On the elimination of the afterglow of CsI(Tl) scintillation detector

被引:9
作者
Alfassi, Z. B. [1 ]
Ifergan, Y.
Wengrowicz, U. [2 ]
Weinstein, Marcelo [2 ]
机构
[1] Ben Gurion Univ Negev, Dept Nucl Engn, IL-84105 Beer Sheva, Israel
[2] Nucl Res Ctr Negev, IL-84190 Beer Sheva, Israel
关键词
Scintillators; Afterglow; Temperature effect; CSI-TL; SUPPRESSION; EU2+;
D O I
10.1016/j.nima.2009.05.002
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The use of CsI(Tl) in survey meters is limited to low dose-rate fields due to the afterglow effect, which can take up to 20 min in case of 120 s exposure to 4 R/h field. Even mixtures of low and high dose-rate fields are prohibited since the afterglow effect of the high dose-rate field prevents the use of the meter in the low dose-rate field. It was found that heating the scintillation detector to 60 degrees C shortened considerably the time of the afterglow effect to 3 s, even after exposure of 120 s to fields of 4 R/h. It is suggested to start heating the detector immediately when reaching high dose-rate field and adjusting the calibration to the measured temperature. (C) 2009 Elsevier B.V. All rights reserved.