Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer

被引:33
作者
Gao, Feng [1 ]
Liu, Hu [1 ]
Zhao, Qing [1 ]
Pan, Yingzhe [1 ]
机构
[1] Hubei Univ Med, Peoples Hosp Xiangyang 1, Dept Ophthalmol, 15 Jiefang Rd, Xiangyang 441000, Hubei, Peoples R China
关键词
iCare rebound tonometer; Goldmann applanation tonometry; intraocular pressure; central corneal thickness; CENTRAL CORNEAL THICKNESS; INTRAOCULAR-PRESSURE; GLAUCOMA;
D O I
10.3892/etm.2017.4164
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Tonometry is a fundamental procedure in routine ophthalmologic examination. Although regarded as the reference standard, the Goldmann applanation tonometer (GAT) has its limitations. A new portable alternative to the GAT is the iCare rebound tonometer (RT). The aim of the present study was to compare the intraocular pressure (IOP) results obtained using the RT and GAT and then correlate the results with the central corneal thickness (CCT). Moreover, the tolerability and safety of the RT were evaluated. The IOP of 336 patients (672 eyes) was determined by the RT and GAT. The patients were divided into three groups (group A, 7-15 mmHg, n=74; group B, 16-22 mmHg, n=218; and group C, 23-50 mmHg, n=44), based on the GAT IOP readings. Pachymetry and slit lamp inspection were also performed. To establish an agreement between the devices, a Bland-Altman analysis and paired t-test were performed. The correlation between CCT and IOP readings obtained by the two devices were assessed using linear regression correlation analysis. The mean IOP values of the RT and the GAT were 18.30 +/- 5.10 and 18.52 +/- 4.46 mmHg, respectively. There were no significant differences between them (t=-1.31, P=0.19). The 95% confidence interval of the differences between the two devices was -5.80-6.24 mmHg. The RT readings are correlated well with those of GAT (r=0.806, P=0.001). However, the RT measurements were significantly (t=-2.84, P=0.007) lower (-1.66 +/- 3.87 mmHg) than those obtained with GAT when GAT =23 mmHg. Both the RT (r=0.390, P=0.001) and the GAT (r=0.191, P=0.001) showed positive correlations with CCT. The IOP measurement with RT was well tolerated. None of the corneal epithelial defects was detected and all subjects denied discomfort. The RT is well tolerated and safe, and can be considered a reliable alternative to GAT for patients in a low to moderate IOP range. However, in patients with high IOP values, the measurements obtained with RT did not correlate well with those obtained by GAT. The RT readings are influenced more by CCT compared to GAT.
引用
收藏
页码:1912 / 1916
页数:5
相关论文
共 50 条
  • [21] Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in patients with glaucoma
    Vandewalle, Evelien
    Vandenbroeck, Sofie
    Stalmans, Ingeborg
    Zeyen, Thierry
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (05) : 783 - 789
  • [22] A comparison of intraocular pressure values obtained with the Tono-Pachymeter NT530P iCare (R) rebound tonometer and Goldmann applanation tonometer
    Rampersad, N.
    Mashige, K. P.
    Jhetamt, S.
    AFRICAN VISION AND EYE HEALTH JOURNAL, 2011, 70 (03):
  • [23] Reliability and reproducibility of introcular pressure (IOP) measurement with the Icare® Home rebound tonometer (model TA022) and comparison with Goldmann applanation tonometer in glaucoma patients
    Valero, B.
    Fenolland, J. -R.
    Rosenberg, R.
    Sendon, D.
    Mesnard, C.
    Sigaux, M.
    Giraud, J. -M.
    Renard, J. -P.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2017, 40 (10): : 865 - 875
  • [24] Comparison of the Icare Tonometer and the Hand-held Goldmann Applanation Tonometer in Pediatric Aphakia
    Li, Yunping
    Tang, Luosheng
    Xiao, Manyi
    Jia, Songbai
    Liu, Ping
    Zhou, Yi
    Chen, Huihui
    Zeng, Jiexi
    JOURNAL OF GLAUCOMA, 2013, 22 (07) : 550 - 554
  • [25] Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry
    Witte, V.
    Glass, A.
    Beck, R.
    Guthoff, R.
    OPHTHALMOLOGE, 2012, 109 (10): : 1008 - 1013
  • [26] Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects
    Kato, Yoshitake
    Nakakura, Shunsuke
    Matsuo, Naoko
    Yoshitomi, Kayo
    Handa, Marina
    Tabuchi, Hitoshi
    Kiuchi, Yoshiaki
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (02) : 687 - 696
  • [27] Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects
    Yoshitake Kato
    Shunsuke Nakakura
    Naoko Matsuo
    Kayo Yoshitomi
    Marina Handa
    Hitoshi Tabuchi
    Yoshiaki Kiuchi
    International Ophthalmology, 2018, 38 : 687 - 696
  • [28] Comparison of the Intraocular Pressure Measured Using the New Rebound Tonometer Icare ic100 and Icare TA01i or Goldmann Applanation Tonometer
    Nakakura, Shunsuke
    Mori, Etsuko
    Fujio, Yuki
    Fujisawa, Yasuko
    Matsuya, Kanae
    Kobayashi, Yui
    Tabuchi, Hitoshi
    Asaoka, Ryo
    Kiuchi, Yoshiaki
    JOURNAL OF GLAUCOMA, 2019, 28 (02) : 172 - 177
  • [29] Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma
    Shunsuke Nakakura
    Ryo Asaoka
    Etsuko Terao
    Yuki Nagata
    Yasuko Fukuma
    Satomi Oogi
    Miku Shiraishi
    Yoshiaki Kiuchi
    Eye and Vision, 8
  • [30] Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma
    Nakakura, Shunsuke
    Asaoka, Ryo
    Terao, Etsuko
    Nagata, Yuki
    Fukuma, Yasuko
    Oogi, Satomi
    Shiraishi, Miku
    Kiuchi, Yoshiaki
    EYE AND VISION, 2021, 8 (01)