Biomechanical Evaluation of Mandibular Condyle Fracture Osteosynthesis Using the Rhombic Three-Dimensional Condylar Fracture Plate

被引:10
作者
Ben Achour, Anas [1 ]
Meissner, Heike [2 ]
Teichet, Uwe [1 ]
Haim, Dominik [3 ]
Range, Ursula [4 ]
Brosius, Alexander [1 ]
Leonhardt, Henry [3 ]
Lauer, Gunter [3 ]
机构
[1] Tech Univ Dresden, Forming Proc, Inst Mfg Technol, George Bahr Str 3c, D-01069 Dresden, Germany
[2] Tech Univ Dresden, Dept Prosthet Dent, Univ Hosp Carl Gustav Carus, Dresden, Germany
[3] Tech Univ Dresden, Dept Oral & Maxillofacial Surg, Univ Hosp Carl Gustav Carus, Dresden, Germany
[4] Tech Univ Dresden, Inst Med Informat & Biometry, Univ Hosp Carl Gustav Carus, Dresden, Germany
关键词
MASTICATORY SYSTEM; INTERNAL-FIXATION; CLOSED TREATMENT; BONE; BITE; STRENGTH; TRAUMA; MODEL; JOINT; PIG;
D O I
10.1016/j.joms.2019.04.020
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The aim of the present study was to evaluate the biomechanical stability of osteosynthesis in mandibular condyle fractures using a newly designed rhombic 3-dimensional (3D) condylar fracture plate and compare it with that using standard two 4-hole miniplates and with that in nonfractured condyles. Materials and Methods: Using 200 porcine mandibles, 3 different monocortical plating techniques were evaluated. The condyles were fractured along a defined line tangentially through the sigmoid notch and perpendicular to the posterior border. After anatomic reduction, osteosynthesis was performed using either standard rhombic 3D condylar fracture plates and standard screws (group A) or locking rhombic 3D condylar fracture plates, which were fixed either with standard screws (group B) or locking screws (group C). For comparison, nonfractured condyles (group D) and condyles fixed with standard two 4-hole miniplates and 8 screws (group E) were included. Using a universal mechanical testing machine (TIRA Test 2720; TIRA GmbH Schalkau, Germany), each group was subjected to linear loading from laterally to medially, medially to laterally, anteriorly to posteriorly, and posteriorly to anteriorly. The maximum axial force and displacement at the maximum force were measured. The mean values were compared for statistical significance using analysis of variance with Bonferroni's correction (statistical significance set at P < .05). Results: The main mode of failure in the plating techniques investigated was the pull out of screws from the proximal fragment. We found no statistically significant differences in the stability of osteosynthesis between the two 4-hole miniplates and the rhombic 3D condylar fracture plate when loading from posteriorly to anteriorly, laterally to medially, and medially to laterally. However, when loading from anteriorly to posteriorly, a statistically significant difference between the standard and locking system and the two 4hole miniplate system was observed, with the latter proving more stable. Conclusions: The results of the present biomechanical study suggest that the rhombic 3D condylar frac- ture plates are suitable for the treatment of condylar neck fractures. Both types of the plate are able to resist physiologic strains comparable to the two 4-hole miniplates. (C) 2019 American Association of Oral and Maxillofacial Surgeons
引用
收藏
页码:1868.e1 / 1868.e15
页数:15
相关论文
共 37 条
  • [1] Interspecies differences in bone composition, density, and quality:: Potential implications for in vivo bone research
    Aerssens, J
    Boonen, S
    Lowet, G
    Dequeker, J
    [J]. ENDOCRINOLOGY, 1998, 139 (02) : 663 - 670
  • [2] Biomechanical comparison of plating techniques for fractures of the mandibular condyle
    Alkan, Alper
    Metin, Murat
    Muglali, Mehtap
    Oezden, Bora
    Celebi, Nuekhet
    [J]. BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2007, 45 (02) : 145 - 149
  • [3] Substitutes for human cadaveric hone in maxillofacial rigid fixation research
    Bredbenner, TL
    Haug, RH
    [J]. ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2000, 90 (05): : 574 - 580
  • [4] Development plates for stable internal fixation: Study of mechanical resistance in simulated fractures of the mandibular condyle
    Celegatti Filho, Toride Sebastiao
    Rodrigues, Danillo Costa
    Lauria, Andrezza
    Fernandes Moreira, Roger William
    Consani, Simonides
    [J]. JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2015, 43 (01) : 158 - 161
  • [5] Choi BH, 2001, J ORAL MAXIL SURG, V59, P734, DOI 10.1053/joms.2001.24283
  • [6] A comparison of open and closed treatment of condylar fractures: a change in philosophy
    De Riu, G
    Gamba, U
    Anghinoni, M
    Sesenna, E
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2001, 30 (05) : 384 - 389
  • [7] Depprich R, 2007, ORAL MAXILLOFAC SURG, V11, P19, DOI 10.1007/s10006-006-0037-1
  • [8] Open versus closed treatment of fractures of the mandibular condylar process - a prospective randomized multi-centre study
    Eckelt, Uwe
    Schneider, Matthias
    Erasmus, Francois
    Gerlach, Klaus Louis
    Kuhlisch, Eberhard
    Loukota, Richard
    Rasse, Michael
    Schubert, Johannes
    Terheyden, Hendrik
    [J]. JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2006, 34 (05) : 306 - 314
  • [9] Surgical complications with open treatment of mandibular condylar process fractures
    Ellis, E
    McFadden, D
    Simon, P
    Throckmorton, G
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2000, 58 (09) : 950 - 958
  • [10] Treatment of mandibular condylar process fractures: Biological considerations
    Ellis, E
    Throckmorton, GS
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 63 (01) : 115 - 134