Enterobacter cloacae outbreak in the NICU related to disinfected thermometers

被引:54
作者
van den Berg, RWA
Claahsen, HL
Niessen, M
Muytjens, HL
Liem, K
Voss, A
机构
[1] Univ Nijmegen Hosp, Dept Med Microbiol, NL-6500 HB Nijmegen, Netherlands
[2] Univ Nijmegen Hosp, Dept Neonatol, NL-6500 HB Nijmegen, Netherlands
关键词
Enterobacter cloacae; NICU; disinfection; thermometers; outbreak;
D O I
10.1053/jhin.1999.0657
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
In the first week of December 1997, an increasing incidence of neonates colonized with multidrug resistant Enterobacter cloacae (MR-E. cloacae) was observed in the neonatal intensive care unit of our 950-bed university hospital. Initially re-enforcement of infection control practices including hand disinfection and cohort isolation seemed to be sufficient to control the outbreak. Nevertheless, an increasing number of newly admitted patients was paralleled by another rise in the incidence of colonized neonates. Since E. cloacae was initially found in urine specimens of the patients, surveillance and environmental cultures were aimed at procedures and instruments that might colonize the gastro-intestinal and/or urinary tract. E. cloacae was isolated from a single cap of an electronic digital thermometer. Despite banning of this possible source, newly admitted neonates still became colonized. The unit was closed for further admissions and a second round of extensive screening was started; this time including all available thermometers and continuous rectal temperature probes. Ready-to-use 'disinfected' thermometers and probes were found to be colonized with MR-E. cloacae. Observation of disinfection procedures and a laboratory investigation revealed that 'rushed' disinfection with alcohol 80% led to a 1 in 10 chance of thermometers still being contaminated. Furthermore, alcoholic hand rub used for convenience disinfection failed to disinfect thermometers in 40% and 20% of the cases when done in a 'rushed' or 'careful' fashion, respectively. Adequate disinfection of the thermometers led to the control of the outbreak, with no new occurrence of MR-E. cloacae in the following months. (C) 2000 The Hospital Infection Society.
引用
收藏
页码:29 / 34
页数:6
相关论文
共 36 条
[1]   ENTEROBACTER-CLOACAE IN A NEONATAL INTENSIVE-CARE UNIT - ACCOUNT OF AN OUTBREAK AND ITS RELATIONSHIP TO USE OF 3RD-GENERATION CEPHALOSPORINS [J].
ACOLET, D ;
AHMET, Z ;
HOUANG, E ;
HURLEY, R ;
KAUFMANN, ME .
JOURNAL OF HOSPITAL INFECTION, 1994, 28 (04) :273-286
[2]   AN OUTBREAK OF GENTAMICIN-RESISTANT ENTEROBACTER-CLOACAE INFECTIONS IN A PEDIATRIC INTENSIVE-CARE UNIT [J].
ANDERSON, EL ;
HIEBER, JP .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 1983, 4 (03) :148-152
[3]   Epidemiological study of an outbreak due to multidrug-resistant Enterobacter aerogenes in a medical intensive care unit [J].
Arpin, C ;
Coze, C ;
Rogues, AM ;
Gachie, JP ;
Bebear, C ;
Quentin, C .
JOURNAL OF CLINICAL MICROBIOLOGY, 1996, 34 (09) :2163-2169
[4]  
Ausubel FM., 1994, Curr. Protoc. Mol. Biol
[5]   OUTBREAK OF SURGICAL WOUND INFECTIONS ASSOCIATED WITH TOTAL HIP-ARTHROPLASTY [J].
BECKSAGUE, CM ;
CHONG, WH ;
ROY, C ;
ANDERSON, R ;
JARVIS, WR .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 1992, 13 (09) :526-534
[6]  
Belkum A. van, 1995, Journal of Clinical Microbiology, V33, P1537
[7]  
Belkum A. van, 1994, Clinical Microbiology Reviews, V7, P174
[8]   WHICH THERMOMETER [J].
BLUMENTHAL, I .
LANCET, 1991, 337 (8755) :1483-1483
[9]   Neonatal airway colonization with gram-negative bacilli: Association with severity of bronchopulmonary dysplasia [J].
Cordero, L ;
Ayers, LW ;
Davis, K .
PEDIATRIC INFECTIOUS DISEASE JOURNAL, 1997, 16 (01) :18-23
[10]   A nosocomial outbreak due to Enterobacter cloacae strains with the E-hormaechei genotype in patients treated with fluoroquinolones [J].
DavinRegli, A ;
Bosi, C ;
Charrel, R ;
Ageron, E ;
Papazian, L ;
Grimont, PAD ;
Cremieux, A ;
Bollet, C .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (04) :1008-1010