Acousto-optic tunable filter: A new generation monochromator and more

被引:21
作者
Tran, CD [1 ]
机构
[1] Marquette Univ, Dept Chem, Milwaukee, WI 53201 USA
关键词
acousto-optic tunable filter; laser; HPLC; flow injection analysis; detector; near-infrared; multispectral imaging;
D O I
10.1080/00032710008543155
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The acousto-optic tunable filter is a compact, all solid state electronic monochromator. It is based on the acousto-optic interaction in an anisotropic crystal. Compared to conventional grating monochromtors, the AOTF has no moving parts, wider spectral tuning range (from UV though visible and near-IR to IR), higher throughput, higher resolution (few angstroms), faster scanning speed (ys) and imaging capability. These features make it possible to use the filter to develop novel instruments which are not possible otherwise. The instrumentation development and unique advantages of such AOTF based instruments including a multidimensional fluorimeter, detectors for HPLC and flow injection analysis, and a multispectral imaging instrument are described.
引用
收藏
页码:1711 / 1732
页数:22
相关论文
共 23 条
[1]   Near-infrared detection of flow injection analysis by acoustooptic tunable filter-based spectrophotometry [J].
Baptista, MS ;
Tran, CD ;
Gao, GH .
ANALYTICAL CHEMISTRY, 1996, 68 (06) :971-976
[2]  
BURNS DA, 1992, HDB NEAR INFRARED AN
[3]  
DESURVIRE E, 1994, ERBIUM DOPED FIBER A
[4]   Investigation of solid phase peptide synthesis by the near-infrared multispectral imaging technique: A detection method for combinatorial chemistry [J].
Fischer, M ;
Tran, CD .
ANALYTICAL CHEMISTRY, 1999, 71 (13) :2255-2261
[5]  
HUBER L, 1993, DIODE ARRAY DETECTIO
[6]  
Morris M. D., 1993, Microscopic and Spectroscopic Imaging of the Chemical State
[7]   Detection of flow injection analysis with pH gradient by acousto-optic tunable filter based spectrophotometry [J].
Pasquini, C ;
Lu, J ;
Tran, CD ;
Smirnov, S .
ANALYTICA CHIMICA ACTA, 1996, 319 (03) :315-324
[8]  
Ruxicka J., 1988, FLOW INJECTION ANAL
[9]   Determination of binding constants by flow injection gradient technique [J].
Tran, CD ;
Baptista, MS ;
Tomooka, T .
LANGMUIR, 1998, 14 (24) :6886-6892
[10]   Detection of flow injection analysis by acousto-optic tunable filter-based fluorimetry and its application in the determination of solvent polarity [J].
Tran, CD ;
Lu, J .
APPLIED SPECTROSCOPY, 1996, 50 (12) :1578-1584