Impact of the Balloon Inflation Time and Pattern on the Coronary Stent Expansion

被引:5
|
作者
Skowronski, Jaroslaw [1 ]
Wolny, Rafal [1 ]
Jastrzebski, Jan [1 ]
Tyczynski, Pawel [1 ]
Szlazak, Karol [2 ]
Pregowski, Jerzy [1 ]
Mintz, Gary S. [3 ]
Lizewska, Karolina [2 ]
Swieszkowski, Wojciech [2 ]
Chmielak, Zbigniew [1 ]
Witkowski, Adam [1 ]
机构
[1] Inst Cardiol, Dept Intervent Cardiol & Angiol, Warsaw, Poland
[2] Warsaw Univ Technol, Fac Mat Sci & Engn, Warsaw, Poland
[3] Cardiovasc Res Fdn, New York, NY USA
关键词
OPTICAL COHERENCE TOMOGRAPHY; INTRAVASCULAR ULTRASOUND; IMPLANTATION; THROMBOSIS; DEPLOYMENT; UNDEREXPANSION; OUTCOMES; IVUS;
D O I
10.1155/2019/6945372
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. To assess the expansion pattern of coronary stents by using different balloon inflation times and pressures. Background. The selection of coronary stent size and its proper deployment is crucial in coronary artery interventions, having an impact on the success of the procedure and further therapy. Methods. Ten pairs of different stents were deployed under nominal pressure using sequential (5, 5, 10, and 10 seconds of repeated inflations, thus 30 seconds of summarized time) and continuous (30 seconds) deployment pattern. After each given time-point, intraluminal stent measurements were performed by optical coherence tomography (OCT) and intravascular ultrasound (IVUS). Results. Both in-stent diameters and cross-section areas (CSA) of paired stents measured by OCT at all sequential time-points were significantly smaller compared to given manufacturers charts' values (90% to 94% for diameters and 81% to 88% for CSA, p<0.05). Significant increase of in-stent diameter and CSA was observed across the step-by-step deployment pattern. In-stent lumen measurements were significantly larger when sequential deployment pattern was applied compared to continuous deployment. Additional measurements were also done for overlapping segments of stents, showing smaller in-stent measurements of the latter compared to nonoverlapping segments. Validation of OCT and IVUS measurements using a phantom metallic tube showed perfect reproducibility with OCT and overestimation with IVUS (8% for diameters and 16% for CSA). Conclusions. Stent diameter after deployment is time-dependent and not only pressure-dependent. Different stent expansion behavior, depending on the applied deployment pattern (sequential and nonsequential), was observed.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Impact of stent strut link location in proximal balloon edge dilation technique for bifurcation percutaneous coronary intervention
    Kume, Teruyoshi
    Nishi, Takeshi
    Murasato, Yoshinobu
    Koto, Satoshi
    Sasahira, Yoshitaka
    Okamoto, Hiroshi
    Yamada, Ryotaro
    Koyama, Terumasa
    Tamada, Tomoko
    Imai, Koichiro
    Neishi, Yoji
    Uemura, Shiro
    CARDIOVASCULAR INTERVENTION AND THERAPEUTICS, 2024, 39 (02) : 137 - 144
  • [32] The duration of balloon inflation affects the luminal diameter of coronary segments after bioresorbable vascular scaffolds deployment
    Sorrentino, Sabato
    De Rosa, Salvatore
    Ambrosio, Giuseppe
    Mongiardo, Annalisa
    Spaccarotella, Carmen
    Polimeni, Alberto
    Sabatino, Jolanda
    Torella, Daniele
    Caiazzo, Gianluca
    Indolfi, Ciro
    BMC CARDIOVASCULAR DISORDERS, 2015, 15
  • [33] Intimal disruption affects drug-eluting cobalt-chromium stent expansion: A randomized trial comparing scoring and conventional balloon predilation
    Jujo, Kentaro
    Saito, Katsumi
    Ishida, Issei
    Kim, Ahsung
    Suzuki, Yuki
    Furuki, Yuho
    Ouchi, Taisuke
    Ishii, Yasuhiro
    Sekiguchi, Haruki
    Yamaguchi, Junichi
    Ogawa, Hiroshi
    Hagiwara, Nobuhisa
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 221 : 23 - 31
  • [34] Differences in Vascular Response between Balloon Overstretch and Stent Overexpansion in Nonatherosclerotic Porcine Coronary Arteries
    Mitsutake, Yoshiaki
    Reifart, Joerg
    Pyun, Wook Bum
    Lyons, Jennifer K.
    Deuse, Tobias
    Schrepfer, Sonja
    Ikeno, Fumiaki
    COMPARATIVE MEDICINE, 2017, 67 (04) : 350 - 355
  • [35] Final kissing balloon inflation for coronary bifurcation lesions treated with single-stent technique: Ameta-analysis
    Liu, G.
    Ke, X.
    Huang, Z. -B.
    Wang, L. -C.
    Huang, Z. -N.
    Guo, Y.
    Long, M.
    Liao, X. -X.
    HERZ, 2019, 44 (04) : 354 - 362
  • [36] Impact of Coronary Plaque Characteristics on Late Stent Malapposition after Drug-Eluting Stent Implantation
    Hong, Sung-Jin
    Kim, Byeong-Keuk
    Shin, Dong-Ho
    Kim, Jung-Sun
    Ko, Young-Guk
    Choi, Donghoon
    Jang, Yangsoo
    Hong, Myeong-Ki
    YONSEI MEDICAL JOURNAL, 2015, 56 (06) : 1538 - 1544
  • [37] Excimer laser coronary angioplasty combined with drug-coated balloon in the treatment of in-stent restenosis
    He, Pan
    Chen, Haiwei
    Yang, Junjie
    Gao, Lei
    Guo, Jun
    Chen, Yundai
    Wang, Qi
    LASERS IN SURGERY AND MEDICINE, 2024, 56 (05) : 474 - 484
  • [38] Impact of in-stent tissue characteristics on excimer laser coronary angioplasty prior to drug-coated balloon treatment
    Ishihara, Takayuki
    Dohi, Tomoharu
    Nakamura, Daisuke
    Kikuchi, Atsushi
    Okamoto, Naotaka
    Mori, Naoki
    Iida, Osamu
    Tsujimura, Takuya
    Mizote, Isamu
    Higuchi, Yoshiharu
    Yamada, Takahisa
    Nishino, Masami
    Mano, Toshiaki
    Sakata, Yasushi
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2021, 339 : 28 - 32
  • [39] Assessment of Post-Dilatation Strategies for Optimal Stent Expansion in Calcified Coronary Lesions: Ex Vivo Analysis With Optical Coherence Tomography
    Dallan, Luis A. P.
    Zimin, Vladislav N.
    Lee, Juhwan
    Gharaibeh, Yazan
    Kim, Justin N.
    Pereira, Gabriel T. R.
    Vergara-Martel, Armando
    Dong, Pengfei
    Gu, Linxia
    Wilson, David L.
    Bezerra, Hiram G.
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2022, 43 : 62 - 70
  • [40] Impact of everolimus-eluting stent length on long-term clinical outcomes of percutaneous coronary intervention
    Yano, Hideki
    Horinaka, Shigeo
    Ishimitsu, Toshihiko
    JOURNAL OF CARDIOLOGY, 2018, 71 (5-6) : 444 - 451