Proportional counting of tritium gas generated by polymer electrolyte membrane (PEM) electrolysis

被引:9
作者
Soreefan, Aurelie M. [1 ]
DeVol, Timothy A. [1 ]
机构
[1] Clemson Univ, Environm Engn & Earth Sci Dept, Clemson, SC 29634 USA
关键词
Proton exchange membrane; Electrolytic reduction; Nafion;
D O I
10.1007/s10967-009-0247-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The combination of an Ir/Pt PEM electrolyzer with a 1 L flow-through gas proportional counter was characterized for the quantification of tritium in water. The goal of the detection system is to quantify at concentrations below the Environmental Protection Agency (EPA) primary drinking water standard (740 Bq/L) with minimal expendables. The detector operating voltage, efficiency, background count rate of the passively shielded counter were measured in order to calculate the minimum detectable concentration of the detection system. The electrolyzer fractionation factor beta (e) value deduced from the measurement of gas phase activity concentrations generated from tritium aqueous standards was found in good agreement with literature values.
引用
收藏
页码:517 / 521
页数:5
相关论文
共 7 条
[1]  
Cameron J.F., 1967, RADIOACTIVE DATING M, P543
[2]  
IVANCHUK OM, 2000, ATOM ENERGY, V89, P3
[3]  
*NAT RES COUNC, 2000, US BOARD RAD WAST MA
[4]   Tritium separation by electrolysis using solid polymer electrolyte [J].
Ogata, Y ;
Sakuma, Y ;
Ohtani, N ;
Kotaka, M .
FUSION SCIENCE AND TECHNOLOGY, 2005, 48 (01) :136-139
[5]  
Ostlund H.G., 1962, TRITIUM PHYS BIOLOGI, P95
[6]  
SOREEFAN AM, 2009, THESIS CLEMSON U
[7]   Determination of tritium enrichment parameters of a commercially available PEM electrolyzer: a comparison with conventional enrichment electrolysis [J].
Soreefan, Aurelie M. ;
DeVol, Timothy A. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 282 (02) :511-515