A method of reducing grating lobes associated with an ultrasound linear phase array intended for transrectal thermotherapy

被引:30
作者
Gavrilov, LR
Hand, JW
Abel, P
Cain, CA
机构
[1] ROYAL POSTGRAD MED SCH,DEPT SURG,LONDON W12 0NN,ENGLAND
[2] UNIV MICHIGAN,BIOENGN PROGRAM,ANN ARBOR,MI 48109
关键词
D O I
10.1109/58.655626
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Some practical aspects of planar linear ultrasound phased arrays for transrectal thermotherapy of prostate diseases are discussed. Several regimens for driving the array are investigated and spatial distributions of ultrasound intensities are measured in water and compared with computer simulations. Practical recommendations for suppressing grating lobes based on the use of subsets of elements and de-activation of several elements in the army are given, Treatment safety could be increased by adopting these measures since the relative intensities and power in grating lobes and other secondary intensity peaks are decreased, as is the overall ultrasound energy introduced into the body without significant reduction in the maximum power at the focus.
引用
收藏
页码:1010 / 1017
页数:8
相关论文
共 33 条
[1]   Influence of interaction between elementary transducer on directivity pattern of an array [J].
Assaad, J ;
Bruneel, C .
ULTRASONICS, 1996, 34 (2-5) :107-110
[2]   DESIGN AND EXPERIMENTAL EVALUATION OF ARM INTRACAVITARY ULTRASOUND PHASED-ARRAY SYSTEM FOR HYPERTHERMIA [J].
BUCHANAN, MT ;
HYNYNEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (12) :1178-1187
[3]  
CHINSON EB, 1996, IEEE T ULTRASON FERR, V43, P1032
[4]  
CHINSON EB, 1996, MED PHYS, V23, P767
[5]   FOCUSED US SYSTEM FOR MR IMAGING-GUIDED TUMOR ABLATION [J].
CLINE, HE ;
HYNYNEN, K ;
WATKINS, RD ;
ADAMS, WJ ;
SCHENCK, JF ;
ETTINGER, RH ;
FREUND, WR ;
VETRO, JP ;
JOLESZ, FA .
RADIOLOGY, 1995, 194 (03) :731-737
[6]   THE DEVELOPMENT OF INTRACAVITARY ULTRASONIC APPLICATORS FOR HYPERTHERMIA - A DESIGN AND EXPERIMENTAL-STUDY [J].
DIEDERICH, CJ ;
HYNYNEN, K .
MEDICAL PHYSICS, 1990, 17 (04) :626-634
[7]   INDUCTION OF HYPERTHERMIA USING AN INTRACAVITARY MULTIELEMENT ULTRASONIC APPLICATOR [J].
DIEDERICH, CJ ;
HYNYNEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (04) :432-438
[8]   THE FEASIBILITY OF USING ELECTRICALLY FOCUSED ULTRASOUND ARRAYS TO INDUCE DEEP HYPERTHERMIA VIA BODY CAVITIES [J].
DIEDERICH, CJ ;
HYNYNEN, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1991, 38 (03) :207-219
[9]  
DIEDERICH CJ, 1987, P IEEE ULTRASON S, V2, P871
[10]  
DUCK FA, 1990, PHYSICAL PROPERTIES