Ultrasonic material characterization using large-aperture PVDF receivers

被引:17
作者
Adamowski, J. C. [2 ]
Buiochi, F. [2 ]
Higuti, R. T. [1 ]
机构
[1] Univ Estadual Paulista, Dep Elect Eng, BR-15385000 Ilha Solteira, SP, Brazil
[2] Univ Sao Paulo, Escola Politecn, BR-05508900 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
Density; Elastic constants; Large-aperture receiver; PVDF; Material characterization; ELASTIC-CONSTANTS; DENSITY; TRANSMISSION; LIQUIDS; PHASE;
D O I
10.1016/j.ultras.2009.09.018
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This work describes the use of a large-aperture PVDF receiver in the measurement of liquid density and composite material elastic constants. The density measurement of several liquids is obtained with accuracy of 0.2% using a conventional NDE emitter transducer and a 70-mm-diameter, 52-mu m P(VDF-TrFE) membrane with gold electrodes. The determination of the elastic constants is based on the phase velocity measurement. Diffraction can lead to errors around 1% in velocity measurement when using alternatively the conventional pair of ultrasonic transducers (1-MHz frequency and 19-mm-diameter) operating in through-transmission mode, separated by a distance of 100 mm. This effect is negligible when using a pair of 10-MHz, 19-mm-diameter transducers. Nevertheless, the dispersion at 10 MHz can result in errors of about 0.5%, when measuring the velocity in composite materials. The use of an 80-mm diameter, 52-mu m-thick PVDF membrane receiver practically eliminates the diffraction effects in phase velocity measurement. The elastic constants of a carbon fiber reinforced polymer were determined and compared with the values obtained by a tensile test. (C) 2009 Elsevier B. V. All rights reserved.
引用
收藏
页码:110 / 115
页数:6
相关论文
共 18 条
[1]   ULTRASONIC MEASUREMENT OF DENSITY OF LIQUIDS [J].
ADAMOWSKI, JC ;
BUIOCHI, F ;
SIMON, C ;
SILVA, ECN ;
SIGELMANN, RA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 97 (01) :354-361
[2]  
ADAMOWSKI JC, 2008, P IEEE INT ULTR S, P1524, DOI DOI 10.1109/ULTSYM.2008.0371
[3]  
ADAMOWSKI JC, 2007, P INT C ULTR VIENN, P4
[4]  
AULD BA, 1990, ACOUSTICS FIELDS WAV, V1
[5]  
BJORNDAL E, 2008, IEEE T ULTRASON FERR, V55, P675
[6]  
Costa ET, 1989, THESIS U LONDON
[7]   SPEED OF SOUND IN PURE WATER [J].
DELGROSS.VA ;
MADER, CW .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1972, 52 (05) :1442-&
[8]   TRANSMISSION AND RECEPTION OF SHORT ULTRASONIC PULSES BY CIRCULAR AND SQUARE TRANSDUCERS [J].
HAYMAN, AJ ;
WEIGHT, JP .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1979, 66 (04) :945-951
[9]   Thermal characterization of an ultrasonic density-measurement cell [J].
Higuti, Ricardo Tokio ;
Galindo, Bruno Shintate ;
Kitano, Claudio ;
Buiochi, Flavio ;
Adamowski, Julio Cezar .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (03) :924-930
[10]   Ultrasonic densitometer using a multiple reflection technique [J].
Higuti, RT ;
Adamowski, JC .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (09) :1260-1268