Internally cooled bipolar radiofrequency ablation:: Is a lower power output more effective?

被引:10
|
作者
Clasen, S.
Geng, A.
Herberts, T.
Boss, A.
Schmidt, D.
Schraml, C.
Fritz, J.
Kroeber, S. M.
Claussen, C. D.
Pereira, P. L.
机构
[1] Univ Klinikum Tubingen, Radiol Klin, D-72076 Tubingen, Germany
[2] Univ Tubingen, Inst Med Biometrie, D-72074 Tubingen, Germany
[3] Univ Tubingen, Inst Pathol, D-72074 Tubingen, Germany
关键词
radiofrequency ablation; abdomen; ablation procedures; technology assessment; liver; bipolar;
D O I
10.1055/s-2006-927360
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Evaluation of bipolar radiofrequency (RF) ablation using internally cooled electrodes in an ex-vivo experiment. Materials and Methods: Bipolar RF ablations (n = 154) were performed in ex-vivo bovine liver. Both electrodes with a total length of the active tip of 4 cm were located on the same shaft of an internally cooled applicator. The power output was systematically varied between 20 and 100 watts (W). The energy application was continuous or modulated depending on the tissue resistance. In relationship to the maximum power output, the volume of coagulation was assessed. Results: In continuous energy application the induced volume of coagulation was increased at lower power outputs up to 33.7 cm(3) (20 watts). Parallel to an increased volume of coagulation, the required duration of energy application was increased up to a maximum of 51.6 minutes. Modulation of the power output as a function of the tissue resistance enabled application of a wide range of power outputs (40-75 watts) leading to a comparable extent of coagulation with a maximum of 14.9 cm(3) (10min.), 16.8 cm(3) (15 min.),and 19.1 cm(3) (20 min.). Conclusion: Continuous application of RF energy leads to an inverse relationship between volume of coagulation and power output. Modulation of the power output as a function of the tissue resistance enables application of a wider range of power outputs compared to continuous application of RIF energy.
引用
收藏
页码:282 / 288
页数:7
相关论文
共 50 条
  • [41] Comparison of ablation performance between dual internally cooled wet tip and conventional dual internally cooled tip radiofrequency electrodes: an experimental study in ex vivo bovine liver
    Cha, Dong Ik
    Lee, Min Woo
    Jeong, Woo Kyoung
    Ha, Sang Yun
    Ahn, Soo Hyun
    Rhim, Hyunchul
    Lim, Hyo Keun
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2021, 38 (01) : 332 - 340
  • [42] Is the use of microwave ablation more effective and/or safe that radiofrequency ablation in the treatment of hepatocellular carcinoma?
    Anzai, Adriano
    Utino, Armelin
    Katayama, Haroldo
    Spir, Ighor A. Z.
    Lemos, Marcio A.
    Nery, Mary Martins
    Anhesini, Mauricio
    Tiezzi, Oswaldo S.
    Spir, Patricia R. N.
    Otani, Pericles
    Bernardo, Wanderley M.
    REVISTA DA ASSOCIACAO MEDICA BRASILEIRA, 2020, 66 (06): : 720 - 720
  • [43] Is the use of microwave ablation more effective and/or safe that radiofrequency ablation in the treatment of hepatocellular carcinoma?
    Anzai, Adriano
    Utino, Armelin
    Katayama, Haroldo
    Spir, Ighor A. Z.
    Lemos, Marcio A.
    Nery, Mary Martins
    Anhesini, Mauricio
    Tiezzi, Oswaldo S.
    Spir, Patricia R. N.
    Otani, Pericles
    Bernardo, Wanderley M.
    REVISTA DA ASSOCIACAO MEDICA BRASILEIRA, 2020, 66 (04): : 392 - 399
  • [44] Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling
    Clasen, S
    Schmidt, D
    Boss, A
    Dietz, K
    Kröber, SM
    Claussen, CD
    Pereira, PL
    RADIOLOGY, 2006, 238 (03) : 881 - 890
  • [45] Percutaneous Radiofrequency Ablation of Hepatocellular Carcinomas Adjacent to the Gallbladder with Internally Cooled Electrodes: Assessment of Safety and Therapeutic Efficacy
    Kim, Sang Won
    Rhim, Hyunchul
    Park, Mihyun
    Kim, Heejung
    Kim, Young-sun
    Choi, Dongil
    Lim, Hyo K.
    KOREAN JOURNAL OF RADIOLOGY, 2009, 10 (04) : 366 - 376
  • [46] Radiofrequency Ablation by a 21-Gauge Internally Cooled Electrode: Ex Vivo and In Vivo Evaluation by Rat Liver
    Haruyuki Takaki
    Yuki Kodama
    Hisao Miyamoto
    Yuto Iijima
    Yoshinari Kikuchi
    Hiroshi Kodama
    Eisuke Ueshima
    Keiji Nakasho
    Reona Wada
    Koichiro Yamakado
    CardioVascular and Interventional Radiology, 2021, 44 : 150 - 155
  • [47] Radiofrequency Ablation by a 21-Gauge Internally Cooled Electrode: Ex Vivo and In Vivo Evaluation by Rat Liver
    Takaki, Haruyuki
    Kodama, Yuki
    Miyamoto, Hisao
    Iijima, Yuto
    Kikuchi, Yoshinari
    Kodama, Hiroshi
    Ueshima, Eisuke
    Nakasho, Keiji
    Wada, Reona
    Yamakado, Koichiro
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2021, 44 (01) : 150 - 155
  • [48] Hepatocellular Carcinoma Radiofrequency Ablation: Which Imaging Technique is more effective?
    Fietz, Cornelia
    GESUNDHEITSOEKONOMIE UND QUALITAETSMANAGEMENT, 2019, 24 (04): : 163 - +
  • [49] A Comparison of Internally Water-perfused and Cryogenically Cooled Monopolar and Bipolar Radiofrequency Applicators in Ex Vivo Liver Samples
    Rempp, Hansjoerg
    Mezger, Dorothee
    Voigtlaender, Matthias
    Scharpf, Marcus
    Hoffmann, Ruediger
    Pereira, Philippe L.
    Enderle, Markus D.
    Claussen, Claus D.
    Clasen, Stephan
    ACADEMIC RADIOLOGY, 2014, 21 (05) : 661 - 666
  • [50] Comparison of microwave ablation and multipolar radiofrequency ablation, both using a pair of internally cooled interstitial applicators: Results in ex vivo porcine livers
    Li, Xin
    Zhang, Liang
    Fan, Weijun
    Zhao, Ming
    Wang, Ligang
    Tang, Tian
    Jiang, Hua
    Zhang, Jianlei
    Liu, Yanfeng
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2011, 27 (03) : 240 - 248