Quantification of the Multiplexing Effects in Multi-Pinhole Small Animal SPECT: A Simulation Study

被引:46
|
作者
Mok, Greta S. P. [1 ]
Wang, Yuchuan [2 ]
Tsui, Benjamin M. W. [3 ]
机构
[1] Chinese Univ Hong Kong, Dept Diagnost Radiol & Organ Imaging, Hong Kong, Hong Kong, Peoples R China
[2] Harvard Univ, Sch Med, Dana Farber Canc Inst, Boston, MA 02115 USA
[3] Johns Hopkins Univ, Dept Radiol, Baltimore, MD 21287 USA
关键词
Multi-pinhole collimator; multiplexing; small animal imaging; SPECT; HIGH-RESOLUTION; RECONSTRUCTION; SYSTEM; DESIGN;
D O I
10.1109/TNS.2009.2023444
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Our goal is to study the trade-off between image degradation and improved detection efficiency and resolution from allowing multiplexing in multi-pinhole (MPH) SPECT, and to determine The optimal pinhole number for; MPH design. We used an analytical 3D MPH projector and two digitized phantoms: the mouse whole body (MOBY) phantom and a hot sphere phantom to generate noise-free and noisy projections, simulating pinhole collimators fitted with pre-studied pinhole patterns. We performed three schemes to achieve different degrees of multiplexing: 1. Fixed magnification and detection efficiency; 2. Fixed detection efficiency and changed magnification; 3. Fixed magnification and changed detection efficiency. We generated various noisy data sets by simulating Poisson noise using differently scaled noise-free projections and obtained 20 noise realizations for each setting. All datasets; were reconstructed using 3D MPH ML-EM reconstruction method. We analyzed the quantitative accuracy by the normalized-mean-square-error. We evaluated the image contrast for the hot sphere phantom simulation, and also the image noise by the average normalized-standard-deviation of certain pixels for different degrees of multiplexing. Generally no apparent artifacts were observed in the reconstructed images, illustrating the effectiveness of reconstructions. Bias increased for increased degree of multiplexing. Contrast was not significantly affected by multiplexing in the specific simulation scheme (1). Scheme (2) showed that excessive multiplexing to improve image resolution would not improve the overall trade-off of bias and noise compared to no multiplexing. However, scheme (3) showed that when comparing to no multiplexing, the trade-off improved initially with increased multiplexing by allowing more number of pinholes to improve detection efficiency. The trade-off reached a maximum and decreased with further multiplexing due to image degradation from increased bias. The optimal pinhole number was 7 for a compact camera with size of 12 cm x 12 cm and 9 for a standard. gamma camera with size of 40 cm x 40 cm in this scheme. We conclude that the gains in improved detection efficiency and resolution by increased multiplexing are offset by increased image degradations. All the aforementioned factors must be considered in the optimum MPH collimator design for small animal SPECT imaging.
引用
收藏
页码:2636 / 2643
页数:8
相关论文
共 50 条
  • [1] Multi-pinhole small animal SPECT
    Schramm, N
    Schurrat, T
    Béhé, M
    Lackas, C
    Hoppin, J
    Ebel, G
    Behr, TM
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S236 - S236
  • [2] Small animal imaging with multi-pinhole SPECT
    Nuyts, Johan
    Vunckx, Kathleen
    Defrise, Michel
    Vanhove, Christian
    METHODS, 2009, 48 (02) : 83 - 91
  • [3] Evaluation of multiplexing in multi-pinhole SPECT
    Jorgensen, Aaron K.
    Zeng, Gengsheng L.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2213 - 2217
  • [4] Spectral Multiplexing Multi-Pinhole Collimator Design forHigh Sensitivity Small Animal SPECT
    Wang, Wenying
    Feng, Tao
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [5] Multi-pinhole SPECT for small animal research.
    Schramm, NU
    Engeland, U
    Ebel, G
    Schurrat, T
    Behe, M
    Behr, TM
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 223P - 223P
  • [6] The effects of object activity distribution on multiplexing multi-pinhole SPECT
    Mok, Greta S. P.
    Tsui, Benjamin M. W.
    Beekman, Freek J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08): : 2635 - 2650
  • [7] Data completeness in multiplexing multi-pinhole SPECT
    Van Audenhaege, Karen
    Vanhove, Christian
    Vandenberghe, Stefaan
    Van Holen, Roel
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [8] Effects of object activity distribution on multiplexing multi-pinhole SPECT
    Mok, Greta
    Tsui, Benjamin
    Beekman, Freek
    Wang, Yuchuan
    Ahuja, Anil
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [9] Performance of a multi-pinhole animal SPECT
    Schramm, NU
    Schipper, M
    Schurrat, T
    Béhé, M
    Alfke, H
    Engeland, U
    Ebel, G
    Behr, TA
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2077 - 2079
  • [10] Simulation of the Effects of Multiplexing in Multi-Pinhole SPECT Using Stacked Si-HPGe Detectors
    Johnson, Lindsay C.
    Shokouhi, Sepideh
    Peterson, Todd E.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2569 - 2575