Non-invasive cardiac output monitor validation study in pediatric cardiac surgery patients

被引:12
作者
Tirotta, Christopher F. [1 ]
Lagueruela, Richard G. [1 ]
Madril, Danielle [1 ]
Velis, Evelio [2 ]
Ojito, Jorge [1 ]
Monroe, Daniel [1 ]
Aguero, David [1 ]
Irizarry, Marysory [1 ]
McBride, John [1 ]
Hannan, Robert L. [1 ]
Burke, Redmond P. [1 ]
机构
[1] Nicklaus Childrens Hosp, Heart Program, 3100 SW 62nd Ave, Miami, FL 33155 USA
[2] Barry Univ, Coll Hlth Sci, 11300 NE Second Ave,Off SNHS 229, Miami Shores, FL 33161 USA
关键词
Cardiac output; Non-invasive monitor; Cardiac surgery; Pediatrics; Congenital heart disease; Fick equation; Oxygen consumption; DIRECTED FLUID MANAGEMENT; CONGENITAL HEART-DISEASE; RISK SURGICAL-PATIENTS; ELECTRICAL VELOCIMETRY; TISSUE PERFUSION; FICK METHOD; THERMODILUTION; CHILDREN; OUTCOMES; THERAPY;
D O I
10.1016/j.jclinane.2017.02.001
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Study objective: To validate a non-invasive cardiac output monitor in pediatric cardiac surgery patients. Design: Prospective trial. Setting: Operating room. Patients: 20 pediatric cardiac surgery patients ASA physical status 3 and 4. Interventions: Aesculon noninvasive cardiac output monitor was used and compared to the cardiac output derived from the Fick equation. Oxygen consumption was measured and blood samples were taken from the arterial line and from the superior and inferior vena cava. Measurements: Noninvasive cardiac output and cardiac index and Fick cardiac output and cardiac index. Oxygen consumption was measured by the TreyMed metabolic monitor. Blood samples were simultaneously drawn from the arterial line and from the superior and inferior vena cava purse string sites by the surgeon, prior to commencing cardiopulmonary bypass. Another data set was obtained right after termination of cardiopulmonary bypass. Results: There was a direct, significant relationship between Fick CO/CI and NICOM CO/CI measurements. More dispersion is detected when the magnitude of the measure increases, i.e., for older and larger patients. Conclusions: There is a strong correlation between the cardiac output values derived from the Fick equation and the Aesculon non-invasive cardiac output monitor. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:129 / 132
页数:4
相关论文
共 30 条
[1]   Lack of agreement between thermodilution and fick cardiac output in critically ill patients [J].
Dhingra, VK ;
Fenwick, JC ;
Walley, KR ;
Chittock, DR ;
Ronco, JJ .
CHEST, 2002, 122 (03) :990-997
[2]  
Falder U, 2007, J THORAC CARDIOVASC, V133, P224
[3]   Goal-Directed Fluid Management Based on the Pulse Oximeter-Derived Pleth Variability Index Reduces Lactate Levels and Improves Fluid Management [J].
Forget, Patrice ;
Lois, Fernande ;
de Kock, Marc .
ANESTHESIA AND ANALGESIA, 2010, 111 (04) :910-914
[4]   Goal-directed intraoperative fluid administration reduces length of hospital stay after major surgery [J].
Gan, TJ ;
Soppitt, A ;
Maroof, M ;
El-Moalem, H ;
Robertson, KM ;
Moretti, E ;
Dwane, P ;
Glass, PSA .
ANESTHESIOLOGY, 2002, 97 (04) :820-826
[5]   Comparison of bedside measurement of cardiac output with the thermodilution method and the Fick method in mechanically ventilated patients [J].
Gonzalez, J ;
Delafosse, C ;
Fartoukh, M ;
Capderou, A ;
Straus, C ;
Zelter, M ;
Derenne, JP ;
Similowski, T .
CRITICAL CARE, 2003, 7 (02) :171-178
[6]   Maintaining Tissue Perfusion in High-Risk Surgical Patients: A Systematic Review of Randomized Clinical Trials [J].
Gurgel, Sanderland T. ;
do Nascimento, Paulo, Jr. .
ANESTHESIA AND ANALGESIA, 2011, 112 (06) :1384-1391
[7]   A Systematic Review and Meta-Analysis on the Use of Preemptive Hemodynamic Intervention to Improve Postoperative Outcomes in Moderate and High-Risk Surgical Patients [J].
Hamilton, Mark A. ;
Cecconi, Maurizio ;
Rhodes, Andrew .
ANESTHESIA AND ANALGESIA, 2011, 112 (06) :1392-1402
[8]   Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension [J].
Hoeper, MM ;
Maier, R ;
Tongers, J ;
Niedermeyer, J ;
Hohlfeld, JM ;
Hamm, M ;
Fabel, H .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 160 (02) :535-541
[9]   Noninvasive Cardiac Output Monitoring in Paediatric Cardiac Surgery: Correlation between Change in Thoracic Fluid Content and Change in Patient Body Weight [J].
Kang, W-S ;
Lee, J. H. ;
Shin, H. J. ;
Kim, S-H ;
Kim, T-Y ;
Seo, D. M. ;
Yoon, T-G .
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2012, 40 (06) :2295-2304
[10]   Advanced Hemodynamic Monitoring in Critically Ill Children [J].
Lemson, Joris ;
Nusmeier, Anneliese ;
van der Hoeven, Johannes G. .
PEDIATRICS, 2011, 128 (03) :560-571