On the influence of noise correlations in measurement data on basis image noise in dual-energylike x-ray imaging

被引:34
作者
Roessl, Ewald [1 ]
Ziegler, Andy [1 ]
Proksa, Roland [1 ]
机构
[1] Philips Res Labs, Sector Tech Syst, Hamburg, Germany
关键词
dual-energy imaging; counting-integrating readout; measurement noise correlations;
D O I
10.1118/1.2514058
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In conventional dual-energy systems, two transmission measurements with distinct spectral characteristics are performed. These measurements are used to obtain the line integrals of two basis material densities. Usually, the measurement process is such that the two measured signals can be treated as independent and therefore uncorrelated. Recently, however, a readout system for x-ray detectors has been introduced for which this is no longer the case. The readout electronics is designed to obtain simultaneous measurements of the total number of photons N and the total energy S they deposit in the sensor material. Practically, this is realized by a signal replication and separate counting and integrating processing units. Since the quantities Nand S are (electronically) derived from one and the same physical sensor signal, they are statistically correlated. Nevertheless, the pair M and S can be used to perform a dual-energy processing following the well-known approach by Alvarez and Macovski. Formally, this means that N is to be identified with the first dual-energy measurement M-1 and epsilon with the second measurement M-2. In the presence of input correlations between M-1 - N and M-2 = epsilon, however, the corresponding analytic expressions for the basis image noise have to be modified. The main observation made in this paper is that for positively correlated data, as is the case for the simultaneous counting and integrating device mentioned above, the basis image noise is suppressed through the influence of the covariance between the two signals. We extend the previously published relations for the basis image noise to the case where the original measurements are not independent and illustrate the importance of the input correlations by comparing dual-energy basis image noise resulting from the device mentioned above and a device measuring the photon numbers and the deposited energies consecutively. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:959 / 966
页数:8
相关论文
共 25 条
  • [1] Comparison of dual energy detector system performance
    Alvarez, RE
    Seibert, JA
    Thompson, SK
    [J]. MEDICAL PHYSICS, 2004, 31 (03) : 556 - 565
  • [2] Active energy selective image detector for dual-energy computed radiography
    Alvarez, RE
    [J]. MEDICAL PHYSICS, 1996, 23 (10) : 1739 - 1748
  • [3] ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY
    ALVAREZ, RE
    MACOVSKI, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) : 733 - 744
  • [4] BREAST IMAGING - DUAL-ENERGY PROJECTION RADIOGRAPHY WITH DIGITAL RADIOGRAPHY
    ASAGA, T
    CHIYASU, S
    MASTUDA, S
    MASTUURA, H
    KATO, H
    ISHIDA, M
    KOMAKI, T
    [J]. RADIOLOGY, 1987, 164 (03) : 869 - 870
  • [5] DETECTOR FOR DUAL-ENERGY DIGITAL RADIOGRAPHY
    BARNES, GT
    SONES, RA
    TESIC, MM
    MORGAN, DR
    SANDERS, JN
    [J]. RADIOLOGY, 1985, 156 (02) : 537 - 540
  • [6] DUAL-ENERGY PROJECTION RADIOGRAPHY - INITIAL CLINICAL-EXPERIENCE
    BRODY, WR
    CASSEL, DM
    SOMMER, FG
    LEHMANN, LA
    MACOVSKI, A
    ALVAREZ, RE
    PELC, NJ
    RIEDERER, SJ
    HALL, AL
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (02) : 201 - 205
  • [7] BROOKS RA, 1978, RADIOLOGY, V126, P255
  • [8] The design of a system for coloured digital radiology with VLSI circuits and GaAs pixel detectors
    Cadeddu, S
    Caligiore, C
    Caria, M
    Lai, A
    LoPresti, D
    Panebianco, S
    Petta, C
    Porcu, P
    Randazzo, N
    Randaccio, P
    Reito, S
    Russo, GV
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 419 (2-3) : 270 - 275
  • [9] COWAN G, 1998, STAT DATA ANAL, P21
  • [10] Ebel H, 1999, X-RAY SPECTROM, V28, P255, DOI 10.1002/(SICI)1097-4539(199907/08)28:4<255::AID-XRS347>3.0.CO