A study of the re-equilibration of hydrophilic interaction columns with a focus on viability for use in two-dimensional liquid chromatography

被引:14
|
作者
Seidl, Claudia [1 ,2 ]
Bell, David S. [3 ]
Stoll, Dwight R. [1 ]
机构
[1] Gustavus Adolphus Coll, Dept Chem, St Peter, MN 56082 USA
[2] Univ Sao Paulo, Fac Filosofia Ciancias & Letras Ribeirao Preto, Dept Chem, Ribeirao Preto, SP, Brazil
[3] Restek Corp, 110 Benner Circle, Bellefonte, PA USA
基金
美国国家科学基金会; 巴西圣保罗研究基金会;
关键词
HILIC; Re-equilibration; 2D-LC; Repeatability; Mobile phase effects; REVERSED-PHASE; GRADIENT ELUTION; PEAK SHAPE; MASS-SPECTROMETRY; BUFFERED ELUENTS; RETENTION; BASES;
D O I
10.1016/j.chroma.2019.460484
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In recent years there has been increasing interest in the use of HILIC separations in two-dimensional liquid chromatography (2D-LC), mainly because the selectivity of HILIC separations complement that of reversed-phase separations for a variety of molecules. Historically, the re-equilibration of HILIC phases following gradient elution has been perceived as too slow to be useful in the second dimension of 2D-LC separations in particular. Recent studies of re-equilibration of HILIC phases by McCalley and coworkers using a limited set of conditions showed that highly repeatable gradient separations could be obtained with re-equilibration times as short as 4.3 min [1,2]. In this study we aimed to study re-equilibration of HILIC phases under a broader set of conditions, and at much shorter re-equilibration times, in the interest of determining whether or not HILIC separations can be generally considered as a viable option for use in the second dimension of 2D-LC separations. To this end we studied the effects of mobile phase pH, buffer concentration, and preparation method, flow rate, analyte and stationary phase chemistry, column length, and re-equilibration time on retention of a variety of small molecule probe solutes following gradient elution. In general, we have found that excellent separation repeatability can be obtained with quite short (<< 10 min) re-equilibration times, even when progress toward full equilibration of the column is quite slow (>> 10 min). In other words, even if the stationary phase is not fully equilibrated, as long as it is partially equilibrated in a highly precise manner, highly repeatable retention times can be obtained. Higher flow rate has a positive effect on both the rate of progress toward full equilibration and the repeatability of separation. No significant, consistent effects of eluent pH or buffer concentration on repeatability of gradient separation were observed for the stationary phases studied. Excellent gradient separation repeatability was obtained with shorter columns (30 mm length) with re-equilibration times as short as 3 s. A proof-of-concept 2D-LC separation of several small molecule probes using HILIC separations in both dimensions was performed to illustrate that re-equilibration of these columns can be fast enough for HILIC columns to be considered as a viable option for the second dimension of comprehensive 2D-LC separations. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:11
相关论文
共 38 条
  • [1] Optimization of comprehensive two-dimensional gradient chromatography coupling in-line hydrophilic interaction and reversed phase liquid chromatography
    Jandera, Pavel
    Hajek, Tomas
    Stankova, Magda
    Vynuchalova, Katerina
    Cesla, Petr
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1268 : 91 - 101
  • [2] Separation of nonionic surfactants according to functionality by hydrophilic interaction chromatography and comprehensive two-dimensional liquid chromatography
    Abrar, Shazia
    Trathnigg, Bernd
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (52) : 8222 - 8229
  • [3] Comprehensive characterization of Stevia Rebaudiana using two-dimensional reversed-phase liquid chromatography/hydrophilic interaction liquid chromatography
    Fu, Qing
    Guo, Zhimou
    Zhang, Xiuli
    Liu, Yanfang
    Liang, Xinmiao
    JOURNAL OF SEPARATION SCIENCE, 2012, 35 (14) : 1821 - 1827
  • [4] The development of an evaluation method for capture columns used in two-dimensional liquid chromatography
    Cao, Liwei
    Yu, Danhua
    Wang, Xinliang
    Ke, Yanxiong
    Jin, Yu
    Liang, Xinmiao
    ANALYTICA CHIMICA ACTA, 2011, 706 (01) : 184 - 190
  • [5] Impact of conditioning runs on hydrophilic interaction chromatography repeatability and its application as a second dimension in online comprehensive two-dimensional liquid chromatography
    Gely, Clemence
    Monneau, Yoan R.
    Hologne, Maggy
    Faure, Karine
    JOURNAL OF SEPARATION SCIENCE, 2024, 47 (9-10)
  • [6] Study of Separation and Identification of the Active Ingredients in Gardenia jasminoides Ellis Based on a Two-Dimensional Liquid Chromatography by Coupling Reversed Phase Liquid Chromatography and Hydrophilic Interaction Liquid Chromatography
    Zhou, Xuan
    Chen, Cen
    Ye, Xiaolan
    Song, Fenyun
    Fan, Guorong
    Wu, Fuhai
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2017, 55 (01) : 75 - 81
  • [7] Development of orthogonal two-dimensional hydrophilic interaction chromatography systems with the introduction of novel stationary phases
    Liu, Yanming
    Guo, Zhimou
    Feng, Jiatao
    Xue, Xingya
    Zhang, Feifang
    Xu, Qing
    Liang, Xinmiao
    JOURNAL OF SEPARATION SCIENCE, 2009, 32 (17) : 2871 - 2876
  • [8] Monolithic capillary columns based on silica and zirconium oxides for use in hydrophilic interaction liquid chromatography
    da Silva, Carla G. A.
    Collins, Carol H.
    Grespan Bottoli, Carla Beatriz
    MICROCHEMICAL JOURNAL, 2014, 116 : 249 - 254
  • [9] Fully automatable two-dimensional hydrophilic interaction liquid chromatography-reversed phase liquid chromatography with online tandem mass spectrometry for shotgun proteomics
    Zhao, Yun
    Kong, Ricky P. W.
    Li, Guohui
    Lam, Maggie P. Y.
    Law, C. H.
    Lee, Simon M. Y.
    Lam, Herman C.
    Chu, Ivan K.
    JOURNAL OF SEPARATION SCIENCE, 2012, 35 (14) : 1755 - 1763
  • [10] Two-dimensional hydrophilic interaction chromatography coupling anion-exchange and hydrophilic interaction columns for separation of 2-pyridylamino derivatives of neutral and sialylated N-glycans
    Deguchi, Kisaburo
    Keira, Takuro
    Yamada, Kuriko
    Ito, Hiroki
    Takegawa, Yasuhiro
    Nakagawa, Hiroaki
    Nishimura, Shin-Ichiro
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1189 (1-2) : 169 - 174