Determining the accuracy of blood pressure measurement by the Omron HEM-907 before and after hemodialysis

被引:15
作者
Czarkowski, Marek [1 ]
Staszkow, Monika [2 ]
Kostyra, Kacper [1 ]
Shebani, Zuhier [2 ]
Niemczyk, Stanislaw [2 ]
Matuszkiewicz-Rowinska, Joanna [2 ]
机构
[1] Med Univ Warsaw, Dept Internal Med & Endocrinol, PL-02097 Warsaw, Poland
[2] Med Univ Warsaw, Dept Nephrol Dialysis & Internal Med, PL-02097 Warsaw, Poland
关键词
after hemodialysis; before hemodialysis; blood pressure measurement; dialyzed patients; end-stage renal disease; Omron HEM-907; oscillometric measurement; SOCIETY-OF-HYPERTENSION; EUROPEAN-SOCIETY; INTERNATIONAL-PROTOCOL; ARTERIAL STIFFNESS; MEASURING DEVICES; VALIDATION; MONITOR; SPHYGMOMANOMETER; POPULATION; REPLACE;
D O I
10.1097/MBP.0b013e328331d5b5
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The accuracy of oscillometric blood pressure (BP) measuring devices determined in resting conditions may be distorted in those under hemodialysis (HD). Methods The evaluation of accuracy of the Omron HEM-907 was performed in accordance with the International Protocol for validation of BP measuring devices (I P). Out of a group of 49 patients with end-stage renal disease (ESRD) (56.7 +/- 15.6 years), four groups of 33 patients each were selected to ascertain their BP before and after HD. Results The Omron HEM-907 fulfilled the IP accuracy requirements for systolic BP measurements before and after HD. In the case of diastolic BP before HD, the Omron HEM-907 did not fulfill both requirements for phase 2.2 IP. In only 21 out of 33 patients, the difference between the mean of two observers and the Omron HEM-907 device (Diff) was not more than 5 mmHg, while the requirement was 22 patients. In seven patients, all three Diff were over 5 mmHg, while the accepted limit is three. After HD, the Omron HEM-907 did not fulfill one out of two criteria of phase 2.2 IP - in five people, all three Diff were over 5 mmHg. Conclusion The Omron HEM-907 oscillometric method of BP measurement should not be used to measure BP before and after HD in ESRD patients. In ESRD patients, an alternative method of BP measurement, in the event of a lack of a mercury sphygmomanometer, could be a classical auscultatory method together with the Omron HEM-907, which allows for this type of BP measurement. Blood Press Monit 14:232-238 (C) 2009 Wolters Kluwer Health | Lippincott Williams & Wilkins.
引用
收藏
页码:232 / 238
页数:7
相关论文
共 26 条
[1]   Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements [J].
Amoore, John N. ;
Vacher, Emilie ;
Murray, Ian C. ;
Mieke, Stephan ;
King, Susan T. ;
Smith, Fiona E. ;
Murray, Alan .
BLOOD PRESSURE MONITORING, 2007, 12 (05) :297-305
[2]  
[Anonymous], 1987, NAT STAND EL AUT SPH
[3]   Tilt test alters accuracy of the oscillometric device Omron HEM-705CP [J].
Czarkowski, M ;
Mikulska, M ;
Zebrowski, M ;
Baran, A ;
Rózanowski, K .
JOURNAL OF HUMAN HYPERTENSION, 2004, 18 (04) :293-294
[4]   Venous blood redistribution alters the accuracy of Omron HEM-705CP [J].
Czarkowski, M ;
Mikulska, M ;
Zebrowski, M ;
Baran, A ;
Rózanowski, K .
BLOOD PRESSURE MONITORING, 2004, 9 (01) :9-12
[5]   Accuracy of SpaceLabs 90207 is altered by venous blood redistribution [J].
Czarkowski, M ;
Mikulska, M ;
Baran, A ;
Zebrowski, M ;
Rózanowski, K .
BLOOD PRESSURE, 2003, 12 (5-6) :334-339
[6]   Validation of the Omron HEM-907 device for blood pressure measurement [J].
El Assaad, MA ;
Topouchian, JA ;
Darné, BM ;
Asmar, RG .
BLOOD PRESSURE MONITORING, 2002, 7 (04) :237-241
[7]   A comparison of two sphygmomanometers that may replace the traditional mercury column in the healthcare workplace [J].
Elliott, William J. ;
Young, Patrick E. ;
DeVivo, Laura ;
Feldstein, Jeffrey ;
Black, Henry R. .
BLOOD PRESSURE MONITORING, 2007, 12 (01) :23-28
[8]   Blood pressure measurement in hemodialysis patients [J].
Lazar, AE ;
Smith, MC ;
Rahman, M .
SEMINARS IN DIALYSIS, 2004, 17 (04) :250-254
[9]   Weekly averaged blood pressure is more important than a single-point blood pressure measurement in the risk stratification of dialysis patients [J].
Moriya, Hidekazu ;
Oka, Machiko ;
Maesato, Kyoko ;
Mano, Tsutomu ;
Ikee, Ryota ;
Ohtake, Takayasu ;
Kobayashi, Shuzo .
CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2008, 3 (02) :416-422
[10]   Comparison of dinamap PRO-100 and mercury sphygmomanometer blood pressure measurements in a population-based study [J].
Ni, HY ;
Wu, C ;
Prineas, R ;
Shea, S ;
Liu, K ;
Kronmal, R ;
Bild, D .
AMERICAN JOURNAL OF HYPERTENSION, 2006, 19 (04) :353-360