Comparison of spectral CT imaging methods based a photon-counting detector: Experimental study

被引:15
作者
Lee, Youngjin [1 ]
Lee, Seungwan [2 ]
Kim, Hee-Joung [3 ]
机构
[1] Eulji Univ, Coll Hlth Sci, Dept Radiol Sci, 553 Sangseong Daero, Songnam 461713, Gyeonggi Do, South Korea
[2] Konyang Univ, Coll Med Sci, Dept Radiol Sci, 158 Gwanjeodong Ro, Daejeon 302812, South Korea
[3] Yonsei Univ, Coll Hlth Sci, Dept Radiol Sci, 1 Yonseidae Gil, Wonju 220710, Kangwon Do, South Korea
关键词
Photon-counting detector; K-edge imaging; Energy-weighting imaging; DUAL-ENERGY CT; X-RAY; K-EDGE; COMPUTED-TOMOGRAPHY; PIXEL DETECTORS; SYSTEM; FEASIBILITY; RADIATION;
D O I
10.1016/j.nima.2016.01.069
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Photon-counting detectors allow spectral computed tomography (CT) imaging using energy-resolved information from a polychromatic X-ray spectrum. The spectral CT images based on the photon-counting detectors are dependent on the energy ranges defined by energy bins for image acquisition. In this study, Kedge and energy weighting imaging methods were experimentally implemented by using a spectral CT system with a cadmium zinc telluride (CZT)-based photon-counting detector. The spectral CT images were obtained by various energy bins and compared in terms of CNR improvement for investigating the effect of energy bins and the efficiency of the spectral CT imaging methods. The results showed that the spectral CT image quality was improved by using the particular energy bins, which were optimized for each spectral CT imaging method and target material. The CNR improvement was different for the spectral CT imaging methods and target materials. It can be concluded that an appropriate selection of imaging method for each target material and the optimization of energy bin can maximize the quality of spectral CT images. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 29 条
[1]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[2]   A System for the Characterization and Testing of CdZnTe/CdTe Pixel Detectors for X-ray and Gamma-ray Imaging [J].
Arodzero, Anatoli ;
Barber, William C. ;
Damron, Matthew Q. ;
Hartsough, Neal E. ;
Iwanczyk, Jan S. ;
Malakhov, Nail ;
Nygard, Einar ;
Moraes, Danielle ;
Weilhammer, Peter ;
Jarron, Pierre .
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, :3638-3642
[3]   Medipix3: A 64 k pixel detector readout chip working in single photon counting mode with improved spectrometric performance [J].
Ballabriga, R. ;
Campbell, M. ;
Heijne, E. ;
Llopart, X. ;
Tlustos, L. ;
Wong, W. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 633 :S15-S18
[4]   A 20 kpixels CdTe photon-counting imager using XPAD chip [J].
Basolo, S. ;
Berar, J. F. ;
Boudet, N. ;
Breugnon, P. ;
Chantepie, B. ;
Clemens, J. C. ;
Delpierre, P. ;
Dinkespiler, B. ;
Hustache, S. ;
Medjoubi, K. ;
Menouni, M. ;
Morel, C. ;
Pangaud, P. ;
Vigeolas, E. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 589 (02) :268-274
[5]   ASIC with multiple energy discrimination for high-rate photon counting applications [J].
De Geronimo, Gianluigi ;
Dragone, Angelo ;
Grosholz, Joe ;
O'Connor, Paul ;
Vernon, Emerson .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (02) :303-312
[6]   Performance of computed tomography for contrast agent concentration measurements with monochromatic x-ray beams:: comparison of K-edge versus temporal subtraction [J].
Elleaume, H ;
Charvet, AM ;
Corde, S ;
Estéve, F ;
Le Bas, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (18) :3369-3385
[7]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[8]   Multienergy Photon-counting K-edge Imaging: Potential for Improved Luminal Depiction in Vascular Imaging [J].
Feuerlein, Sebastian ;
Roessl, Ewald ;
Proksa, Roland ;
Martens, Gerhard ;
Klass, Oliver ;
Jeltsch, Martin ;
Rasche, Volker ;
Brambs, Hans-Juergen ;
Hoffmann, Martin H. K. ;
Schlomka, Jens-Peter .
RADIOLOGY, 2008, 249 (03) :1010-1016
[9]   Material differentiation by dual energy CT: initial experience [J].
Johnson, Thorsten R. C. ;
Krauss, Bernhard ;
Sedlmair, Martin ;
Grasruck, Michael ;
Bruder, Herbert ;
Morhard, Dominik ;
Fink, Christian ;
Weckbach, Sabine ;
Lenhard, Miriam ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Reiser, Maximilian F. ;
Becker, Christoph R. .
EUROPEAN RADIOLOGY, 2007, 17 (06) :1510-1517
[10]   About the Scattering of Radiation by free Electrons after the new relativistic Quantum dynamics of Dirac. [J].