Self-Fitting Algorithm for Digital Hearing Aid Based on Interactive Evolutionary Computation and Expert System

被引:14
作者
Liang, Ruiyu [1 ,2 ]
Guo, Ruxue [1 ]
Xi, Ji [1 ,3 ]
Xie, Yue [1 ]
Zhao, Li [1 ]
机构
[1] Southeast Univ, Sch Informat Sci & Engn, Nanjing 210096, Jiangsu, Peoples R China
[2] Nanjing Inst Technol, Sch Commun Engn, Nanjing 211167, Jiangsu, Peoples R China
[3] Changzhou Inst Technol, Sch Comp Informat & Engn, Changzhou 213002, Peoples R China
来源
APPLIED SCIENCES-BASEL | 2017年 / 7卷 / 03期
基金
中国国家自然科学基金; 中国博士后科学基金;
关键词
hearing aid; self-fitting; interactive evolutionary computation; expert system; AUDIOLOGY; LOUDNESS; DESIGN; NEED;
D O I
10.3390/app7030272
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The traditional hearing aid fitting method, which mainly relies on the audiologist, is time-consuming and messy. To improve this situation, a self-fitting algorithm based on an improved interactive evolutionary computation (IEC) algorithm and expert system, which enables the patients to fit the hearing aid by themselves, is proposed. The algorithm takes the band gain as the fitting target and uses the patient's subjective evaluation to iteratively update the algorithm parameters based on the improved IEC algorithm. In addition, a real-time updated expert system is constructed to assist in the optimization of the initial and iterative parameters of the fitting based on the patient's audiogram and personal information. To verify the performance of the algorithm, a self-fitting software for the hearing aid is designed. Through this software, the test signal is generated for the patient to evaluate the audio quality on a five-level scale. Based on the evaluation results, the algorithm iteratively optimizes the algorithm parameters until the patient is satisfied with the generated audio. Compared with the fitting algorithm based on Gaussian processes algorithm or the interactive evolutionary algorithm, the average subjective speech recognition rate of the proposed algorithm increase at least 11%. The average recognition rate for environmental sound is also improved by at least 2.9%. In addition, the fitting time of the proposed algorithm is shortened by at least 10 min compared to others two algorithms.
引用
收藏
页数:19
相关论文
共 40 条
  • [1] An Introduction to the Second Starkey Research Summit
    Abrams, Harvey B.
    [J]. AMERICAN JOURNAL OF AUDIOLOGY, 2012, 21 (02) : 329 - 330
  • [2] [Anonymous], 2001, J Am Acad Audiol, DOI [DOI 10.1055/S-0041-1741117, DOI 10.1055/s-0041-1741117]
  • [3] Bench J, 1979, Br J Audiol, V13, P108, DOI 10.3109/03005367909078884
  • [4] Audiologist-Driven Versus Patient-Driven Fine Tuning of Hearing Instruments
    Boymans, Monique
    Dreschler, Wouter A.
    [J]. TRENDS IN AMPLIFICATION, 2012, 16 (01): : 49 - 58
  • [5] Cai Liang-wei, 2006, Systems Engineering and Electronics, V28, P925
  • [6] Enabling Query Technologies for the Semantic Sensor Web
    Calbimonte, Jean-Paul
    Jeung, Hoyoung
    Corcho, Oscar
    Aberer, Karl
    [J]. INTERNATIONAL JOURNAL ON SEMANTIC WEB AND INFORMATION SYSTEMS, 2012, 8 (01) : 43 - 63
  • [7] Assembly and Insertion of a Self-Fitting Hearing Aid: Design of Effective Instruction Materials
    Caposecco, Andrea
    Hickson, Louise
    Meyer, Carly
    [J]. TRENDS IN AMPLIFICATION, 2011, 15 (04): : 184 - 195
  • [8] A Self-Fitting Hearing Aid: Need and Concept
    Convery, Elizabeth
    Keidser, Gitte
    Dillon, Harvey
    Hartley, Lisa
    [J]. TRENDS IN AMPLIFICATION, 2011, 15 (04): : 157 - 166
  • [9] Perception of a Self-Fitting Hearing Aid Among Urban-Dwelling Hearing-Impaired Adults in a Developed Country
    Convery, Elizabeth
    Keidser, Gitte
    Hartley, Lisa
    [J]. TRENDS IN AMPLIFICATION, 2011, 15 (04): : 175 - 183
  • [10] Verification of In Situ Thresholds and Integrated Real-Ear Measurements
    DiGiovanni, Jeffrey J.
    Pratt, Ryan M.
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2010, 21 (10) : 663 - 670