Challenges of implementing Point-of-Care Testing (POCT) glucose meters in a pediatric acute care setting

被引:20
作者
Kavsak, PA
Zielinski, N
Li, DL
McNamara, PJ
Adeli, K
机构
[1] Univ Toronto, Hosp Sick Children, Div Clin Biochem, Toronto, ON M5G 1X8, Canada
[2] Univ Toronto, Hosp Sick Children, Div Neonatol, Toronto, ON M5G 1X8, Canada
关键词
glucose measurements; NICU; glucose meters; Point-of-Care Testing; POCT;
D O I
10.1016/j.clinbiochem.2004.06.003
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To investigate factors contributing to analytical bias in POCT glucose values generated by the NICU versus the core laboratory. Methods: The LifeScan Flexx hospital system glucose meters (SureStep) were used in precision and comparison studies between the NICU and laboratory (ABL715 and Vitros 950). Results: Analysis of 40 neonatal blood samples revealed a positive bias between the NICU glucose meters versus either the laboratory glucose meter or instrument (mean difference of 0.28 and 0.21 mmol/L, respectively). Linear regression analysis (R-2 = 0.0584) of the difference in glucose results versus time elapsed between measurements indicated that the bias observed between the NICU and laboratory glucose meters was not due to in vitro glycolysis for samples transported on ice. Further analysis indicated that the bias appeared to be mostly operator driven, with different NICU operators exhibiting different mean biases. Increasing the amount of blood applied to the SureStep Pro test strip (e.g., 60 vs. 20 muL), led to higher values for glucose concentration for the same blood. Nearly 50% of all glucose values reported for the NICU were obtained by the SureStep Flexx glucose meters in a 3-month period following the introduction of POCT, yet the number of laboratory-reported glucose results for the same period increased by 21% as compared to the previous year. Conclusions: Operator error appears to be a source of bias present between the NICU and laboratory, and despite glucose meter utilization in the NICU, the number of glucose measurements by the central laboratory increased after POCT introduction. (C) 2004 The Canadian Society of Clinical Chemists. All rights reserved.
引用
收藏
页码:811 / 817
页数:7
相关论文
共 50 条
[41]   Point-of-care testing on admission to the intensive care unit: lactate and glucose independently predict mortality [J].
Martin, Jan ;
Blobner, Manfred ;
Busch, Raymonde ;
Moser, Norman ;
Kochs, Eberhard ;
Luppa, Peter B. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (02) :405-412
[42]   My One Wish for Point-of-Care Testing Evidence-Based Regulatory Decisions Regarding Use of Glucose Meters in the Hospital [J].
Karon, Brad S. .
POINT OF CARE, 2016, 15 (02) :83-85
[43]   Point-of-care testing in critical care: the clinician's point of view [J].
Casagranda, Ivo .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (07) :931-934
[44]   Barriers to hospital-based clinical adoption of point-of-care testing (POCT): A systematic narrative review [J].
Quinn, Alistair D. ;
Dixon, Dorian ;
Meenan, Brian J. .
CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2016, 53 (01) :1-12
[45]   Point-of-Care Testing: General Aspects [J].
Ferreira, Carlos E. S. ;
Guerra, Joao C. C. ;
Slhessarenko, Natasha ;
Scartezini, Marileia ;
Franca, Carolina N. ;
Colombini, Marjorie P. ;
Berlitz, Fernando ;
Machado, Antonia M. O. ;
Campana, Gustavo A. ;
Faulhaber, Adriana C. L. ;
Galoro, Cesar A. ;
Dias, Claudia M. ;
Shcolnik, Wilson ;
Martino, Marines D. V. ;
Cesar, Katia R. ;
Sumita, Nairo M. ;
Mendes, Maria E. ;
Faulhaber, Marcelo H. W. ;
Pinho, Joao R. R. ;
Barbosa, Ismar V. ;
Batista, Marcelo C. ;
Khawali, Cristina ;
Pariz, Vitor M. ;
Andriolo, Adagmar .
CLINICAL LABORATORY, 2018, 64 (1-2) :1-9
[46]   Point-of-Care Testing for Pharyngitis in the Pharmacy [J].
Essack, Sabiha ;
Bell, John ;
Burgoyne, Douglas ;
Tongrod, Wirat ;
Duerden, Martin ;
Sessa, Aurelio ;
Altiner, Attila ;
Shephard, Adrian .
ANTIBIOTICS-BASEL, 2020, 9 (11) :1-16
[47]   Improving the quality of point-of-care testing [J].
Price, Christopher P. ;
Smith, Ian ;
Van den Bruel, Ann .
FAMILY PRACTICE, 2018, 35 (04) :358-364
[48]   Artificial Intelligence in Point-of-Care Testing [J].
Khan, Adil I. ;
Khan, Mazeeya ;
Khan, Raheeb .
ANNALS OF LABORATORY MEDICINE, 2023, 43 (05) :401-407
[49]   Point-of-care testing: A critical review [J].
Mor, M ;
Waisman, Y .
PEDIATRIC EMERGENCY CARE, 2000, 16 (01) :45-48
[50]   What Is New in Point-of-Care Testing? [J].
Melanson, Stacy E. F. .
POINT OF CARE, 2009, 8 (04) :166-170