Validation of use of annular once-punched filter paper bloodspot samples for repeat lead testing

被引:19
作者
El-Hajjar, Dana F.
Swanson, Katherine H.
Landmark, James D.
Stickle, Douglas F. [1 ]
机构
[1] Univ Nebraska, Ctr Med, Dept Pathol & Microbiol, Omaha, NE 68198 USA
[2] Nebraska Publ Hlth Lab, Omaha, NE USA
关键词
lead; filter paper; bloodspot; ICP-MS;
D O I
10.1016/j.cca.2006.09.020
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Lack of a second sample often precludes the ability to perform repeat lead measurements on filter paper bloodspot samples. We investigated whether annular specimens remaining from once-punched filter paper bloodspot specimens could provide accurate Pb measurements when measurements were scaled for the remainder area relative to the original punch area. Methods: 50 mu l bloodspot specimens were prepared using Pb-spiked EDTA whole blood. After removal of 6 mm punches, bitmap images of the remainder specimens were obtained using a scanner. Image analysis was used to determine the bloodspot area of the remainder sample relative to the area of the original punch. Measurement of Pb for punches and for remainder specimens was performed by ICP-MS. Results: Area-corrected Pb measurements for remainder samples were significantly higher than for the punches, by an average factor of 1.52 +/- 0.12 (p < 0.05, n=28). The difference was due to a discontinuity (an increase) in the per-area Pb at the bloodspot perimeter. Area-corrected results for annular specimens that excluded the perimeter were identical to those of the punch. Conclusion: Area-corrected-Pb measurement using annular once-punched bloodspot remainder specimens can accurately reproduce lead measurements obtained from the original punch when the bloodspot perimeter area is excluded. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:179 / 184
页数:6
相关论文
共 19 条
[1]  
Adam BW, 2000, CLIN CHEM, V46, P126
[2]  
Ashley K, 1996, AM IND HYG ASSOC J, V57, P161, DOI 10.1202/0002-8894(1996)057<0161:EOACSK>2.0.CO
[3]  
2
[4]  
CERNIK AA, 1974, BRIT J IND MED, V31, P239
[5]  
Christensen J. M., 1994, Handbook of Metals in Clinical and Analytical Chemistry, P425
[6]  
*CTR DIS CONTR PRE, 1997, SCREEN YOUNG CHILDR
[7]  
*CTR DIS CONTR PRE, 1991, PREV LEAD POIS YOUNG
[8]  
*CTR DIS CONTR PRE, 2000, MMWR-MORBID MORTAL W, V49, pRR14
[9]   EDTA is essential to recover lead from dried blood spots on filter paper [J].
Di Martino, MT ;
Michniewicz, A ;
Martucci, M ;
Parlato, G .
CLINICA CHIMICA ACTA, 2004, 350 (1-2) :143-150
[10]   Use of filter paper for the collection and analysis of human whole blood specimens [J].
Mei, JV ;
Alexander, JR ;
Adam, BW ;
Hannon, WH .
JOURNAL OF NUTRITION, 2001, 131 (05) :1631S-1636S