Application of a 3D volumetric display for radiation therapy treatment planning I: quality assurance procedures

被引:2
作者
Gong, Xing [1 ]
Kirk, Mike [1 ]
Zusag, Tom [1 ]
Khelashvili, Gocha [1 ]
Chu, James [1 ]
Napoli, Josh [2 ]
Stutsman, Sandy [2 ]
机构
[1] Rush Univ, Med Ctr, Dept Radiat Oncol, Chicago, IL 60612 USA
[2] Actual Med Syst, Bedford, MA 01730 USA
关键词
Three-dimensional displays; treatment planning; quality assurance;
D O I
10.1120/jacmp.v10i3.2900
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To design and implement a set of quality assurance tests for an innovative 3D volumetric display for radiation treatment planning applications. A genuine 3D display (Perspecta Spatial 3D, Actuality-Systems Inc., Bedford, MA) has been integrated with the Pinnacle TPS (Philips Medical Systems, Madison WI), for treatment planning. The Perspecta 3D display renders a 25 cm diameter volume that is viewable from any side, floating within a translucent dome. In addition to displaying all 3D data exported from Pinnacle, the system provides a 3D mouse to define beam angles and apertures and to measure distance. The focus of this work is the design and implementation of a quality assurance program for 3D displays and specific 3D planning issues as guided by AAPM Task Group Report 53. A series of acceptance and quality assurance tests have been designed to evaluate the accuracy of CT images, contours, beams, and dose distributions as displayed on Perspecta. Three-dimensional matrices, rulers and phantoms with known spatial dimensions were used to check Perspecta's absolute spatial accuracy. In addition, a system of tests was designed to confirm Perspecta's ability to import and display Pinnacle data consistently. CT scans of phantoms were used to confirm beam field size, divergence, and gantry and couch angular accuracy as displayed on Perspecta. Beam angles were verified through Cartesian coordinate system measurements and by CT scans of phantoms rotated at known angles. Beams designed on Perspecta were exported to Pinnacle and checked for accuracy. Dose at sampled points were checked for consistency with Pinnacle and agreed within 1% or 1 mm. All data exported from Pinnacle to Perspecta was displayed consistently. The 3D spatial display of images, contours, and dose distributions were consistent with Pinnacle display. When measured by the 3D ruler, the distances between any two points calculated using Perspecta agreed with Pinnacle within the measurement error
引用
收藏
页码:96 / 114
页数:19
相关论文
共 21 条
[1]   Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems [J].
Able, Charles M. ;
Thomas, Michael D. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (03) :13-22
[2]   Holographic image guided radiation therapy (HIGRT) treatment planning: a multi-institutional study [J].
Chu, J. C. H. ;
Gong, X. ;
Kirk, M. ;
Khan, A. ;
Rivard, M. ;
Melhus, C. ;
Buscher, M. ;
Cardarelli, G. ;
Hurley, A. ;
Hepel, J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03) :S664-S665
[3]  
CHU JCH, 2007, ASTRO 2007 IN PRESS
[4]   A quality assurance phantom for three-dimensional radiation treatment planning [J].
Craig, T ;
Brochu, D ;
Van Dyk, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 44 (04) :955-966
[5]   100 Million-voxel volumetric display [J].
Favalora, GE ;
Napoli, J ;
Hall, DM ;
Dorval, RK ;
Giovinco, MG ;
Richmond, MJ ;
Chun, WS .
COCKPIT DISPLAYS IX: DISPLAYS FOR DEFENSE APPLICATIONS, 2002, 4712 :300-312
[6]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[7]  
HALLER M, 1999, EUROGRAPHICS 99
[8]   Stereo display of nested 3D volume data using automatic tunnelling [J].
Hubbold, RJ ;
Hancock, DJ .
STEREOSCOPIC DISPLAYS AND VIRTUAL REALITY SYSTEMS VI, 1999, 3639 :200-207
[9]  
HUBBOLD RJ, 1997, P EL IM 97 S C STER, V3012
[10]   Quality assurance of a virtual simulation software:: application to IMAgo and SIMAgo (ISOgray™) [J].
Isambert, A. ;
Beaudre, A. ;
Ferreira, I. ;
Lefkopoulos, D. .
CANCER RADIOTHERAPIE, 2007, 11 (04) :178-187