Simple and convenient analytical method for the direct determination of chlorine species by ETV-ICP-AES using tungsten boat furnace vaporiser and exchangeable small sample cuvettes

被引:4
作者
Kojima, Norihisa [1 ]
Mizoguchi, Yasuhiko [1 ]
Tanabe, Kenji [1 ]
Iida, Yutaka [1 ]
Hashimoto, Bunji [2 ]
Uchihara, Hiroshi [2 ]
Ohshita, Yuji [3 ]
Okamoto, Yasuaki [4 ]
机构
[1] Toray Res Ctr Ltd, Otsu, Shiga 5208567, Japan
[2] Horiba Ltd, Applicat R&D Ctr, Kyoto 6018510, Japan
[3] Rakuyo Giken Co Ltd, Kumiyama Cho, Kyoto 6130023, Japan
[4] Hiroshima Univ, Grad Sch Sci, Higashihiroshima 7398526, Japan
关键词
INDUCTIVELY-COUPLED PLASMA; ATOMIC-EMISSION-SPECTROMETRY; MICROWAVE-INDUCED COMBUSTION; MASS-SPECTROMETRY; ION CHROMATOGRAPHY; ELECTROTHERMAL VAPORIZATION; ABSORPTION-SPECTROMETRY; POLYMERS; BROMINE; SYSTEM;
D O I
10.1016/j.polymertesting.2017.02.008
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
A tungsten boat furnace vaporiser and an exchangeable sample cuvette system were applied for the direct determination of chlorine in plastic samples by electrothermal vaporisation inductively coupled plasma atomic emission spectrometry. Each piece of plastic samples was moulded into a thin film, and weighed into small sample cuvettes. An ethanolic solution of the modifier potassium hydroxide and an organic solvent were added to each sample cuvette. The contents including the plastic piece were spread over the bottom of each sample cuvette by warming on a hotplate. After the preparation, one of the cuvettes was placed in the tungsten boat furnace equipped with the electrothermal vaporiser. The temperature was gradually elevated. At the flash vaporisation step of 1800 degrees C, the analyte was vaporised and introduced into the ICP to measure an atomic emission assignable to Cl I 134.724 nm. A widely spread thin layer of the plastic facilitated its rapid evaporation and introduction into an ICP atomic emission spectrometer. Since all the chlorine species were on-furnace decomposed to form potassium chloride during the pretreatment, the solid samples were analysed using a calibration curve prepared from the aqueous standard solutions. By using this system, a detection limit of 1.5 mu g g(-1) of chlorine in solid plastic samples was established when a sample amount taken of 2 mg was studied. The exchangeable sample cuvette technique makes it possible to measure a number of sample sequentially by preparing a lot of sample cuvettes prior to analysis. Approximately 20 batches could be vaporised per hour. Analytical results of some plastic samples that were analysed by this technique are given in this report. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:262 / 267
页数:6
相关论文
共 34 条
  • [1] Development of a vaporization system for direct determination of chlorine in petroleum coke by ICP-MS
    Antes, Fabiane Goldschmidt
    Dullius, Eduardo
    Ben da Costa, Adilson
    Molz, Rolf Fredi
    Gottfried Paniz, Jose Neri
    Moraes Flores, Erico Marlon
    Dressler, Valderi L.
    [J]. MICROCHEMICAL JOURNAL, 2013, 109 : 117 - 121
  • [2] BALLSCHMITER K, 1996, DIOXINE
  • [3] Budavari S., 1989, MERCK INDEX, VEleventh
  • [4] Determination of halides by microwave induced plasma and stabilized capacitive plasma atomic emission spectrometry after on-line continuous halogen generation
    Camuna, JF
    Montes, M
    Pereiro, R
    SanzMedel, A
    Katschthaler, C
    Gross, R
    Knapp, G
    [J]. TALANTA, 1997, 44 (04) : 535 - 544
  • [5] Simultaneous determination of bromine and chlorine in coal using electrothermal vaporization inductively coupled plasma mass spectrometry and direct solid sample analysis
    de Gois, Jefferson S.
    Pereira, Ederson R.
    Welz, Bernhard
    Borges, Daniel L. G.
    [J]. ANALYTICA CHIMICA ACTA, 2014, 852 : 82 - 87
  • [6] Application of direct solid sample analysis for the determination of chlorine in biological materials using electrothermal vaporization inductively coupled plasma mass spectrometry
    de Gois, Jefferson Santos
    Pereira, Ederson R.
    Welz, Bernhard
    Borges, Daniel L. G.
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2015, 105 : 12 - 17
  • [7] Duarte FA, 2013, QUIM NOVA, V36, P716, DOI 10.1590/S0100-40422013000500017
  • [8] TURBIDIMETRIC DETERMINATION OF CHLORINE IN CHLOROBUTYL AND OTHER CHLORINE CONTAINING POLYMERS AT LOW-LEVELS
    FALCON, JZ
    LOVE, JL
    GAETA, LJ
    ALTENAU, AG
    [J]. ANALYTICAL CHEMISTRY, 1975, 47 (01) : 171 - 173
  • [9] Determination of iodine in food samples by inductively coupled plasma mass spectrometry after alkaline extraction
    Fecher, PA
    Goldmann, I
    Nagengast, A
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (09) : 977 - 982
  • [10] Determination of halogens in coal after digestion using the microwave-induced combustion technique
    Flores, Erico M. M.
    Mesko, Marcia F.
    Moraes, Diogo P.
    Pereira, Juliana S. F.
    Mello, Paola A.
    Barin, Juliano S.
    Knapp, Guenter
    [J]. ANALYTICAL CHEMISTRY, 2008, 80 (06) : 1865 - 1870