Image quality and exposure dose in digital projection radiography

被引:28
作者
Busch, HP [1 ]
Busch, S [1 ]
Decker, C [1 ]
Schilz, C [1 ]
机构
[1] Abt Radiol, Krankenhaus Barmherzigen Bruder, D-54292 Trier, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2003年 / 175卷 / 01期
关键词
flat plate detector radiography; storage phosphor radiography; dose; image quality; digital radiography;
D O I
10.1055/s-2003-36595
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Comparison of the imaging capabilities of storage phosphor (computed) radiography and flat plate radiography with conventional film-screen radiography to find new strategies for quality and dose management, i.e., optimizing imaging quality and dose depending on the imaging method and clinical situation. Materials and Methods: Images of a CDRAD-phantom, hand-phantom, abdomen-phantom and chest-phantom obtained with different exposure voltages (50 kV, 73 kV, 109 W) and different speeds (200, 400, 800, 1600) were processed with various digital systems (flat plate detector: Digital Diagnost [Philips]; storage phosphors: ADC-70 [Agfa], ADC-Solo [Agfa], FCRXG1 [Fuji]) and a conventional film-screen system (HT100G/Ortho Regular [Agfa]). Results: The evaluation of CDRAD images found the flat plate detector system to have the highest contrast detectability for all dose levels, followed by the FCRXG1, ADC-Solo and ADC-70 systems. Comparison of the organ-phantom images found the flat plate detector system to be equal to film-screen radiography and especially to storage phosphor systems even for low exposure doses. Conclusions: Flat plate radiography systems demonstrate the highest potential for high image quality when reducing the exposure dose. Depending on the system generation, the storage phosphor systems also show an improved image quality, but the possibility of a dose reduction is limited in comparison with the flat plate detector system.
引用
收藏
页码:32 / 37
页数:6
相关论文
共 19 条
[1]   Comparison of low contrast detectability between a digital amorphous silicon and a screen-film based imaging system for thoracic radiography [J].
Aufrichtig, R .
MEDICAL PHYSICS, 1999, 26 (07) :1349-1358
[2]  
Busch HP, 1999, RADIOLOGE, V39, P710, DOI 10.1007/s001170050569
[3]  
BUSCH HP, 2002, EUROPEAN RADIOLOGY
[4]   Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography [J].
Floyd, CE ;
Warp, RJ ;
Dobbins, JT ;
Chotas, HG ;
Baydush, AH ;
Vargas-Voracek, R ;
Ravin, CE .
RADIOLOGY, 2001, 218 (03) :683-688
[5]   Image quality vs radiation dose for a flat-panel amorphous silicon detector: a phantom study [J].
Geijer, H ;
Beckman, KW ;
Andersson, T ;
Persliden, J .
EUROPEAN RADIOLOGY, 2001, 11 (09) :1704-1709
[6]   Digital radiography with a large-scale electronic flat-panel detector vs screen-film radiography: observer preference in clinical skeletal diagnostics [J].
Hamers, S ;
Freyschmidt, J ;
Neitzel, U .
EUROPEAN RADIOLOGY, 2001, 11 (09) :1753-1759
[7]  
HARRELL G, 2001, RADIOLOGY, V218, P679
[8]   Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography [J].
Hennigs, SP ;
Garmer, M ;
Jaeger, HJ ;
Classen, R ;
Jacobs, A ;
Gissler, HM ;
Christmann, A ;
Mathias, K .
EUROPEAN RADIOLOGY, 2001, 11 (09) :1688-1696
[9]   Chest imaging with flat-panel detector at low and standard doses:: comparison with storage phosphor technology in normal patients [J].
Herrmann, KA ;
Bonél, H ;
Stäbler, A ;
Kulinna, C ;
Glaser, C ;
Holzknecht, N ;
Geiger, B ;
Schätzl, M ;
Reiser, MF .
EUROPEAN RADIOLOGY, 2002, 12 (02) :385-390
[10]   Reduction of radiation dose by using digital luminescence radiography on a hand phantom. [J].
Heyne, JP ;
Merbold, H ;
Sehner, J ;
Neumann, R ;
Adler, R ;
Freesmeyer, M ;
Kaiser, WA .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (04) :386-390