Tendon-driven continuum robot for neuroendoscopy: validation of extended kinematic mapping for hysteresis operation
被引:67
作者:
Kato, Takahisa
[1
,2
,3
]
Okumura, Ichiro
[4
]
Kose, Hidekazu
[4
]
Takagi, Kiyoshi
[4
]
Hata, Nobuhiko
[1
,2
,5
]
机构:
[1] Brigham & Womens Hosp, Natl Ctr Image Guided Therapy, 75 Francis St, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
[3] Canon USA Inc, Healthcare Opt Res Lab, Cambridge, MA USA
[4] Canon Inc, Mech R&D Ctr, Tokyo, Japan
[5] Brigham & Womens Hosp, Dept Radiol, 75 Francis St, Boston, MA 02115 USA
基金:
美国国家卫生研究院;
关键词:
Neuroendoscopy;
Continuum robot;
Tendon-driven robot;
Robot kinematics;
Robot control;
CATHETER;
DESIGN;
D O I:
10.1007/s11548-015-1310-2
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
The hysteresis operation is an outstanding issue in tendon-driven actuation-which is used in robot-assisted surgery-as it is incompatible with kinematic mapping for control and trajectory planning. Here, a new tendon-driven continuum robot, designed to fit existing neuroendoscopes, is presented with kinematic mapping for hysteresis operation. With attention to tension in tendons as a salient factor of the hysteresis operation, extended forward kinematic mapping (FKM) has been developed. In the experiment, the significance of every component in the robot for the hysteresis operation has been investigated. Moreover, the prediction accuracy of postures by the extended FKM has been determined experimentally and compared with piecewise constant curvature assumption. The tendons were the most predominant factor affecting the hysteresis operation of the robot. The extended FKM including friction in tendons predicted the postures in the hysteresis operation with improved accuracy (2.89 and 3.87 mm for the single and the antagonistic-tendons layouts, respectively). The measured accuracy was within the target value of 5 mm for planning of neuroendoscopic resection of intraventricle tumors. The friction in tendons was the most predominant factor for the hysteresis operation in the robot. The extended FKM including this factor can improve prediction accuracy of the postures in the hysteresis operation. The trajectory of the new robot can be planned within target value for the neuroendoscopic procedure by using the extended FKM.