Design of Position and Force Sensors of the Haptic Interface for Training Simulation of Radial Artery Puncture

被引:0
|
作者
Li, Chengjie [1 ]
Li, Dezhou [1 ]
Lee, Doo Yong [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Mech Engn, Daejeon 34141, South Korea
来源
2016 16TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS) | 2016年
关键词
Position Sensor; Force Sensor; Haptic Interface; Radial Artery Puncture;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper proposes position and force sensors of the haptic interface for training simulation of radial artery puncture. The position sensor with two optocouplers and one code stick is designed to measure translational movement of the guidewire. The position sensor can measure linear motion with speed up to 160mm/s with 1mm error. The resolution of the designed position sensor is 0.195mm. A leverage structure of the designed force sensor is employed with two force sensitive resistors (FSR) to measure the interaction force. The force sensor can measure 0 similar to 10N in the needle-insertion direction, and 0 similar to 5N in the needle-withdrawal direction. The force sensing errors are 8.85% and 12.22% in the needle-insertion and needle-withdrawal directions, respectively. The RMS errors are 0.52N and 0.37N for each direction.
引用
收藏
页码:93 / 97
页数:5
相关论文
共 11 条
  • [1] Haptic Interface with Hydraulic Actuator for Training Simulation of Radial Artery Puncture
    Li, Chengjie
    Lee, Doo Yong
    2018 18TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS), 2018, : 24 - 28
  • [2] Design of a Haptic Interface for Simulation of Needle Intervention
    Kang, Seung Gyu
    Lee, Doo Yong
    2013 9TH ASIAN CONTROL CONFERENCE (ASCC), 2013,
  • [3] Design of a Haptic Interface for a Gastrointestinal Endoscopy Simulation
    Yu, In Sik
    Woo, Hyun Soo
    Son, Hyoung Il
    Ahn, Woojin
    Jung, Hoeryong
    Lee, Doo Yong
    Yi, Sun Young
    ADVANCED ROBOTICS, 2012, 26 (18) : 2115 - 2143
  • [4] A Design of Hardware Haptic Interface for Gastrointestinal Endoscopy Simulation
    Gu, Yunjin
    Lee, Doo Yong
    MEDICINE MEETS VIRTUAL REALITY 18, 2011, 163 : 199 - 201
  • [5] WeHAPTIC: A Wearable Haptic interface for Accurate Position Tracking and Interactive force Control
    Park, Yeongyu
    Jo, Inseong
    Lee, Jeongsoo
    Bae, Joonbum
    MECHANISM AND MACHINE THEORY, 2020, 153
  • [6] Integration of force-position control and haptic interface facilities for a virtual excavator simulator
    Torres-Rodríguez, HI
    Parra-Vega, V
    Ruiz-Sánchez, FJ
    2005 12th International Conference on Advanced Robotics, 2005, : 761 - 768
  • [7] Application of Two-port Network in Bilateral Control for a Haptic Interface with Force-position Compensation
    Meng, Jian Jun
    Yun, Jin Tian
    ADVANCES IN MECHANICAL DESIGN, PTS 1 AND 2, 2011, 199-200 : 1211 - 1216
  • [8] A shared-control approach to haptic interface design for minimally invasive telesurgical training
    Nudehi, SS
    Mukherjee, R
    Ghodoussi, M
    IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 2005, 13 (04) : 588 - 592
  • [9] The application of the acoustic shadowing facilitates guidance in radial artery puncture and cannulation teaching in standardized training for residents: a randomized controlled trial
    Dong, Rui
    Chen, Jingyan
    Wang, Hong
    Liu, Zhilin
    Sun, Xiaopeng
    Guo, Yuwei
    Wang, Mingshan
    Sun, Lixin
    Gu, Xiaoping
    BMC MEDICAL EDUCATION, 2022, 22 (01)
  • [10] The application of the acoustic shadowing facilitates guidance in radial artery puncture and cannulation teaching in standardized training for residents: a randomized controlled trial
    Rui Dong
    Jingyan Chen
    Hong Wang
    Zhilin Liu
    Xiaopeng Sun
    Yuwei Guo
    Mingshan Wang
    Lixin Sun
    Xiaoping Gu
    BMC Medical Education, 22