Ultrashort echo time Cones double echo steady state (UTE-Cones-DESS) for rapid morphological imaging of short T2 tissues

被引:10
|
作者
Jang, Hyungseok [1 ]
Ma, Yajun [1 ]
Carl, Michael [2 ]
Jerban, Saeed [1 ]
Chang, Eric Y. [1 ,3 ]
Du, Jiang [1 ]
机构
[1] Univ Calif San Diego, Dept Radiol, 200 West Arbor Dr, San Diego, CA 92103 USA
[2] GE Healthcare, San Diego, CA USA
[3] Vet Affairs San Diego Healthcare Syst, Radiol Serv, San Diego, CA USA
关键词
DESS; dixon; knee; MSK; SSFP; tibia; UTE; MAGNETIC-RESONANCE; ARTICULAR-CARTILAGE; DIXON TECHNIQUE; FLIP ANGLES; DUAL-ECHO; IN-VIVO; KNEE; DIFFUSION; SEQUENCES; WATER;
D O I
10.1002/mrm.28769
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In this study, we aimed to develop a new technique, ultrashort echo time Cones double echo steady state (UTE-Cones-DESS), for highly efficient morphological imaging of musculoskeletal tissues with short T(2)s. We also proposed a novel, single-point Dixon (spDixon)-based approach for fat suppression. Methods: The UTE-Cones-DESS sequence was implemented on a 3T MR system. It uses a short radiofrequency (RF) pulse followed by a pair of balanced spiral-out and spiral-in readout gradients separated by an unbalanced spoiling gradient in-between. The readout gradients are applied immediately before or after the RF pulses to achieve a UTE image (S+) and a spin/stimulated echo image (S-). Weighted echo subtraction between S+ and S- was performed to achieve high contrast specific to short T-2 tissues, and spDixon was applied to suppress fat by using the intrinsic complex signal of S+ and S-. Six healthy volunteers and five patients with osteoarthritis were recruited for whole-knee imaging. Additionally, two healthy volunteers were recruited for lower leg imaging. Results: The UTE-Cones-DESS sequence allows fast volumetric imaging of musculoskeletal tissues with excellent image contrast for the osteochondral junction, tendons, menisci, and ligaments in the knee joint as well as cortical bone and aponeurosis in the lower leg within 5 min. spDixon yields efficient fat suppression in both S+ and S- images without requiring any additional acquisitions or preparation pulses. Conclusion: The rapid UTE-Cones-DESS sequence can be used for high contrast morphological imaging of short T-2 tissues, providing a new tool to assess their association with musculoskeletal disorders.
引用
收藏
页码:881 / 892
页数:12
相关论文
共 50 条
  • [1] Correction of B0 and linear eddy currents: Impact on morphological and quantitative ultrashort echo time double echo steady state (UTE-DESS) imaging
    Jang, Hyungseok
    Athertya, Jiyo S.
    Jerban, Saeed
    Ma, Yajun
    Lombardi, Alecio F.
    Chung, Christine B.
    Chang, Eric Y.
    Du, Jiang
    NMR IN BIOMEDICINE, 2023, 36 (09)
  • [2] Short T2 imaging using a 3D double adiabatic inversion recovery prepared ultrashort echo time cones (3D DIR-UTE-Cones) sequence
    Ma, Ya-Jun
    Zhu, Yanchun
    Lu, Xing
    Carl, Michael
    Chang, Eric Y.
    Du, Jiang
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (05) : 2555 - 2563
  • [3] Imaging and T2 relaxometry of short-T2 connective tissues in the knee using ultrashort echo-time double-echo steady-state (UTEDESS)
    Chaudhari, Akshay S.
    Sveinsson, Bragi
    Moran, Catherine J.
    McWalter, Emily J.
    Johnson, Ethan M.
    Zhang, Tao
    Gold, Garry E.
    Hargreaves, Brian A.
    MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (06) : 2136 - 2148
  • [4] Three-dimensional ultrashort echo time cones T1ρ (3D UTE-cones-T1ρ) imaging
    Ma, Ya-Jun
    Carl, Michael
    Shao, Hongda
    Tadros, Anthony S.
    Chang, Eric Y.
    Du, Jiang
    NMR IN BIOMEDICINE, 2017, 30 (06)
  • [5] Rapid Estimation of Cartilage T2 Based on Double Echo at Steady State (DESS) With 3 Tesla
    Welsch, Goetz H.
    Scheffler, Klaus
    Mamisch, Tallal C.
    Hughes, Timothy
    Millington, Steven
    Deimling, Michael
    Trattnig, Siegfried
    MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (02) : 544 - 549
  • [6] Weighted subtraction in 3D ultrashort echo time (UTE) imaging for visualization of short T2 tissues of the knee
    Lee, Young Han
    Kim, Sungjun
    Song, Ho-Taek
    Kim, InSeong
    Suh, Jin-Suck
    ACTA RADIOLOGICA, 2014, 55 (04) : 454 - 461
  • [7] Single- and Bicomponent Analyses of T2? Relaxation in Knee Tendon and Ligament by Using 3D Ultrashort Echo Time Cones (UTE Cones) Magnetic Resonance Imaging
    Liu, Jin
    Nazaran, Amin
    Ma, Yajun
    Chen, Haimei
    Zhu, Yanchun
    Du, Jiang
    Li, Shaolin
    Zhou, Quan
    Zhao, Yinghua
    BIOMED RESEARCH INTERNATIONAL, 2019, 2019
  • [8] Overview of Short and Ultrashort T-2/T-2* Echo Time (UTE) Imaging
    Young, Ian
    EMAGRES, 2012, 1 (04): : 747 - 755
  • [9] Three-dimensional ultrashort echo time cones (3D UTE-Cones) magnetic resonance imaging of entheses and tendons
    Chen, Bimin
    Zhao, Yinghua
    Cheng, Xin
    Ma, Yajun
    Chang, Eric Y.
    Kavanaugh, Arthur
    Liu, Sirun
    Du, Jiang
    MAGNETIC RESONANCE IMAGING, 2018, 49 : 4 - 9
  • [10] Accurate T1 mapping of short T2 tissues using a three-dimensional ultrashort echo time cones actual flip angle imaging-variable repetition time (3D UTE-Cones AFI-VTR) method
    Ma, Ya-Jun
    Lu, Xing
    Carl, Michael
    Zhu, Yanchun
    Szeverenyi, Nikolaus M.
    Bydder, Graeme M.
    Chang, Eric Y.
    Du, Jiang
    MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (02) : 598 - 608