A method of reducing grating lobes associated with an ultrasound linear phase array intended for transrectal thermotherapy

被引:30
作者
Gavrilov, LR
Hand, JW
Abel, P
Cain, CA
机构
[1] ROYAL POSTGRAD MED SCH,DEPT SURG,LONDON W12 0NN,ENGLAND
[2] UNIV MICHIGAN,BIOENGN PROGRAM,ANN ARBOR,MI 48109
关键词
D O I
10.1109/58.655626
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Some practical aspects of planar linear ultrasound phased arrays for transrectal thermotherapy of prostate diseases are discussed. Several regimens for driving the array are investigated and spatial distributions of ultrasound intensities are measured in water and compared with computer simulations. Practical recommendations for suppressing grating lobes based on the use of subsets of elements and de-activation of several elements in the army are given, Treatment safety could be increased by adopting these measures since the relative intensities and power in grating lobes and other secondary intensity peaks are decreased, as is the overall ultrasound energy introduced into the body without significant reduction in the maximum power at the focus.
引用
收藏
页码:1010 / 1017
页数:8
相关论文
共 33 条
[12]   A SPHERICAL-SECTION ULTRASOUND PHASED-ARRAY APPLICATOR FOR DEEP LOCALIZED HYPERTHERMIA [J].
EBBINI, ES ;
CAIN, CA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (07) :634-643
[13]  
FOSTER RS, 1993, EUR UROL, V23, P29
[14]  
GAVRILOV LR, 1974, SOV PHYS ACOUST+, V20, P16
[15]  
GAVRILOV LR, 1996, HYPERTHERMIC ONCOLOG, V2, P399
[16]  
GELET A, 1993, EUR UROL, V23, P44
[17]  
HAND JW, 1993, P IEEE ULTR S, P1225
[18]  
HUTCHINSON EB, 1995, P 1995 IEEE ULTR S, P1601
[19]   THE THRESHOLD FOR THERMALLY SIGNIFICANT CAVITATION IN DOG THIGH MUSCLE INVIVO [J].
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (02) :157-169
[20]  
MADERSBACHER S, 1993, EUR UROL, V23, P39