Reliability of point-of-care prothrombin time testing in a community clinic: a randomized crossover comparison with hospital laboratory testing

被引:44
作者
Shiach, CR
Campbell, B
Poller, L
Keown, M
Chauhan, N
机构
[1] Manchester Royal Infirm, Dept Clin Haematol, Manchester M13 9WL, Lancs, England
[2] Univ Manchester, Sch Biol Sci, ECAA Cent Facil, Manchester, Lancs, England
关键词
point-of-care prothrombin time testing; warfarin; community medicine; ISI calibration;
D O I
10.1046/j.1365-2141.2002.03888.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The success in achieving therapeutic international normalized ratio (INR) targets in the control of warfarin using a whole-blood point-of-care testing (POCT) monitor (CoaguChek) in a community clinic was compared with hospital laboratory coagulometer prothrombin time (PT) testing in a randomized crossover study. Forty-six patients were randomized into two groups. At each visit, capillary blood was taken for the POCT monitor and venous blood for the laboratory coagulometer. In Group 1, for 6 months, dosage was based on the CoaguChek and for the second 6 months on the coagulometer. In the second group, the order was reversed. Dosages were determined using the DAWN AC computer programme. Success was assessed by the percentage of time patients were maintained within the INR targets. Agreement between laboratory and monitor INR, and patient satisfaction were also assessed. Results with the POCT monitor compared well with the hospital coagulometer. Time in INR target range between the groups was similar, with 60.9% on the POCT monitor and 59.3% with the laboratory coagulometer in Group 1 and in Group 2, respectively, 64.3% and 63.4% with no significant difference in mean INR. An INR above 4.0 gave some discrepant results. International Sensitivity Index calibrations of the two test systems indicated that the INRs were dependable. Patient questionnaires showed greater satisfaction with community POCT monitoring.
引用
收藏
页码:370 / 375
页数:6
相关论文
共 18 条
  • [1] EVALUATION OF A PORTABLE PROTHROMBIN TIME MONITOR FOR HOME-USE BY PATIENTS WHO REQUIRE LONG-TERM ORAL ANTICOAGULANT-THERAPY
    ANDERSON, DR
    HARRISON, L
    HIRSH, J
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1993, 153 (12) : 1441 - 1447
  • [2] Managing oral anticoagulant therapy
    Ansell, J
    Hirsh, J
    Dalen, J
    Bussey, H
    Anderson, D
    Poller, L
    Jacobson, A
    Deykin, D
    Matchar, D
    [J]. CHEST, 2001, 119 (01) : 22S - 38S
  • [3] COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1995, 346 (8982): : 1085 - 1087
  • [4] Cai ZJ, 1999, WHO TECH REP SER, V887, P1
  • [5] Oral anticoagulation self-management and management by a specialist anticoagulation clinic:: a randomised cross-over comparison
    Cromheecke, ME
    Levi, M
    Colly, LP
    de Mol, BJM
    Prins, MH
    Hutten, BA
    Mak, R
    Keyzers, KCJ
    Büller, HR
    [J]. LANCET, 2000, 356 (9224) : 97 - 102
  • [6] Accuracy of a portable international normalization ratio monitor in outpatients receiving long-term oral anticoagulant therapy: Comparison with a laboratory reference standard using clinically relevant criteria for agreement
    Douketis, JD
    Lane, A
    Milne, J
    Ginsberg, JS
    [J]. THROMBOSIS RESEARCH, 1998, 92 (01) : 11 - 17
  • [7] Oral anticoagulation management in primary care with the use of computerized decision support and near-patient testing -: A randomized, controlled trial
    Fitzmaurice, DA
    Hobbs, FDR
    Murray, ET
    Holder, RL
    Allan, TF
    Rose, PE
    [J]. ARCHIVES OF INTERNAL MEDICINE, 2000, 160 (15) : 2343 - 2348
  • [8] 2-PERIOD CROSSOVER CLINICAL-TRIAL
    HILLS, M
    ARMITAGE, P
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1979, 8 (01) : 7 - 20
  • [9] Oral anticoagulants: Mechanism of action, clinical effectiveness, and optimal therapeutic range
    Hirsh, J
    Dalen, JE
    Anderson, DR
    Poller, L
    Bussey, H
    Ansell, J
    Deykin, D
    [J]. CHEST, 2001, 119 (01) : 8S - 21S
  • [10] ACCURACY AND PRECISION OF A PORTABLE ANTICOAGULATION MONITOR IN A CLINICAL SETTING
    MCCURDY, SA
    WHITE, RH
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1992, 152 (03) : 589 - 592