombelli, Sara ;
Minunni, Maria ;
Mascini, Marco .
ANALYTICAL CHEMISTRY, 2007, 79 (04) :1466-1473
[5]   Use of metal chelate affinity chromatography and membrane-based ion-exchange as clean-up procedure for trace residue analysis of tetracyclines in animal tissues and egg [J].
Croubels, SM ;
Vanoosthuyze, KEI ;
VanPeteghem, CH .
JOURNAL OF CHROMATOGRAPHY B, 1997, 690 (1-2) :173-179
[6]  
De Wasch K, 1998, ANALYST, V123, P2737, DOI 10.1039/a804909b
[7]   COMPETITION BETWEEN TETRACYCLINE AND TRANSFER-RNA AT BOTH P-SITE AND A-SITE OF THE RIBOSOME OF ESCHERICHIA-COLI [J].
EPE, B ;
WOOLLEY, P ;
HORNIG, H .
FEBS LETTERS, 1987, 213 (02) :443-447
[8]   Electrochemical immunosensor with aptamer-based enzymatic amplification [J].
Feng, Kejun ;
Kang, Yan ;
Zhao, Jing-Jin ;
Liu, Ya-Li ;
Jiang, Jian-Hui ;
Shen, Guo-Li ;
Yu, Ru-Qin .
ANALYTICAL BIOCHEMISTRY, 2008, 378 (01) :38-42
[9]   Label-free electrochemical detection of nanomolar adenosine based on target-induced aptamer displacement [J].
Feng, Kejun ;
Sun, Chenhu ;
Kang, Yan ;
Chen, Jiwei ;
Jiang, Jian-Hui ;
Shen, Guo-Li ;
Yu, Ru-Qin .
ELECTROCHEMISTRY COMMUNICATIONS, 2008, 10 (04) :531-535
[10]   CAN TETRACYCLINE-INDUCED FATTY LIVER IN PREGNANCY BE ATTRIBUTED TO CHOLINE DEFICIENCY [J].
GWEE, MCE .
MEDICAL HYPOTHESES, 1982, 9 (02) :157-162