Detection of Ischemic Electroencephalography Changes During Carotid Endarterectomy Using Synchronization Likelihood Analysis

被引:6
|
作者
Rijsdijk, Mienke [1 ]
Ferrier, Cyrille [2 ]
Laman, Martin [3 ]
Kesecioglu, Jozef [1 ]
Stam, Kees
Slooter, Arjen [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Intens Care, NL-3508 GA Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Clin Neurophysiol, NL-3508 GA Utrecht, Netherlands
[3] Vrije Univ Amsterdam Med Ctr, St Lucas Andreas Hosp, Dept Clin Neurophysiol, Amsterdam, Netherlands
关键词
electroencephalography; synchronization likelihood; carotid endarterectomy; brain ischemia; INTENSIVE-CARE-UNIT; CONTINUOUS EEG; SEIZURE DETECTION; STROKE;
D O I
10.1097/ANA.0b013e3181ada2bb
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Introduction: The intensive care unit is short of a suitable tool for instant detection of cerebral ischemia. Synchronization likelihood (SL) electroencephalography (EEG) analysis has proven to be a promising approach for instant online seizure detection. The aim of this study was to investigate the value of SL analysis in the detection of cerebral ischemia. Patients and Methods: SL analysis was studied using conventional EEG recordings during 143 carotid endarterectomy procedures. We compared EEG data before and after clamping of the carotid artery for 2 different anesthesia protocols (isoflurane or propofol) in patients with and without development of cerebral ischemia, making use of SL analysis. Cerebral ischemia was defined by using EEG variables. Results: Fifty-eight patients received isoflurane and 85 propofol anesthesia of whom overall 27% developed ischemia. In patients from the isoflurane group who developed ischemia, the mean overall SL decreased [from 0.220 (SD 0.052) before clamping to 0.208 (SD 0.044) after clamping; P = 0.06] In patients with ischemia in the propofol group, the mean overall SL remained stable (0.185; P = 0.87) Patients from both groups without development of ischemia had increased mean overall SL values after clamping [isoflurane: from 0.210 (SD 0.041) before clamping to 0.219 (SD 0.051) after clamping; P = 0.08. Propofol: from 0.188 (SD 0.019) to 0.189 (SD 0.021); P = 0.57]. Conclusions: No significant changes in the mean SL were observed after development of ischemia during either isoflurane or propofol anesthesia. SL analysis does not seem suitable for detection of cerebral ischemia in anesthetized patients.
引用
收藏
页码:302 / 306
页数:5
相关论文
共 50 条
  • [1] Diagnostic accuracy of somatosensory evoked potential and electroencephalography during carotid endarterectomy
    Thirumala, Parthasarathy D.
    Natarajan, Piruthiviraj
    Thiagarajan, Karthy
    Crammond, Donald J.
    Habeych, Miguel E.
    Chaer, Rabih A.
    Avgerinos, Efthymios D.
    Friedlander, Robert
    Balzer, Jeffrey R.
    NEUROLOGICAL RESEARCH, 2016, 38 (08) : 698 - 705
  • [2] Carotid endarterectomy during the acute period of ischemic stroke
    Berek, P.
    Kopolovets, I.
    Sihotsky, V.
    Kubikova, M.
    Stefanic, P.
    Toth, S.
    Dzsinich, Cs.
    Frankovicova, M.
    COR ET VASA, 2018, 60 (02) : E169 - E173
  • [3] COMPUTER-DERIVED DENSITY SPECTRAL ARRAY IN DETECTION OF MILD ANALOG ELECTROENCEPHALOGRAPHIC ISCHEMIC PATTERN CHANGES DURING CAROTID ENDARTERECTOMY
    KEARSE, LA
    MARTIN, D
    MCPECK, K
    LOPEZBRESNAHAN, M
    JOURNAL OF NEUROSURGERY, 1993, 78 (06) : 884 - 890
  • [4] Diagnostic accuracy of EEG changes during carotid endarterectomy in predicting perioperative strokes
    Thirumala, Parthasarathy D.
    Thiagarajan, Karthy
    Gedela, Satyanarayana
    Crammond, Donald J.
    Balzer, Jeffrey R.
    JOURNAL OF CLINICAL NEUROSCIENCE, 2016, 25 : 1 - 9
  • [5] Management of neurophysiological monitoring changes during carotid endarterectomy exposure
    Gigliotti, Michael J.
    Feidt, Amy E.
    Cockroft, Kevin M.
    Wilkinson, D. Andrew
    Simon, Scott D.
    Church, Ephraim W.
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2021, 211
  • [6] Cerebral injury predicted by transcranial Doppler ultrasonography but not electroencephalography during carotid endarterectomy
    Costin, M
    Rampersad, A
    Solomon, RA
    Connolly, ES
    Heyer, EJ
    JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 2002, 14 (04) : 287 - 292
  • [7] Benefits of using dexmedetomidine during carotid endarterectomy: A review
    Nair, Abhijit S.
    SAUDI JOURNAL OF ANAESTHESIA, 2014, 8 (02) : 264 - 267
  • [8] Seizure detection in adult ICU patients based on changes in EEG synchronization likelihood
    Slooter, A. J. C.
    Vriens, E. M.
    Leijten, F. S. S.
    Spijkstra, J. J.
    Girbes, A. R. J.
    van Huffelen, A. C.
    Stam, C. J.
    NEUROCRITICAL CARE, 2006, 5 (03) : 186 - 192
  • [9] Seizure detection in adult ICU patients based on changes in EEG synchronization likelihood
    A. J. C. Slooter
    E. M. Vriens
    F. S. S. Leijten
    J. J. Spijkstra
    A. R. J. Girbes
    A. C. van Huffelen
    C. J. Stam
    Neurocritical Care, 2006, 5 : 186 - 192
  • [10] ANTERIOR COMMUNICATING ARTERY COLLATERAL FLOW PROTECTION AGAINST ISCHEMIC CHANGE DURING CAROTID ENDARTERECTOMY
    LOPEZBRESNAHAN, MV
    KEARSE, LA
    YANEZ, P
    YOUNG, TI
    JOURNAL OF NEUROSURGERY, 1993, 79 (03) : 379 - 382