Long-term thermal stability of fluorescent dye used for chiral amino acid analysis on future spaceflight missions

被引:12
作者
Creamer, Jessica S. [1 ]
Mora, Maria F. [1 ]
Noell, Aaron C. [1 ]
Willis, Peter A. [1 ]
机构
[1] CALTECH, Jet Prop Lab, Pasadena, CA USA
关键词
Amino acids; Biosignatures; Extraterrestrial; In situ analysis; Life detection; CAPILLARY-ELECTROPHORESIS; SEPARATION;
D O I
10.1002/elps.201900268
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Future spaceflight missions focused on life detection will carry with them new, state-of-the-art instrumentation capable of highly selective and sensitive organic analysis. CE-LIF is an ideal candidate for such a mission due to its high separation efficiency and low LODs. One perceived risk of utilizing this technique on a future mission is the stability of the chemical reagents in the spaceflight environment. Here, we present an investigation of the thermal stability of the fluorescent dye (5-carboxyfluorescein succinimidyl ester) used for amino acid analysis. The dye was stored at 4, 25, and 60 degrees C for 1 month, 6 months, 1 year, and 2 years. When stored at 4 degrees C for 2 years, 25 degrees C for 6 months, or 60 degrees C for 1 month there was no effect on CE-LIF assay performance due to dye degradation. Beyond these time points, while the dye degradation begins to interfere with the analysis, it is still possible to perform the analysis and achieve the majority of amino acid biosignature science goals described in the science definition team report for the potential Europa Lander mission. This work indicates that thermal control of the dye at <= 4 degrees C will be needed during transit on future spaceflight missions to maintain dye stability.
引用
收藏
页码:3117 / 3122
页数:6
相关论文
共 16 条
[1]   Enhanced Amine and Amino Acid Analysis Using Pacific Blue and the Mars Organic Analyzer Microchip Capillary Electrophoresis System [J].
Chiesl, Thomas N. ;
Chu, Wai K. ;
Stockton, Amanda M. ;
Amashukeli, Xenia ;
Grunthaner, Frank ;
Mathies, Richard A. .
ANALYTICAL CHEMISTRY, 2009, 81 (07) :2537-2544
[2]  
Council N.R., 2011, Vision and Voyages for Planetary Science in the Decade 2013-2022
[3]   Stability of reagents used for chiral amino acid analysis during spaceflight missions in high-radiation environments [J].
Creamer, Jessica S. ;
Mora, Maria F. ;
Willis, Peter A. .
ELECTROPHORESIS, 2018, 39 (22) :2864-2871
[4]   Enhanced Resolution of Chiral Amino Acids with Capillary Electrophoresis for Biosignature Detection in Extraterrestrial Samples [J].
Creamer, Jessica S. ;
Mora, Maria F. ;
Willis, Peter A. .
ANALYTICAL CHEMISTRY, 2017, 89 (02) :1329-1337
[5]   Chiral separation with gradient elution isotachophoresis for future in situ extraterrestrial analysis [J].
Danger, Gregoire ;
Ross, David .
ELECTROPHORESIS, 2008, 29 (19) :4036-4044
[6]   Chance and Necessity in Biochemistry: Implications for the Search for Extraterrestrial Biomarkers in Earth-like Environments [J].
Davila, Alfonso F. ;
McKay, Christopher P. .
ASTROBIOLOGY, 2014, 14 (06) :534-540
[7]   Assessing Antibody Microarrays for Space Missions: Effect of Long-Term Storage, Gamma Radiation, and Temperature Shifts on Printed and Fluorescently Labeled Antibodies [J].
de Diego-Castilla, Graciela ;
Cruz-Gil, Patricia ;
Mateo-Marti, Eva ;
Fernandez-Calvo, Patricia ;
Rivas, Luis A. ;
Parro, Victor .
ASTROBIOLOGY, 2011, 11 (08) :759-773
[8]  
Getty S. A., AER C 2013 IEEE AER, P1
[9]  
Gilmore D. G., 1999, SPACE MISSION ANAL D
[10]  
Hand K.P., 2017, D97667 JPL