Disparity between adjusted and actual opening cerebrospinal fluid pressure in a patient with the Codman Hakim programmable valve: occult form of shunt failure due to head banging - Case report

被引:8
作者
Sato, Kimitoshi [1 ]
Smmizu, Satoru [1 ]
Utsuki, Satosth [1 ]
Suzuki, Sachio [1 ]
Oka, Hidehiro [1 ]
Fujii, Kiyotaka [1 ]
机构
[1] Kitasato Univ, Sch Med, Dept Neurosurg, Sagamihara, Kanagawa 2288555, Japan
关键词
opening pressure; programmable valve; shunt failure; pediatric neurosurgery;
D O I
10.3171/ped.2006.105.5.425
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Although the Codman Hakim programmable valve (CHPV) system is popular, several problems arising from its design have been described. The authors report an additional cause of shunt failure associated with the use of the CHPV system. A 7-year-old girl who had previously undergone placement of a ventriculoperitoneal shunt with the CHPV system was admitted in a comatose state. Radiography did not demonstrate any breakage or change in the opening pressure of the valve. Manual pumping of the valve resulted in neurological improvement and shrinkage of the girl's ventricles. However, ventriculornegaly was again observed several hours later, and an emergency valve revision was per-formed. The manufacturer's examination of the extracted valve revealed a crack on the surface of the hard plastic housing covering the valve chamber, which appeared to be suggestive of some type of blunt trauma. In addition, the flat spring that transmitted resistance from the pressure control cam to the valve ball was deformed in such a manner that it caused excessive pressure against the valve ball. Because of this deformity, the actual opening pressure in this patient had reached a maximum level of 226 mm H2O instead of the adjusted pressure of 60 mm H2O. In patients with the CHPV system whose medical history is difficult to ascertain, the possibility of damage to the valve spring and the need for a prompt revision surgery should be considered. The authors conclude that placement of the CHPV system may be contraindicated in patients with a known habit of head banging.
引用
收藏
页码:425 / 427
页数:3
相关论文
共 12 条
[1]  
Belliard H, 1996, NEUROCHIRURGIE, V42, P139
[2]   THE USE OF THE CODMAN-MEDOS PROGRAMMABLE HAKIM VALVE IN THE MANAGEMENT OF PATIENTS WITH HYDROCEPHALUS - ILLUSTRATIVE CASES [J].
BLACK, PM ;
HAKIM, R ;
BAILEY, NO .
NEUROSURGERY, 1994, 34 (06) :1110-1113
[3]   Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus [J].
Drake, JM ;
Kestle, JRW ;
Milner, R ;
Cinalli, G ;
Boop, F ;
Piatt, J ;
Haines, S ;
Schiff, SJ ;
Cochrane, DD ;
Steinbok, P ;
MacNeil, N ;
Haines, S ;
Sainte-Rose, C ;
Poskitt, K ;
Goumnerova, L ;
Albright, AL ;
Cinalli, G ;
Pierre-Kahn, A ;
Renier, D ;
Zerah, M ;
Rutka, J ;
Humphreys, R ;
Hoffman, H ;
Lamberti-Pasculi, M ;
Teo, C ;
Cherny, B ;
Aureli, S ;
Vandertop, P ;
Broomstra, S ;
Chadduck, W ;
Donahue, D ;
Hall, W ;
Parent, A ;
Turmel, A ;
Myles, T ;
Hamilton, M ;
Oakes, J ;
Mapstone, T .
NEUROSURGERY, 1998, 43 (02) :294-303
[4]  
HAKIM C, 1985, PHYS PATHOPHYSIOLOGY
[5]  
HAKIM S, 1973, DEV MED CHILD NEUROL, V15, P230
[6]   A rare case of shunt malfunction attributable to blockage of a Codman-Hakim programmable shunt valve [J].
Kurosaki, K ;
Hamada, H ;
Hayashi, N ;
Kurimoto, M ;
Hirashima, Y ;
Endo, S .
CHILDS NERVOUS SYSTEM, 2002, 18 (3-4) :183-185
[7]   Shunt malfunction caused by mischievous twisting of the valve by the patient: Case report [J].
Nabika, S ;
Oki, S ;
Sumida, M ;
Isobe, N .
NEUROSURGERY, 2003, 53 (01) :238-239
[8]   A rare case of shunt malfunction attributable to a broken Codman-Hakim programmable shunt valve after a blow to the head [J].
Okazaki, T ;
Oki, S ;
Migita, K ;
Kurisu, K .
PEDIATRIC NEUROSURGERY, 2005, 41 (05) :241-243
[9]   The Medos Hakim programmable valve in the treatment of pediatric hydrocephalus [J].
Di Rocco, C .
CHILDS NERVOUS SYSTEM, 1997, 13 (11-12) :594-594
[10]   Obstruction of a Codman-Hakim programmable valve by a migrating pressure control cam - Case report [J].
Shimizu, S ;
Utsuki, S ;
Suzuki, S ;
Oka, H ;
Fujii, K .
JOURNAL OF NEUROSURGERY, 2005, 103 (03) :270-271