PET/CT image fusion error due to urinary bladder filling changes: consequence and correction

被引:8
作者
Heiba, Sherif I. [1 ,2 ]
Raphael, Barbara [3 ]
Castellon, Ivan [3 ]
Altinyay, Erkan [3 ]
Sandella, Nick [3 ]
Rosen, Gerald [4 ]
Abdel-Dayem, Hussein M. [3 ]
机构
[1] Mt Sinai Med Ctr, Div Nucl Med, New York, NY 10029 USA
[2] Mt Sinai Sch Med, New York, NY 10029 USA
[3] St Vincents Comprehens Canc Ctr, Nucl Med Serv, New York, NY USA
[4] St Vincents Comprehens Canc Ctr, Div Oncol, New York, NY USA
关键词
PET/CT; (18)FDG; Image fusion; Urinary bladder; ARTIFACTS; PITFALLS;
D O I
10.1007/s12149-009-0304-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective A considerable change of urinary bladder (UB) shape in PET compared with CT in integrated PET/CT system is frequently noted. This study initially evaluated this finding with and without oral contrast (OC) use. In addition, a one bed pelvic section (PLV) repeat acquisition was investigated as a solution to this problem. Methods (18)FDG PET/CTs of 88 patients were analyzed. OC was administered in 68 patients, of whom 31 had PLV images taken 5-10 min later. Three-dimensional mid-UB CT and PET matching measurements were compared. In addition, UB walls displacement between CT and PET were analyzed. Results The mean UB height was significantly increased (P < 0.001) in PET when compared with CT, both anteriorly and posteriorly; however, UB width and depth were not significantly different. An upward shift of superior UB wall in PET from equivalent CT images was noted, whereas there was no appreciable displacement of the other UB walls. The percent UB height increase on PET from CT was significantly greater with than without OC use. The UB height difference between PET and CT was markedly reduced on PLV when compared with the original scans. Conclusions Caution should be exerted during the interpretation of PET/CT scans of the pelvis as there is significant upward expansion of UB on PET compared with CT that appears to be exaggerated by OC use, likely due to additional fluid load. The PET/CT fusion errors of UB can be substantially resolved through a separate PLV acquisition presumably due to the shorter time interval of UB scan completion between CT and PET.
引用
收藏
页码:739 / 744
页数:6
相关论文
共 9 条
[1]  
Beyer T, 2004, J NUCL MED, V45, p25S
[2]   Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology [J].
Beyer, T ;
Antoch, G ;
Blodgett, T ;
Freudenberg, LF ;
Akhurst, T ;
Mueller, S .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) :588-596
[3]   Positron emission tomography/computed tomography: Protocol issues and options [J].
Blodgett, TM ;
McCook, BM ;
Federle, MP .
SEMINARS IN NUCLEAR MEDICINE, 2006, 36 (02) :157-168
[4]   Pitfalls and artifacts in 18FDG PET and PET/CT oncologic imaging [J].
Cook, GJR ;
Wegner, EA ;
Fogelman, I .
SEMINARS IN NUCLEAR MEDICINE, 2004, 34 (02) :122-133
[5]   Whole-body PET/CT:: Spectrum of physiological variants, artifacts and interpretative pitfalls in cancer patients [J].
Gorospe, L ;
Raman, S ;
Echeveste, J ;
Avril, N ;
Herrero, Y ;
Hernández, S .
NUCLEAR MEDICINE COMMUNICATIONS, 2005, 26 (08) :671-687
[6]   PET/CT imaging techniques, considerations, and artifacts [J].
Malvlawi, Osama ;
Pan, Tinsu ;
Macapinlac, Homer A. .
JOURNAL OF THORACIC IMAGING, 2006, 21 (02) :99-110
[7]   Anterior layering of excreted 18F-FDG in the bladder on PET/CT:: Frequency and cause [J].
Purcell, Derk D. ;
Coakley, Fergus V. ;
Franc, Benjamin L. ;
Hawkins, Randall A. ;
Boddington, Sophie E. ;
Yeh, Benjamin M. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (02) :W96-W99
[8]  
Vogel WV, 2004, J NUCL MED, V45, p15S
[9]  
Wahl RL, 2004, J NUCL MED, V45, p82S