The low-noise read-out electronics of the 187Re experiment

被引:3
作者
Gatti, F
Parodi, L
机构
[1] Univ Genoa, Dipartimento Fis, I-16146 Genoa, Italy
[2] Ist Nazl Fis Nucl, I-16146 Genoa, Italy
关键词
D O I
10.1016/S0168-9002(99)01344-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The low-noise read-out electronics for the rhenium experiment is presented. We have built three source followers with two matched J-FET in cascode configuration. These are cooled at about 150 K in a thermally shielded box in the inner vacuum chamber of the dilution cryostat. The output is forwarded to a noise-matched amplifier at room temperature. In the operating conditions, the overall voltage noise density ranges between 0.8 and 1.1 nV/Hz(-1/2) at 1 kHz. The 1/f noise contributes with an excess noise lower than 3 nV/Hz(-1/2) at 1 Hz. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:129 / 131
页数:3
相关论文
共 1 条
[1]   First results of the calorimetric spectrometer for the beta decay of rhenium-187 [J].
Gatti, F ;
Fontanelli, F ;
Galeazzi, M ;
Vitale, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) :88-91