Functional alignment achieves soft-tissue balance in total knee arthroplasty as measured with quantitative sensor-guided technology

被引:92
作者
Chang, J. S. [1 ,2 ,3 ]
Kayani, B. [1 ,2 ,3 ]
Wallace, C. [1 ,4 ]
Haddad, F. S. [1 ,2 ,3 ]
机构
[1] Univ Coll Hosp, London, England
[2] Univ Coll London Hosp, Dept Trauma & Orthopaed Surg, Orthopaed Surg, London, England
[3] Princess Grace Hosp, Dept Trauma & Orthopaed Surg, London, England
[4] Univ Coll London Hosp, Dept Trauma & Orthopaed Surg, London, England
关键词
CRUCIATE LIGAMENT RESECTION; FLEXION-EXTENSION GAPS; MEDIOLATERAL LAXITY; CLINICAL-OUTCOMES; REPLACEMENT; REVISION; VARUS;
D O I
10.1302/0301-620X.103B.BJJ-2020-0940.R1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aims Total knee arthroplasty (TKA) using functional alignment aims to implant the components with minimal compromise of the soft-tissue envelope by restoring the plane and obliquity of the non-arthritic joint. The objective of this study was to determine the effect of TKA with functional alignment on mediolateral soft-tissue balance as assessed using intraoperative sensor-guided technology. Methods This prospective study included 30 consecutive patients undergoing robotic-assisted TKA using the Stryker PS Triathlon implant with functional alignment. Intraoperative soft-tissue balance was assessed using sensor-guided technology after definitive component implantation; soft-tissue balance was defined as intercompartmental pressure difference (ICPD) of < 15 psi. Medial and lateral compartment pressures were recorded at 10 degrees, 45 degrees, and 90 degrees of knee flexion. This study included 18 females (60%) and 12 males (40%) with a mean age of 65.2 years (SD 9.3). Mean preoperative hip-knee-ankle deformity was 6.3 degrees varus (SD 2.7 degrees). Results TKA with functional alignment achieved balanced medial and lateral compartment pressures at 10 degrees (25.0 psi (SD 6.1) vs 23.1 psi (SD 6.7), respectively; p = 0.140), 45 degrees (21.4 psi (SD 5.9) vs 20.6 psi (SD 5.9), respectively; p = 0.510), and 90 degrees (21.2 psi (SD 7.1) vs 21.6 psi (SD 9.0), respectively; p = 0.800) of knee flexion. Mean ICPD was 6.1 psi (SD 4.5; 0 to 14) at 10 degrees, 5.4 psi (SD 3.9; 0 to 12) at 45 degrees, and 4.9 psi (SD 4.45; 0 to 15) at 90 degrees of knee flexion. Mean postoperative limb alignment was 2.2 degrees varus (SD 1.0 degrees). Conclusion TKA using the functional alignment achieves balanced mediolateral soft-tissue tension through the arc of knee flexion as assessed using intraoperative pressure-sensor technology. Further clinical trials are required to determine if TKA with functional alignment translates to improvements in patient satisfaction and outcomes compared to conventional alignment techniques.
引用
收藏
页码:507 / 514
页数:8
相关论文
共 39 条
[1]   Risk factors, diagnosis and management of prosthetic joint infection after total hip arthroplasty [J].
Ahmed, Syed S. ;
Begum, Fahima ;
Kayani, Babar ;
Haddad, Fares S. .
EXPERT REVIEW OF MEDICAL DEVICES, 2019, 16 (12) :1063-1070
[2]   The effect of patient age at intervention on risk of implant revision after total replacement of the hip or knee: a population-based cohort study [J].
Bayliss, Lee E. ;
Culliford, David ;
Monk, A. Paul ;
Glyn-Jones, Sion ;
Prieto-Alhambra, Daniel ;
Judge, Andrew ;
Cooper, Cyrus ;
Carr, Andrew J. ;
Arden, Nigel K. ;
Beard, David J. ;
Price, Andrew J. .
LANCET, 2017, 389 (10077) :1424-1430
[3]   The Chitranjan Ranawat Award: Is Neutral Mechanical Alignment Normal for All Patients?: The Concept of Constitutional Varus [J].
Bellemans, Johan ;
Colyn, William ;
Vandenneucker, Hilde ;
Victor, Jan .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (01) :45-53
[4]   A functional comparison of medial pivot and condylar knee designs based on patient outcomes and parameters of gait [J].
Benjamin, B. ;
Pietrzak, J. R. T. ;
Tahmassebi, J. ;
Haddad, F. S. .
BONE & JOINT JOURNAL, 2018, 100B (01) :76-82
[5]   Influence of patient factors on TKA outcomes at 5 to 11 years followup [J].
Bourne, Robert B. ;
McCalden, Richard W. ;
MacDonald, Steven J. ;
Mokete, Lipalo ;
Guerin, Jeff .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (464) :27-31
[6]   Patient Satisfaction after Total Knee Arthroplasty Who is Satisfied and Who is Not? [J].
Bourne, Robert B. ;
Chesworth, Bert M. ;
Davis, Aileen M. ;
Mahomed, Nizar N. ;
Charron, Kory D. J. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (01) :57-63
[7]   Why are Total Knee Arthroplasties Being Revised? [J].
Dalury, David F. ;
Pomeroy, Donald L. ;
Gorab, Robert S. ;
Adams, Mary Jo .
JOURNAL OF ARTHROPLASTY, 2013, 28 (08) :120-121
[8]   A randomised controlled trial of kinematically and mechanically aligned total knee replacements TWO-YEAR CLINICAL RESULTS [J].
Dossett, H. G. ;
Estrada, N. A. ;
Swartz, G. J. ;
LeFevre, G. W. ;
Kwasman, B. G. .
BONE & JOINT JOURNAL, 2014, 96B (07) :907-913
[9]   Are Patients More Satisfied With a Balanced Total Knee Arthroplasty? [J].
Golladay, Gregory J. ;
Bradbury, Thomas L. ;
Gordon, Alexander C. ;
Fernandez-Madrid, Ivan J. ;
Krebs, Viktor E. ;
Patel, Preetesh D. ;
Suarez, Juan C. ;
Rueda, Carlos A. Higuera ;
Barsoum, Wael K. .
JOURNAL OF ARTHROPLASTY, 2019, 34 (07) :S195-S200
[10]   Increased satisfaction after total knee replacement using sensor-guided technology [J].
Gustke, K. A. ;
Golladay, G. J. ;
Roche, M. W. ;
Jerry, G. J. ;
Elson, L. C. ;
Anderson, C. R. .
BONE & JOINT JOURNAL, 2014, 96B (10) :1333-1338