The effect of measurement geometry on patient specific QA pass/fail rates for stereotactic body radiation therapy (SBRT) Plans

被引:2
作者
Hadsell, Courtney S. [1 ]
Lenards, Nishele. [1 ]
Hunzeker, Ashley. [1 ]
Tallhamer, Michael J. [2 ]
Hadsell, Michael J. [2 ]
机构
[1] Univ Wisconsin, Med Dosimetry Program, La Crosse, WI 54601 USA
[2] Centura Hlth, Dept Radiat Oncol, Centennial, CO USA
关键词
Patient QA; Gamma Index; SBRT; Pass Rate; Replan; MODULATED ARC THERAPY; IMRT;
D O I
10.1016/j.meddos.2021.05.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Patient quality assurance (QA) is a required part of the treatment care path, and plan failure can lead to increased personnel hours or delay of treatment. The recommendation by the American Association of Physicists in Medicine is that gamma analysis be used to evaluate measured volumetric modulated arc therapy plans. Vendors have developed many different measurement geometries for patient QA devices which could yield varying pass rates when used with the recommended tolerances, normalization, and criterion. For this study, clinically treated stereotactic body radiation therapy plans were used to evaluate differences in gamma dose tolerances and sampled dose distribution complexity for centralized or peripheral measurement geometries on a cylindrical phantom. Random errors were then introduced into a subset of these plans, and the differences in pass rates between the geometries were correlated with differences in the observed mathematical differences. Finally, a single clinically relevant target coverage deviation was introduced to another subset of plans to evaluate whether a particular geometry is measurably better at identifying clinically relevant errors. It was found that centralized geometries resulted in more lenient dose tolerances and less complex sampled dose distributions compared to peripheral geometries. Pass rates were uniformly lower in the peripheral measurement geometry, and the difference in pass rates between the geometries correlated strongly with the difference in dose tolerance and weakly with the difference in the chosen complexity metrics. However, neither of the geometries were sufficiently sensitive enough to detect clinically relevant changes to target coverage when using recommended tolerances and criteria, and no statistically significant difference was found between their pass rates. Given these findings, the authors concluded that stereotactic body radiation therapy plans could fail patient QA when measured in the peripheral geometry but pass in the centralized geometry, with possibly neither having correlation to true clinical deviation.
引用
收藏
页码:389 / 397
页数:9
相关论文
共 20 条
  • [1] Evaluation of the Delta4 phantom for IMRT and VMAT verification
    Bedford, James L.
    Lee, Young K.
    Wai, Philip
    South, Christopher P.
    Warrington, Alan P.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) : N167 - N176
  • [2] 3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification
    Carrasco, Pablo
    Jornet, Nuria
    Latorre, Artur
    Eudaldo, Teresa
    Ruiz, Agusti
    Ribas, Montserrat
    [J]. MEDICAL PHYSICS, 2012, 39 (08) : 5040 - 5049
  • [3] Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics
    Coleman, Linda
    Skourou, Christina
    [J]. MEDICAL PHYSICS, 2013, 40 (11)
  • [4] Gimeno-Olmos Jose, 2013, RADIOTHER ONCOL, DOI [10.1016/S0167- 8140(15, DOI 10.1016/S0167-8140(15]
  • [5] Jabbari Keyvan, 2018, J Med Signals Sens, V8, P31, DOI 10.4103/jmss.JMSS_15_17
  • [6] Dosimetric Discrepancies due to Positional Errors in MLC Movement During Stereotactic Lung VMAT
    Karan, T.
    Kim, S.
    Abbas, A.
    Moseley, D. J.
    Taremi, M. M.
    Yeung, I.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S190 - S190
  • [7] Evaluation of the ArcCHECK QA system for IMRT and VMAT verification
    Li, Guangjun
    Zhang, Yingjie
    Jiang, Xiaoqin
    Bai, Sen
    Peng, Guang
    Wu, Kui
    Jiang, Qingfeng
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (03): : 295 - 303
  • [8] Patient-Specific Quality Assurance Protocol for Volumetric Modulated Arc Therapy using Dose Volume Histogram
    Low, Christopher
    Toye, Warren
    Phung, Peter
    Huston, Christopher
    [J]. JOURNAL OF MEDICAL PHYSICS, 2018, 43 (02) : 112 - 118
  • [9] A technique for the quantitative evaluation of dose distributions
    Low, DA
    Harms, WB
    Mutic, S
    Purdy, JA
    [J]. MEDICAL PHYSICS, 1998, 25 (05) : 656 - 661
  • [10] Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218
    Miften, Moyed
    Olch, Arthur
    Mihailidis, Dimitris
    Moran, Jean
    Pawlicki, Todd
    Molineu, Andrea
    Li, Harold
    Wijesooriya, Krishni
    Shi, Jie
    Xia, Ping
    Papanikolaou, Nikos
    Low, Daniel A.
    [J]. MEDICAL PHYSICS, 2018, 45 (04) : E53 - E83