Radiation Dose From Cone Beam CT in a Pediatric Phantom: Risk Estimation of Cancer Incidence

被引:40
作者
Kim, Sangroh [2 ]
Yoshizumi, Terry T. [1 ,3 ,4 ]
Frush, Donald P. [1 ]
Toncheva, Greta [1 ,3 ]
Yin, Fang-Fang [4 ]
机构
[1] Duke Univ, Dept Radiol, Med Ctr, Durham, NC 27710 USA
[2] Duke Univ, Med Phys Grad Program, Med Ctr, Durham, NC 27710 USA
[3] Duke Univ, Div Radiat Safety, Med Ctr, Durham, NC 27710 USA
[4] Duke Univ, Dept Radiat Oncol, Med Ctr, Durham, NC 27710 USA
关键词
CBCT; cone beam CT; effective dose; lifetime attributable risk; MOSFET; pediatric phantom; CBCT;
D O I
10.2214/AJR.08.2168
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The objective of our study was to measure absorbed doses and calculate effective dose (ED) from cone beam CT (CBCT) with metal oxide semiconductor field effect transistor (MOSFET) detectors in an anthropomorphic phantom and to estimate the risk of cancer incidence for CBCT. MATERIALS AND METHODS. Abdominal CBCT was performed in an anthropomorphic phantom of a 5-year-old child using the On-Board Imager with arbitrarily designated standard-dose (125 kVp, 80 mA, 25 milliseconds) and low-dose (125 kVp, 40 mA, 10 milliseconds) modes. The full-fan mode was used, and 20 MOSFET dosimeters were used to measure the absorbed doses in various organs. We calculated the ED, the lifetime attributable risk (LAR) for cancer incidence, and relative risk (RR) of cancer induction from a single scan for both standard- and low-dose modes in 5-year-old children. RESULTS. The highest absorbed doses were found in the skin, ascending colon, and stomach. The mean ED was 37.8 +/- 0.7 (SD) mSv for the standard- dose mode and 8.1 +/- 0.2 mSv for the low-dose mode. The LAR of cancer incidence ranged from 23 to 144 cases per 100,000 exposed persons for the standard- dose mode and from five to 31 cases per 100,000 exposed persons for the low-dose mode. The RR of cancer incidence ranged from 1.003 to 1.054 for the standard-dose mode and from 1.001 to 1.012 for the low-dose mode. CONCLUSION. The ED from pediatric CBCT using the standard- dose mode was considerably higher than that of MDCT, whereas the ED for CBCT using the low-dose mode was comparable to that of abdominal MDCT. For abdominal CBCT in the pediatric phantom, the highest LARs were for colon and bladder cancers and the highest RRs were for stomach and liver cancers.
引用
收藏
页码:186 / 190
页数:5
相关论文
共 15 条
[1]  
[Anonymous], 1993, ICRP PUBL
[2]  
[Anonymous], 1975, ICRP Publication, V23
[3]   Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative [J].
Brenner, DJ .
PEDIATRIC RADIOLOGY, 2002, 32 (04) :228-231
[4]  
CRISTY M, 1987, ORNLTM8381, V1
[5]  
Goldstone K., 1989, ICRU report 44, DOI DOI 10.1093/JICRU/OS23.1.REPORT44
[6]   Radiation dose from contemporary cardiothoracic multidetector CT Protocols with an anthropomorphic female phantom: Implications for cancer induction [J].
Hurwitz, Lynne M. ;
Reiman, Robert E. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Toncheva, Greta ;
Nguyen, Giao ;
Lowry, Carolyn .
RADIOLOGY, 2007, 245 (03) :742-750
[7]  
*ICRP, 1996, ICRP PUBLICATION, V70
[8]   Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy [J].
Islam, Mohammad K. ;
Purdie, Thomas G. ;
Norrlinger, Bernhard D. ;
Alasti, Hamideh ;
Moseley, Douglas J. ;
Sharpe, Michael B. ;
Siewerdsen, Jeffrey H. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2006, 33 (06) :1573-1582
[9]   Organ and effective dose evaluation in diagnostic radiology based on in-phantom dose measurements with novel photodiode-dosemeters [J].
Kawaura, Chiyo ;
Aoyama, Takahiko ;
Koyama, Shuji ;
Achiwa, Masataka ;
Mori, Masaki .
RADIATION PROTECTION DOSIMETRY, 2006, 118 (04) :421-430
[10]  
National Research Council of the National Academies, 2006, HLTH RISKS EXPOSURE