Complications of foldable intraocular lenses requiring explantation or secondary intervention - 1998 survey

被引:52
作者
Mamalis, N [1 ]
机构
[1] Univ Utah, Moran Eye Ctr, Salt Lake City, UT 84132 USA
关键词
D O I
10.1016/S0886-3350(00)00424-7
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A survey of the complications associated with foldable intraocular lenses (IOLs) that required explantation or secondary intervention was sent to domestic members of the American Society of Cataract and Refractive Surgery in 1998. Preoperative data as well as signs and symptoms were obtained and postoperative results and outcomes assessed. The types of foldable IOLs explanted were tabulated. The percentage of each type removed and the common reasons for removal were as follows: 38%-3-piece monofocal silicone IOLs for incorrect lens power followed by dislocation/decentration; 31%-3-piece multifocal silicone IOLs for glare/optical aberrations; 16%-3-piece acrylic IOLS for glare/optical aberrations followed by incorrect lens power and dislocation; 15%-1-piece or plate-haptic silicone IOLs for dislocation/decentration followed by incorrect lens power. Visual outcomes were uniformly good in the 4 groups. Meticulous surgical technique , IOL insertion, and IOL power measurements, along with proper patient selection, are important factors in avoiding complications with foldable IOLs. J Cataract Refract Surg 2000; 26:766-772 (C) 2000 ASCRS and ESCRS.
引用
收藏
页码:766 / 772
页数:7
相关论文
共 17 条
[1]   COMPLICATIONS OF INTRAOCULAR LENSES - A HISTORICAL AND HISTOPATHOLOGICAL REVIEW [J].
APPLE, DJ ;
MAMALIS, N ;
LOFTFIELD, K ;
GOOGE, JM ;
NOVAK, LC ;
KAVKAVANNORMAN, D ;
BRADY, SE ;
OLSON, RJ .
SURVEY OF OPHTHALMOLOGY, 1984, 29 (01) :1-54
[2]   A comparison of patient comfort during cataract surgery with topical anesthesia versus topical anesthesia and intracameral lidocaine [J].
Crandall, AS ;
Zabriskie, NA ;
Patel, BCK ;
Burns, TA ;
Mamalis, N ;
Malmquist-Carter, LA ;
Yee, R .
OPHTHALMOLOGY, 1999, 106 (01) :60-66
[3]   COMPLICATIONS OF SILICONE DISC INTRAOCULAR LENSES [J].
DUNCKER, GIW ;
WESTPHALEN, S ;
BEHRENDT, S .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (05) :562-566
[4]   Analysis of edge glare phenomena in intraocular lens edge designs [J].
Holladay, JT ;
Lang, A ;
Portney, V .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (06) :748-752
[5]   Decentration of 3-piece versus plate-haptic silicone intraocular lenses [J].
Hwang, IP ;
Clinch, TE ;
Moshifar, M ;
Malmquist, L ;
Cason, M ;
Crandall, AS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (11) :1505-1508
[6]   Postoperative decentration of three-piece silicone intraocular lenses [J].
Kimura, W ;
Kimura, T ;
Sawada, T ;
Taguchi, H ;
Hirokane, C ;
Ohte, A ;
Hitani, H .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1277-1280
[7]   A SURVEY OF INTRAOCULAR-LENS EXPLANTATIONS [J].
KRAFF, MC ;
SANDERS, DR ;
RAANAN, MG .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1986, 12 (06) :644-650
[8]  
Leaming DV, 1999, J CATARACT REFR SURG, V25, P851
[9]  
Lyle WA, 1996, J CATARACT REFR SURG, V22, P1456
[10]   Comparison of two plate-haptic intraocular lenses in a rabbit model [J].
Mamalis, N ;
Omar, O ;
Veiga, J ;
Tanner, D ;
Pirayesh, A ;
Fernquist, DS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1291-1295