On Synthetic Instrument Response Functions of Time-Correlated Single-Photon Counting Based Fluorescence Lifetime Imaging Analysis

被引:7
作者
Xiao, Dong [1 ]
Sapermsap, Natakorn [2 ]
Safar, Mohammed [1 ]
Cunningham, Margaret Rose [1 ]
Chen, Yu [2 ]
Li, David Day-Uei [1 ]
机构
[1] Univ Strathclyde, Strathclyde Inst Pharm & Biomed Sci, Glasgow, Lanark, Scotland
[2] Univ Strathclyde, Dept Phys, Glasgow, Lanark, Scotland
基金
英国生物技术与生命科学研究理事会; 英国工程与自然科学研究理事会;
关键词
time-resolved imaging; photon counting; deconvolution; fluorescence microscopy; instrument response function;
D O I
10.3389/fphy.2021.635645
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Time-correlated single-photon counting (TCSPC) has been the gold standard for fluorescence lifetime imaging (FLIM) techniques due to its high signal-to-noize ratio and high temporal resolution. The sensor system's temporal instrument response function (IRF) should be considered in the deconvolution procedure to extract the real fluorescence decay to compensate for the distortion on measured decays contributed by the system imperfections. However, to measure the instrument response function is not trivial, and the measurement setup is different from measuring the real fluorescence. On the other hand, automatic synthetic IRFs can be directly derived from the recorded decay profiles and provide appropriate accuracy. This paper proposed and examined a synthetic IRF strategy. Compared with traditional automatic synthetic IRFs, the new proposed automatic synthetic IRF shows a broader dynamic range and better accuracy. To evaluate its performance, we examined simulated data using nonlinear least square deconvolution based on both the Levenberg-Marquardt algorithm and the Laguerre expansion method for bi-exponential fluorescence decays. Furthermore, experimental FLIM data of cells were also analyzed using the proposed synthetic IRF. The results from both the simulated data and experimental FLIM data show that the proposed synthetic IRF has a better performance compared to traditional synthetic IRFs. Our work provides a faster and precise method to obtain IRF, which may find various FLIM-based applications. We also reported in which conditions a measured or a synthesized IRF can be applied.
引用
收藏
页数:11
相关论文
共 27 条
  • [1] Kilometer range depth imaging using time-correlated single-photon counting
    Buller, Gerald S.
    Krichel, Nils J.
    McCarthy, Aongus
    Gemmell, Nathan R.
    Tanner, Michael G.
    Natarajan, Chandra M.
    Ren, Ximing
    Hadfield, Robert H.
    INFRARED SENSORS, DEVICES, AND APPLICATIONS AND SINGLE PHOTON IMAGING II, 2011, 8155
  • [2] Ranging and three-dimensional imaging using time-correlated single-photon counting
    Buller, Gerald S.
    Wallace, Andrew M.
    McCarthy, Aongus
    Lamb, Robert A.
    ELECTRO-OPTICAL AND INFRARED SYSTEMS: TECHNOLOGY AND APPLICATIONS IV, 2007, 6737
  • [3] Time-of-flight sensing using time-correlated single-photon counting
    Buller, Gerald S.
    McCarthy, Aongus
    Collins, Robert J.
    Hiskett, Philip A.
    Fernandez, Veronica
    Hemandez-Marin, Sergio
    Wallace, Andrew M.
    QUANTUM SENSING AND NANOPHOTONIC DEVICES V, 2008, 6900
  • [4] Scanning of Low Signature Targets Using Time-Correlated Single-Photon Counting
    Buller, G. S.
    Collins, R. J.
    Krichel, N. J.
    Wallace, A. M.
    McCarthy, A.
    ADVANCED PHOTON COUNTING TECHNIQUES III, 2009, 7320
  • [5] Hybrid lidar system using combination of streak tube imaging and time-correlated single-photon counting
    Chen, Zhaodong
    Li, Xudong
    Zhou, Zhigang
    Ye, Guangchao
    Guan, Jiayu
    Luoi, Tong
    Fan, Rongwei
    Chen, Deying
    LASER PHYSICS, 2018, 28 (12)
  • [6] SPAD Smart Pixel for Time-of-Flight and Time-Correlated Single-Photon Counting Measurements
    Villa, F.
    Markovic, B.
    Bellisai, S.
    Bronzi, D.
    Tosi, A.
    Zappa, F.
    Tisa, S.
    Durini, D.
    Weyers, S.
    Paschen, U.
    Brockherde, W.
    IEEE PHOTONICS JOURNAL, 2012, 4 (03): : 795 - 804
  • [7] Comparative Study of Image Quality in Time-Correlated Single-Photon Counting Computed Tomography
    Gaudreault, David
    Rossignol, Julien
    Berube-Lauziere, Yves
    Fontaine, Rejean
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (03) : 343 - 349
  • [8] Depth imaging at kilometer range using time-correlated single-photon counting at wavelengths of 850 nm and 1560 nm
    Buller, Gerald S.
    McCarthy, Aongus
    Ren, Ximing
    Gemmell, Nathan R.
    Collins, Robert J.
    Krichel, Nils J.
    Tanner, Michael G.
    Wallace, Andrew M.
    Dorenbos, Sandor
    Zwiller, Val
    Hadfield, Robert H.
    BIOSENSING AND NANOMEDICINE V, 2012, 8460
  • [9] A correction method for range walk error in time-correlated single-photon counting using photomultiplier tube
    Chen, Zhaodong
    Li, Xudong
    Li, Xiaohui
    Ye, Guangchao
    Zhou, Zhigang
    Lu, Liutao
    Sun, Tongying
    Fan, Rongwei
    Chen, Deying
    OPTICS COMMUNICATIONS, 2019, 434 : 7 - 11
  • [10] The key technology of picosecond time-correlated single photon counting spectrometer
    Zhang, XF
    Song, KF
    Yu, T
    Han, BD
    Sheng, CX
    Lin, JL
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2004, 24 (12) : 1489 - 1493