Internally cooled bipolar radiofrequency ablation:: Is a lower power output more effective?

被引:10
|
作者
Clasen, S.
Geng, A.
Herberts, T.
Boss, A.
Schmidt, D.
Schraml, C.
Fritz, J.
Kroeber, S. M.
Claussen, C. D.
Pereira, P. L.
机构
[1] Univ Klinikum Tubingen, Radiol Klin, D-72076 Tubingen, Germany
[2] Univ Tubingen, Inst Med Biometrie, D-72074 Tubingen, Germany
[3] Univ Tubingen, Inst Pathol, D-72074 Tubingen, Germany
关键词
radiofrequency ablation; abdomen; ablation procedures; technology assessment; liver; bipolar;
D O I
10.1055/s-2006-927360
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Evaluation of bipolar radiofrequency (RF) ablation using internally cooled electrodes in an ex-vivo experiment. Materials and Methods: Bipolar RF ablations (n = 154) were performed in ex-vivo bovine liver. Both electrodes with a total length of the active tip of 4 cm were located on the same shaft of an internally cooled applicator. The power output was systematically varied between 20 and 100 watts (W). The energy application was continuous or modulated depending on the tissue resistance. In relationship to the maximum power output, the volume of coagulation was assessed. Results: In continuous energy application the induced volume of coagulation was increased at lower power outputs up to 33.7 cm(3) (20 watts). Parallel to an increased volume of coagulation, the required duration of energy application was increased up to a maximum of 51.6 minutes. Modulation of the power output as a function of the tissue resistance enabled application of a wide range of power outputs (40-75 watts) leading to a comparable extent of coagulation with a maximum of 14.9 cm(3) (10min.), 16.8 cm(3) (15 min.),and 19.1 cm(3) (20 min.). Conclusion: Continuous application of RF energy leads to an inverse relationship between volume of coagulation and power output. Modulation of the power output as a function of the tissue resistance enables application of a wider range of power outputs compared to continuous application of RIF energy.
引用
收藏
页码:282 / 288
页数:7
相关论文
共 50 条
  • [11] Hepatic radiofrequency ablation with internally cooled probes: Effect of coolant temperature on lesion size
    Haemmerich, D
    Chachati, L
    Wright, AS
    Mahvi, DM
    Lee, FT
    Webster, JG
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2003, 50 (04) : 493 - 500
  • [12] Radiofrequency ablation of subcapsular lesions with internally cooled perfusion electrode: ex vivo study
    Kim, Myung Sub
    Hong, Hyun Pyo
    Kang, Kyung A.
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2022, 31 (02) : 216 - 222
  • [13] Radiofrequency ablation of the porcine liver in vivo: Increased coagulation with an internally cooled perfusion electrode
    Lee, JM
    Han, JK
    Chang, JM
    Chung, SY
    Kim, SH
    Lee, JY
    Lee, MW
    Choi, BI
    ACADEMIC RADIOLOGY, 2006, 13 (03) : 343 - 352
  • [14] Safety and efficacy of radiofrequency ablation with internally cooled electrode for perivascular hepatic malignancy.
    Wang, Liping
    Chen, Liyu
    Xu, Dong
    Yang, Chen
    Shi, Kaiyuan
    Qian, Chaowen
    Liu, Junping
    BIOMEDICAL RESEARCH-INDIA, 2015, 26 (03): : 485 - 492
  • [15] Bipolar Radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver -: Prediction of coagulation volume from applied energy
    Clasen, Stephan
    Schmidt, Diethard
    Dietz, Klaus
    Boss, Andreas
    Kroeber, Stefan M.
    Schraml, Christina
    Fritz, Jan
    Claussen, Claus D.
    Pereira, Philippe L.
    INVESTIGATIVE RADIOLOGY, 2007, 42 (01) : 29 - 36
  • [16] Radiofrequency Ablation Zones in Ex Vivo Bovine and In Vivo Porcine Livers: Comparison of the Use of Internally Cooled Electrodes and Internally Cooled Wet Electrodes
    Cha, Jihoon
    Choi, Dongil
    Lee, Min Woo
    Rhim, Hyunchul
    Kim, Young-sun
    Lim, Hyo K.
    Yoon, Jung Hwan
    Park, Cheol Keun
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 32 (06) : 1235 - 1240
  • [17] Radiofrequency Ablation Zones in Ex Vivo Bovine and In Vivo Porcine Livers: Comparison of the Use of Internally Cooled Electrodes and Internally Cooled Wet Electrodes
    Jihoon Cha
    Dongil Choi
    Min Woo Lee
    Hyunchul Rhim
    Young-sun Kim
    Hyo K. Lim
    Jung Hwan Yoon
    Cheol Keun Park
    CardioVascular and Interventional Radiology, 2009, 32 : 1235 - 1240
  • [18] Internally gas-cooled radiofrequency applicators as an alternative to conventional radiofrequency and microwave ablation devices: An in vivo comparison
    Rempp, Hansjoeg
    Voigtlaender, Matthias
    Schenk, Martin
    Enderle, Markus D.
    Scharpf, Marcus
    Greiner, Tim O.
    Neugebauer, Alexander
    Hoffmann, Ruediger
    Claussen, Claus D.
    Clasen, Stephan
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (08) : E350 - E355
  • [19] Finite Element Study of an Internally Cooled Bipolar Electrode for Biological Tissue Coagulation with Radiofrequency Currents
    Gonzalez Suarez, A.
    Trujillo, M.
    Berjano, E. J.
    5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 13 - 16
  • [20] Radiofrequency ablation for small hepatocellular carcinoma: Prospective comparison of internally cooled electrode and expandable electrode
    Shibata, T
    Shibata, T
    Maetani, Y
    Isoda, H
    Hiraoka, M
    RADIOLOGY, 2006, 238 (01) : 346 - 353