Internally cooled bipolar radiofrequency ablation:: Is a lower power output more effective?

被引:10
|
作者
Clasen, S.
Geng, A.
Herberts, T.
Boss, A.
Schmidt, D.
Schraml, C.
Fritz, J.
Kroeber, S. M.
Claussen, C. D.
Pereira, P. L.
机构
[1] Univ Klinikum Tubingen, Radiol Klin, D-72076 Tubingen, Germany
[2] Univ Tubingen, Inst Med Biometrie, D-72074 Tubingen, Germany
[3] Univ Tubingen, Inst Pathol, D-72074 Tubingen, Germany
关键词
radiofrequency ablation; abdomen; ablation procedures; technology assessment; liver; bipolar;
D O I
10.1055/s-2006-927360
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Evaluation of bipolar radiofrequency (RF) ablation using internally cooled electrodes in an ex-vivo experiment. Materials and Methods: Bipolar RF ablations (n = 154) were performed in ex-vivo bovine liver. Both electrodes with a total length of the active tip of 4 cm were located on the same shaft of an internally cooled applicator. The power output was systematically varied between 20 and 100 watts (W). The energy application was continuous or modulated depending on the tissue resistance. In relationship to the maximum power output, the volume of coagulation was assessed. Results: In continuous energy application the induced volume of coagulation was increased at lower power outputs up to 33.7 cm(3) (20 watts). Parallel to an increased volume of coagulation, the required duration of energy application was increased up to a maximum of 51.6 minutes. Modulation of the power output as a function of the tissue resistance enabled application of a wide range of power outputs (40-75 watts) leading to a comparable extent of coagulation with a maximum of 14.9 cm(3) (10min.), 16.8 cm(3) (15 min.),and 19.1 cm(3) (20 min.). Conclusion: Continuous application of RF energy leads to an inverse relationship between volume of coagulation and power output. Modulation of the power output as a function of the tissue resistance enables application of a wider range of power outputs compared to continuous application of RIF energy.
引用
收藏
页码:282 / 288
页数:7
相关论文
共 50 条
  • [21] Percutaneous radiofrequency ablation using internally cooled wet electrodes for the treatment of patients with lung tumors
    Iezzi, R.
    Larici, A. R.
    Posa, A.
    Carchesio, F.
    Congedo, M. T.
    Tagliaferri, L.
    Cassano, A.
    D'argento, E.
    Mantini, G.
    Rodolfino, E.
    Margaritora, S.
    Bria, E.
    Tortora, G.
    Valentini, V.
    Manfredi, R.
    EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 2019, 23 (15) : 6554 - 6561
  • [22] Porcine liver: Morphologic characteristics and cell viability at experimental radiofrequency ablation with internally cooled electrodes
    Ng, KK
    Lam, CM
    Poon, RT
    Shek, TW
    Yu, WC
    To, JY
    Wo, YH
    Lau, CP
    Tang, TC
    Ho, DW
    Fan, ST
    RADIOLOGY, 2005, 235 (02) : 478 - 486
  • [23] Microwave Ablation versus Radiofrequency Ablation in the Kidney: High-power Triaxial Antennas Create Larger Ablation Zones than Similarly Sized Internally Cooled Electrodes
    Laeseke, Paul F.
    Lee, Fred T., Jr.
    Sampson, Lisa A.
    van der Weide, Daniel W.
    Brace, Christopher L.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (09) : 1224 - 1229
  • [24] Radiofrequency ablation in pig lungs:: in vivo comparison of internally cooled, perfusion and multitined expandable electrodes
    Lee, J. M.
    Han, J. K.
    Chang, J. M.
    Chung, S. Y.
    Kim, S. H.
    Lee, J. Y.
    Choi, B. I.
    BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (943): : 562 - 571
  • [25] Percutaneous radiofrequency ablation using internally cooled wet electrodes for treatment of colorectal liver metastases
    Yoon, H. M.
    Kim, J. H.
    Shin, Y. M.
    Won, H. J.
    Kim, P. N.
    CLINICAL RADIOLOGY, 2012, 67 (02) : 122 - 127
  • [26] A clinically-oriented computer model for radiofrequency hepatic ablation with internally cooled wet electrode
    Ewertowska, E.
    Quesada, R.
    Radosevic, A.
    Andaluz, A.
    Moll, X.
    Berjano, E.
    Burdio, F.
    Trujillo, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2018, 194 (05) : 479 - 479
  • [27] A clinically oriented computer model for radiofrequency ablation of hepatic tissue with internally cooled wet electrode
    Ewertowska, E.
    Quesada, R.
    Radosevic, A.
    Andaluz, A.
    Moll, X.
    Garcia Arnas, F.
    Berjano, E.
    Burdio, F.
    Trujillo, M.
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2019, 35 (01) : 194 - 204
  • [28] Image-guided percutaneous radiofrequency ablation of unresectable spinal metastases with an internally cooled electrode system
    Gevargez, A
    Braun, M
    Schirp, S
    Groenemeyer, DH
    RADIOLOGY, 2002, 225 : 614 - 614
  • [29] Radiofrequency ablation with internally cooled versus perfused electrodes for the treatment of small hepatocellular carcinoma in patients with cirrhosis
    Seror, Olivier
    N'Kontchou, Gisele
    Tin-Tin-Htar, Myint
    Barrucand, Corinne
    Ganne, Nathalie
    Coderc, Emmanuelle
    Trinchet, Jean Claude
    Sellier, Nicolas
    Beaugrand, Michel
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (05) : 718 - 724
  • [30] Radiofrequency ablation in the liver using two cooled-wet electrodes in the bipolar mode
    Han, JK
    Lee, JM
    Kim, SH
    Lee, JY
    Park, HS
    Eo, H
    Choi, BI
    EUROPEAN RADIOLOGY, 2005, 15 (10) : 2163 - 2170