Retention of bonded titanium copings fixed to implant abutments

被引:1
|
作者
Mehl, Christian J. [1 ]
Okano, Nobuhiko [2 ]
Ludwig, Klaus [1 ]
Kern, Matthias [1 ]
机构
[1] Univ Kiel, Dept Prosthodont Propaedeut & Dent Mat, Kiel, Germany
[2] Tokyo Med & Dent Univ, Grad Sch, Removable Partial Prosthodont, Tokyo, Japan
来源
JOURNAL OF PROSTHETIC DENTISTRY | 2016年 / 115卷 / 01期
关键词
MECHANICAL-PROPERTIES; DENTAL IMPLANTS; ALLOY; CORROSION; STRENGTH; WEAR; OVERDENTURES; ATTACHMENTS; VANADIUM; NIOBIUM;
D O I
10.1016/j.prosdent.2015.09.012
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. Conical abutments can be laser welded to the abutment base to compensate for differing implant axes. As laser welding requires expensive equipment, alternative methods for bonding the conical abutment part with the abutment base should be considered. Purpose. The purpose of this in vitro study was to evaluate the retention of Ti-6AI-7Nb copings bonded adhesively to Ti-6Al-7Nb abutments and to compare it with the stability achieved by laser welding. Material and methods. A total of 104 two-part Ti-6Al-7Nb specimens were prepared and divided into 13 groups with 8 specimens each. In this 2-part study, 3 luting resins (Panavia F 2.0 [PF]; Multilink Automix [MA]; Superbond C&B [S13]) were used with or without metal priming (PR). The laser welding group (LW) served as the control. After storage for 1 or 150 days (150 days with thermal cycling [TC]), push-out retention and welded joint stability were tested. The data were analyzed with ANOVA and the Tukey HSD multiple comparison tests. Results. The choice of resin, thermal cycling, and metal priming had a significant effect on resin push-out retention. LW provided the strongest retention, followed by PF, MA, and SB. For PF and SB, TC decreased retention. PR did not lead to higher retention but provided better bonding stability when IC was applied. Conclusions. The retention values suggest that considering the maximum mastication forces, resin bonding is an appropriate substitute for the laser welding method.
引用
收藏
页码:26 / 34
页数:9
相关论文
共 50 条
  • [1] Retention of zirconia ceramic copings bonded to titanium abutments
    Ebert, Agnes
    Hedderich, Juergen
    Kern, Matthias
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2007, 22 (06) : 921 - 927
  • [2] Retention of cast crown copings cemented to implant abutments
    Dudley, J. E.
    Richards, L. C.
    Abbott, J. R.
    AUSTRALIAN DENTAL JOURNAL, 2008, 53 (04) : 332 - 339
  • [3] Retention of titanium copings to implant-supported fixed dental prostheses
    Chiam, Sieu Yien
    Lee, Hsin Lin
    Bedrossian, Armand E.
    Xu, Qianhui
    Kuykendall, William
    Ren, Anna
    Hess, Timothy A.
    Ramos Jr, Van
    Chung, Kwok-Hung
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2024, 33 (04): : 340 - 347
  • [4] Effect of reduced airborne-particle abrasion pressure on the retention of zirconia copings resin bonded to titanium abutments
    Hansen, Nadine Anika
    Wille, Sebastian
    Kern, Matthias
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (01): : 60 - 67
  • [5] Retention of zirconia copings on zirconia implant abutments cemented with provisional luting agents
    Kokubo, Y.
    Kano, T.
    Tsumita, M.
    Sakurai, S.
    Itayama, A.
    Fukushima, S.
    JOURNAL OF ORAL REHABILITATION, 2010, 37 (01): : 48 - 53
  • [6] Evaluation of Retention Forces of Implant-Supported Zirconia Copings on Titanium Abutments Coated with Metal Opaquers Using Different Cements
    Lei, Wenlong
    Guo, Jingmei
    Du, Rui
    Shi, Bin
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (02) : 339 - 345
  • [7] The effect of resin cements and primer on retentive force of zirconia copings bonded to zirconia abutments with insufficient retention
    Kim, Seung-Mi
    Yoon, Ji-Young
    Lee, Myung-Hyun
    Oh, Nam-Sik
    JOURNAL OF ADVANCED PROSTHODONTICS, 2013, 5 (02): : 198 - 203
  • [8] Effect of hemispherical dimples at titanium implant abutments for the retention of cemented crowns
    Choi, Jung-Hoon
    Heo, Seong-Joo
    Koak, Jai-Young
    Kim, Seong-Kyun
    Park, Ji-Man
    Ahn, Jin-Soo
    JOURNAL OF ADVANCED PROSTHODONTICS, 2023, 15 (02): : 63 - 71
  • [9] Evaluation of Heat Transfer to the Implant-Bone Interface During Removal of Metal Copings Cemented onto Titanium Abutments
    Cakan, Umut
    Cakan, Murat
    Delilbasi, Cagri
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2016, 29 (03) : 290 - 292
  • [10] Dislodgement Resistance of Zirconia Copings Cemented onto Zirconia and Titanium Abutments
    Guler, Umut
    Budak, Yasemin
    Cavalcanti Queiroz, Jose Renato
    Ozcan, Mutlu
    IMPLANT DENTISTRY, 2017, 26 (04) : 510 - 515